Philips North America recalls Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM
Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance…
Read the official notice →- Date
- 2025-12-17
- Country
- United States
- Brand
- Philips
- Product
- Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM
- Classification
- Class II
- Units
- U.S.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Philips recalls
- 2026-02-04
US
Philips North America recalls SmartPath to dStream for 3.0T. Product Code (REF): 782145. MR systems with SW version R11.1 and R12.1.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
FDA · DevicePhilips North AmericaWorldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA,…Official notice →- 2026-01-28
US
Philips North America recalls Patient Information Center iX; Software Version Number: 4.5.0
A recent software (SW) patch modifies Mobile Event Notification filter settings without providing any indication to the user when upgrading the system.
FDA · DevicePhilips North AmericaWorldwide distribution - US Nationwide and the countries of Austria, Belgium,…Official notice →- 2026-01-14
US
Philips North America Llc recalls Philips Smart-hopping 2.0 AP 1.4 GHz. Patient Monitor.
It was found that the MX40 device could not reconnect to the PIC iX when moving between Radiohead and Trident 1.4 GHz access points if the signal strength changed quickly.
- 2025-12-10
US
Philips North America Llc recalls CT 5300; Software Version Number: 4.5, 5.0, 5.1;
If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems…
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Albania, Algeria,…Official notice →US
Philips North America Llc recalls IntelliVue MP90. Product Number: M8010A.
Potential issue where the IntelliVue monitors did not alarm.
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Afghanistan,…Official notice →- 2025-10-22
US
Philips North America recalls Cardiac Workstation 5000; Model Number: 860439;
Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.
FDA · DevicePhilips North AmericaUS Nationwide distribution in the states of CA, IA, TN, TX, VA.Official notice →
Other drugs & medical recalls in the United States
- 2025-12-17
US
Bard Peripheral Vascular Inc recalls Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,
Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt…
FDA · DeviceBard Peripheral Vascular IncUS Nationwide distribution in the states of CA, MN, IN, MI, LA.Official notice →US
Medline Industries, LP recalls MEDLINE PERFUSION SYRINGE, REF: DYNJ39223C
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing BD Luer Tip Caps. BD has issued a recall of their Luer Tip Caps after confirming through internal testing that…
US
GE Medical Systems, LLC recalls Senographe Pristina, Pristina Serena, and Pristina Serena 3D. for screening and diagnostic mammography
Some Senographe Pristina systems either have a missing or an incorrect X-Ray Warning label, which is a nonconformance to FDA Regulation 21 CFR ¿1020.30(j).
US
Draeger, Inc. recalls ErgoStar CM 60, Model/Catalog Number: MP01860, Airway connector for conduction of respiratory gases between the breathing system and the…
Multiple complaints were reported in which cracks formed in the hose of specific catheter mounts.
FDA · DeviceDraeger, Inc.Worldwide distribution. US states of CA, FL, IL, NY, OH, OK, PA, SC, TX, and…Official notice →US
Lupin Pharmaceuticals Inc. recalls Sertraline Hydrochloride Tablets USP, 100 mg, 90 Tablets bottles, Rx Only, Manufactured by: Lupin Pharmaceuticals, Inc., Naples, FL…
Defective container - seal not adhering to bottles
US
Cipla USA, Inc. recalls Cinacalcet Hydrochloride Tablets, 90 mg, 30 Tablets per bottle, Rx Only, Manufactured for: Cipla USA, Inc., 10 Independence Boulevard,…
CGMP Deviations: the presence of a nitrosamine impurity, N-nitroso-cinacalcet, above the acceptable daily intake (ADI) limits.