Philips North America Llc recalls CT 5300; Software Version Number: 4.5, 5.0, 5.1;
If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems…
Read the official notice →- Date
- 2025-12-10
- Country
- United States
- Brand
- Philips
- Product
- CT 5300; Software Version Number: 4.5, 5.0, 5.1;
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Philips recalls
- 2026-01-28
US
Philips North America recalls Patient Information Center iX; Software Version Number: 4.5.0
A recent software (SW) patch modifies Mobile Event Notification filter settings without providing any indication to the user when upgrading the system.
FDA · DevicePhilips North AmericaWorldwide distribution - US Nationwide and the countries of Austria, Belgium,…Official notice →- 2026-01-14
US
Philips North America Llc recalls Philips Smart-hopping 2.0 AP 1.4 GHz. Patient Monitor.
It was found that the MX40 device could not reconnect to the PIC iX when moving between Radiohead and Trident 1.4 GHz access points if the signal strength changed quickly.
- 2025-12-17
US
Philips North America recalls Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM
Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance…
- 2025-12-10
US
Philips North America Llc recalls IntelliVue MP90. Product Number: M8010A.
Potential issue where the IntelliVue monitors did not alarm.
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Afghanistan,…Official notice →- 2025-10-22
US
Philips North America recalls Cardiac Workstation 5000; Model Number: 860439;
Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.
FDA · DevicePhilips North AmericaUS Nationwide distribution in the states of CA, IA, TN, TX, VA.Official notice →- 2025-10-01
US
Philips North America Llc recalls Philips CT systems labeled as the following: 1. Brilliance iCT, Model Number: 728306.
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
Other drugs & medical recalls in the United States
- 2025-12-11
US
Uuoeebb Infant Walkers Recalled Due to Risk of Serious Injury or Death from Fall and Entrapment Hazards; Violate Mandatory Standard for Infant Walkers; Sold on Amazon…
The recalled infant walkers violate the mandatory standard for infant walkers because they can fit through a standard doorway and fail to stop at the edge of a step, posing a deadly fall hazard. In addition, the infant…
- 2025-12-10
US
Illumina, Inc. recalls Reagent Kits/ containing flow cells, components of NextSeq 550Dx Instrument: NextSeq 550Dx High Output Reagent Kit v2.5 (75 cycles) IVD,…
An instrument intended for sequencing DNA libraries with in vitro diagnostic assays is used with reagent kits including flow cells that have flow cell gaskets that have a quality issue, which may lead to flow cell…
FDA · DeviceIllumina, Inc.Worldwide - US Nationwide distribution in the states of PA, NJ, CA, NC, UT,…Official notice →US
Philips Ultrasound, LLC recalls Lumify Diagnostic Ultrasound System models : 795005, 795161, 989605449841,
Ultrasound system compatibility issues with Apple devices running iOS 18 may cause a failure to perform live imagining.
FDA · DevicePhilips Ultrasound, LLCWorldwide US Nationwide distribution including in the states of AL, AR, AZ,…Official notice →US
Accriva Diagnostics, Inc. recalls Hemochron Activated Clotting Time Low-Range Test Cuvette, REF: ACT-LR, to be used with the Hemochron Signature ELITE Whole Blood…
Test cuvettes, used with an assay for monitoring heparin anticoagulation, may have a foreign object that has sharp edges and could pose a risk of injury if not detected prior to handling.
FDA · DeviceAccriva Diagnostics, Inc.Worldwide - US Nationwide distribution including in the states of OH, IL, NY,…Official notice →US
Siemens Medical Solutions USA, Inc recalls MAMMOMAT Inspiration;
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
FDA · DeviceSiemens Medical Solutions USA, IncWorldwide - US Nationwide distribution in the states of AR, CA, NY and the…Official notice →US
Microbiologics Inc recalls KWIK-STIK, Quality control kit for culture media, Vibrio parahaemolyticus derived from ATCC 17802, Catalog Number 0818K
Some products packaged for lot 818-111-7 may have been packaged in pouches labeled as lot 857-53-10.
FDA · DeviceMicrobiologics IncInternational distribution to the countries of UAE, China, Singapore,…Official notice →