Philips North America recalls IntelliVue G7m Anesthesia Gas Module, Product Number 866173
The 866173 lntelliVue G7m Anesthesia Gas Modules could have a component issue that may trigger an interruption of the gas measurement. This issue concerns the G7m's sample gas pump which could stop working, producing a…
Read the official notice →- Date
- 2024-10-09
- Country
- United States
- Brand
- Philips
- Product
- IntelliVue G7m Anesthesia Gas Module, Product Number 866173
- Classification
- Class II
- Units
- US Nationwide. Global Distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Philips recalls
- 2024-10-16
US
Philips North America recalls Patient Information Center iX, Product Code 866386, and Patient Information Center iX Expand, Product Code 866390, Software Revision 4.0.1
During extended operation of the MX40 with the PIC iX, the DHCP (Dynamic Host Configuration Protocol) lease can expire on the MX40. If the DHCP lease expires while the device is offline, upon reconnecting with the PIC…
US
Philips North America Llc recalls IntelliVue Patient Monitor MX400/MX450/MX500/MX550 with software version P.01.01 and ordered with Option CP2 are affected as follows:…
Devices with Option CP2 and initially programmed to software version P.01.01 were configured incorrectly at the factory. Therefore, Enhanced ECG Capabilities provided by Option CP2 were not enabled in these devices.…
FDA · DevicePhilips North America LlcForeign Only: Denmark France Germany Japan United KingdomOfficial notice →- 2024-10-09
US
Philips North America Llc recalls DIVA 24 Inch Widescreen LCD Touch Display - Used with the Philips PIC iX system central monitoring system. Model: RGD2461AMI Philips…
If the DIVA touch display generates touch inputs without user interaction, due to possibility of the erroneous acknowledgement of a patient alarm, there is a potential for a delay in the detection of a change in a…
FDA · DevicePhilips North America LlcWorldwide - US Nationwide distribution in the states of CA, IL, ME, NC, NJ,…Official notice →- 2024-10-02
US
Philips North America recalls IntelliVue Patient Monitor MX400 (Model Number 866060), MX430 (Model Number 866061), MX450 (Model Number 866062), MX500 (Model Number…
During a production process, Philips became aware of one IntelliVue power supply with a broken ground bolt upon disconnection of the ground cable from the equipotential ground connector. Loss of electrical grounding…
US
Philips North America Llc recalls Ingenia 3.0T - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal…
Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May…
FDA · DevicePhilips North America LlcWorldwide distribution - US Nationwide and the countries of Argentina,…Official notice →- 2024-08-14
US
Philips North America Llc recalls IntelliVue Patient Monitor MX500 with L.x and M.x; K.x provided with L.x entitlements Model Number: 866064. Indicated for use by health…
Philips lntelliVue Patient Monitors MX400/450/500/550 Shipped or Upgraded with Incorrect Software Options. Changes were made to reflect some options becoming standard capabilities for software-version N.x. In the…
FDA · DevicePhilips North America LlcNationwide Foreign: Australia Austria Bangladesh Brazil Canada Chile China…Official notice →
Other drugs & medical recalls in the United States
- 2024-10-09
US
Outset Medical, Inc. recalls Tablo Console, REF: PN-0003000, PN-0006000, part of the Tablo Hemodialysis System. The Tablo Hemodialysis System is indicated for use in…
Blood leak sensor, a component of hemodialysis device, may show decreased sensitivity, at the lowest level, specified as 0.35 mL/min blood leak rate, which could delay triggering the ALARM_DIALYZER_BLOOD_LEAK alarm, if…
FDA · DeviceOutset Medical, Inc.US: CA, NY, IL, MO, KY, AZ, PA, FL, GA, AK, NM, VA, NC, SC, OK, AL, MA, OH,…Official notice →US
Becton, Dickinson and Company, BD Bio Sciences recalls BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSDuet Sample Preparation System…
Sample preparation system with software may cause: 1) Multidispense feature enabled with 2-tube assay: Specimen volume dispensed into tube1 at least 3% higher than volume in tube2, or 2) Absolute counts calculated…
FDA · DeviceBecton, Dickinson and Company, BD Bio SciencesUS: NJ, UT, WI, TX, ID, AK, MN, MA, MD, TN, MI, CA, VA OUS: Ireland, Austria,…Official notice →US
Young Dental Manufacturing I, LLC recalls BIOTROL PURIT Clean-It General Purpose Ultrasonic Cleaner, REF PC016, net contents 16 fl. oz., makes 16 gallons of solution.
The bulk product used in the recalling firm's repackaged finished product was recalled due to bacterial contamination.
FDA · DeviceYoung Dental Manufacturing I, LLCDistribution was made to AZ, ID, MD, NV, NY, PA, SC, TN, TX, WI. There was no…Official notice →US
Ossur H / F recalls Miami J Select Collar, Catalog: MJS-101; Miami J Select Collar Set, Catalog: MJSR-101
Pressure sensitive adhesive hooks attached to the sides of the anterior cervical orthosis collar panel can occasionally exhibit poor adhesion, which may lead to them gradually peeling off with repeated doffing,…
FDA · DeviceOssur H / FUS: NH, MA, NY, NJ, PA, GA, MD, NC, IN, FL, AR, TN, KY, OH, MI, IL, TX, MO,…Official notice →US
Mercury Enterprises, Inc. dba Mercury Medical recalls Neo-Tee T-Piece Resuscitator. Part Numbers 1050805, 1050832. For pediatric emergency respiratory support.
Manual resuscitator circuit flow controller may come apart due to a missing manufacturing step.
FDA · DeviceMercury Enterprises, Inc. dba Mercury MedicalUS distribution to MI, FL, SC, PA, IN. International distribution to Albania.Official notice →US
Medtronic MiniMed, Inc. recalls MiniMed 620G Insulin Pump, REF: MMT-1510, MMT-1710, MMT-1750; MiniMed 630G Insulin Pump, REF: MMT-1515, MMT-1714, MMT-1715, MMT-1754,…
Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life…
FDA · DeviceMedtronic MiniMed, Inc.US: CT, MI, PA, WA, IA, NY, ND, AZ, TX, OH, NC, AL, MN, IN, NJ, KY, UT, CA,…Official notice →