RAYSEARCH LABORATORIES AB recalls RayCare 2024A SP1. Oncology Information Systems.
Beam set delivery note issues in which there is potential data loss if the beam set delivery note is modified simultaneously and/or if a note is updated during a treatment session, the user is not notified of the change.
Read the official notice →- Date
- 2026-07-08
- Country
- United States
- Brand
- Raysearch Laboratories Ab
- Product
- RayCare 2024A SP1. Oncology Information Systems.
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Raysearch Laboratories Ab recalls
- 2026-01-28
US
RAYSEARCH LABORATORIES AB recalls Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0, 17.0.1 Software Version: RayStation v2025,…
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material…
FDA · DeviceRAYSEARCH LABORATORIES ABRefer to the attached documents: Distribution list 167168 US - updated,…Official notice →- 2025-09-10
US
RAYSEARCH LABORATORIES AB recalls RayStation with the following product descriptions: 1. RayStation 11B, Software Version: 12.0.0.932. 2. RayStation 11B Service Pack 1,…
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
FDA · DeviceRAYSEARCH LABORATORIES ABWorldwide - US Nationwide distribution in the states of AL, CA, FL, GA, IL,…Official notice →- 2025-02-26
US
RAYSEARCH LABORATORIES AB recalls Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 1.0.0.951, 11.0.1.29,…
Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT…
FDA · DeviceRAYSEARCH LABORATORIES ABWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2024-05-08
US
RAYSEARCH LABORATORIES AB recalls RayStation 13.0.0.1547, 13.1.0.144 & 13.1.1.89. Radiation Therapy Treatment Planning System.
Potential for reported SSD to be too high.
FDA · DeviceRAYSEARCH LABORATORIES ABWorldwide distribution - US Nationwide and the countries of Algeria,…Official notice →- 2024-04-17
US
RAYSEARCH LABORATORIES AB recalls RayStation Software Version Numbers version numbers 8B, 9A, 9B, 10A, 10B, 11A, 11B, 12A, 12B, 2023B, 2024A including some service packs.
A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.
FDA · DeviceRAYSEARCH LABORATORIES ABWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →- 2023-11-08
US
RAYSEARCH LABORATORIES AB recalls RayCare 5B and 6A including service packs- An oncology information system used to support workflows and scheduling, as well as clinical…
An issue was found in RayCare 5B, 6A, including service packs, where values will not be visible for documents in a read-only state when opened outside the Documents workspace when using document fields configured with…
Other drugs & medical recalls in the United States
- 2026-07-08
US
Covidien LLC recalls Covidien Shiley Disposable Decannulation Plug REF DDCP UDI-DI code: 10884522004060 Product Codes that include the Covidien Shiley…
Due to product labeling incorrectly including the symbol indicating non-DEHP (Bis(2-ethylhexyl) phthalate as the product includes 30 to 35% DEHP.
FDA · DeviceCovidien LLCU.S. Nationwide distribution including in the states and territories of AK,…Official notice →US
Carl Zeiss Meditec AG recalls Zeiss Intrabeam Spherical Applicator 1.5 cm REF:304534-6000-557
Spherical applicators (radiation therapy accessory) failed to meet manufacturing specifications.
FDA · DeviceCarl Zeiss Meditec AGWorldwide - US Nationwide distribution in the state and territories of Florida…Official notice →US
Siemens Medical Solutions USA, Inc recalls ARTIS icono ceiling. Model Number: 11328100.
Potential issue with the Display ceiling suspension (DCS Fix XL) that holds the monitor in place. Over time some screws that keep the monitor attached to the suspension may gradually loosen.
FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide distribution in the states of CA, GA, IA, IN, LA, MO, NC, NY,…Official notice →US
Medela Inc recalls Brand Name: Medela Product Name: ENFit ExSet Strl 60 2.0ml PV 1ct EN Model/Catalog Number: ENF060182LD / 101033086 Software Version: N/A…
May contain incorrect connectors and will not attach or misconnect to ENFit connections and is thus incompatible. Issue could lead to slight delay in delivery of enteral nutritional fluids or oral medication.
FDA · DeviceMedela IncUS distribution to the following states: AZ, CA, CO, FL, HI, IL, IN, MD, MI,…Official notice →US
Accuray Incorporated recalls Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator
In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room…
FDA · DeviceAccuray IncorporatedWorldwide - US Nationwide and and the countries of Australia, Austria,…Official notice →US
Siemens Medical Solutions USA, Inc recalls ARTIS icono floor. Model Number: 11327700.
Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.