Drugs & medicalClass II

Philips North America recalls SmartPath to dStream for 3.0T. Product Code (REF): 782145. MR systems with SW version R11.1 and R12.1.

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Read the official notice →
Date
2026-02-04
Country
United States
Brand
Philips
Product
SmartPath to dStream for 3.0T. Product Code (REF): 782145. MR systems with SW version R11.1 and R12.1.
Classification
Class II
Units
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Philips recalls

  1. 2026-06-17
  2. US
    Drugs & medicalClass II

    Philips North America Llc recalls DigitalDiagnost C90 Flex/Value/Chest/ER. Model Number: 712035.

    System was delivered without the required Seismic kit for Wall Stand VS2, which is designed to prevent damage and ensure stability of the Wall Stand VS2 during seismic events.

    FDA · DevicePhilips North America LlcUS distribution to California.Official notice →
  3. 2026-05-06
  4. US
    Drugs & medicalClass II

    Philips North America recalls Philips SmartPath to dStream for 3.0T with MR Elastography (MRE). 1. Model Number (REF): 782145.

    The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE…

    FDA · DevicePhilips North AmericaWorldwide - US Nationwide and the countries of Argentina, Australia,…Official notice →
  5. 2026-02-18
  6. US
    Drugs & medicalClass II

    Philips North America Llc recalls Brilliance iCT; Product Code (REF): 728306;

    As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners…

    FDA · DevicePhilips North America LlcWorldwide - US Nationwide distribution in the states of AL, AR, CA, CO, CT,…Official notice →
  7. 2026-01-28
  8. US
    Drugs & medicalClass II

    Philips North America recalls Patient Information Center iX; Software Version Number: 4.5.0

    A recent software (SW) patch modifies Mobile Event Notification filter settings without providing any indication to the user when upgrading the system.

    FDA · DevicePhilips North AmericaWorldwide distribution - US Nationwide and the countries of Austria, Belgium,…Official notice →
  9. 2026-01-14
  10. US
    Drugs & medicalClass II

    Philips North America Llc recalls Philips Smart-hopping 2.0 AP 1.4 GHz. Patient Monitor.

    It was found that the MX40 device could not reconnect to the PIC iX when moving between Radiohead and Trident 1.4 GHz access points if the signal strength changed quickly.

    FDA · DevicePhilips North America LlcUS Nationwide distribution.Official notice →
  11. 2025-12-17
  12. US
    Drugs & medicalClass II

    Philips North America recalls Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM

    Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance…

    FDA · DevicePhilips North AmericaU.S.Official notice →
All 100 Philips recalls →

Other drugs & medical recalls in the United States

  1. 2026-02-04
  2. US
    Drugs & medicalClass II

    LSL Healthcare Inc. recalls LSL Healthcare, Central Line Dressing Kit, Model/Catalog Number: 2519CP, medical procedure convenience kit

    BD ChloraPrep Triple Swabsticks, a component within the kits, exhibit an open seal on the packaging of the applicators.

    FDA · DeviceLSL Healthcare Inc.US Nationwide distribution in the state of Idaho.Official notice →
  3. US
    Drugs & medicalClass II

    LimFlow, Inc. recalls Brand Name: LimFlow Vector Product Name: LimFlow Vector Model/Catalog Number: VT-US-23

    Incorrect expiration date

    FDA · DeviceLimFlow, Inc.US Nationwide distribution in the states of AL, CT, District of Columbia, FL,…Official notice →
  4. US
    Drugs & medicalClass II

    CareFusion 303, Inc. recalls BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a modular infusion pump…

    Due to a number of modules flashed with a date and time associated with the daylight savings time (DST) adjustment that may cause connectivity issue with hospital networks. This issue only affects the Interoperability…

    FDA · DeviceCareFusion 303, Inc.U.S. Nationwide distribution in the states of AZ, IN, OH, and VA.Official notice →
  5. US
    Drugs & medicalClass II

    Cardinal Health 200, LLC recalls Chest Drainage Units and Accessories: Product Code Description 1. 8888571562 Cardinal Health Sentinel Seal CDU 2. 8888571513 Cardinal…

    The firm is updating the IFU to clarify the intended patient population for the identified chest drainage units and accessories as adults, 18 years and older. When used on infants, the CDU may not demonstrate…

    FDA · DeviceCardinal Health 200, LLCWorldwide distribution - US Nationwide and the countries of Brazil, Canada,…Official notice →
  6. US
    Drugs & medicalClass II

    Zimmer Surgical Inc recalls Brand Name: Zimmer Tourniquet Systems Product Name: A.T.S.¿ 3200TS Tourniquet Systems Model/Catalog Number: 60320010100 Software Version:…

    Device user interface (screen) may freeze or become unresponsive to touch when operating in certain non-English language settings. Issue may cause delay or extension while alternate device is located.

    FDA · DeviceZimmer Surgical IncInternational distribution to the countries of Canada and EMEA only.Official notice →
  7. US
    Drugs & medicalClass II

    Medline Industries, LP recalls Medline medical procedure convenience kits labeled as: 1) WATKINS SPINE PACK-LF, Kit SKU DYNJ0275635M; 2) SM-SPINE PACK-LF, Kit SKU…

    Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.

    FDA · DeviceMedline Industries, LPUS Nationwide distribution.Official notice →
See all drugs & medical recalls →