Outset Medical, Inc. recalls Tablo Console, REF: PN-0003000, PN-0006000, part of the Tablo Hemodialysis System. The Tablo Hemodialysis System is indicated for use in…
Blood leak sensor, a component of hemodialysis device, may show decreased sensitivity, at the lowest level, specified as 0.35 mL/min blood leak rate, which could delay triggering the ALARM_DIALYZER_BLOOD_LEAK alarm, if…
Read the official notice →- Date
- 2024-10-09
- Country
- United States
- Brand
- Outset Medical
- Product
- Tablo Console, REF: PN-0003000, PN-0006000, part of the Tablo Hemodialysis System. The Tablo Hemodialysis System is indicated for use in…
- Classification
- Class II
- Units
- US: CA, NY, IL, MO, KY, AZ, PA, FL, GA, AK, NM, VA, NC, SC, OK, AL, MA, OH,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Outset Medical recalls
- 2024-06-26
US
Outset Medical, Inc. recalls TabloCart with Prefiltration Drawer PN-0006813, with power cords: PN-0005746 (10-foot), PN-0006578 (15-foot), PN-0005747 (20-foot)10-, 15-…
A cart with prefiltration drawer, an optional hemodialysis system accessory, has a cord with a female connector that has internal contact pins that may become damaged that could create localized heat, sparks, smoke,…
FDA · DeviceOutset Medical, Inc.US Nationwide distribution in the states of FL, CA, NC, GA, MD, DE, AK, OK,…Official notice →- 2024-04-24
US
Outset Medical, Inc. recalls Tablo Hemodialysis System, REF: PN-0003000, PN-0006000 and PN-0007001
Hemodialysis System, pre and post dialyzer peroxide-cured silicone tubing, with less than 336 hours usage, will be replaced because tubing may contain non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs), that…
FDA · DeviceOutset Medical, Inc.US:FL, CA, NM, SC, AZ, TN, NC, TX, IL, MD, IN, DE, AL, MI, AK, KY, LA, ID, OK,…Official notice →- 2023-04-19
US
Outset Medical, Inc. recalls Outset Tablo Model Numbers (REF): PN-0006000 PN-0007001 Tablo is a next generation, self-contained hemodialysis system (Hemodialysis…
An observed trend of high conductivity dialysate alarms in a hemodialysis system as a result of introducing a software tool into the manufacturing process.
FDA · DeviceOutset Medical, Inc.U.S. Nationwide distribution including in the states of AK, AL, AR, CA, CO,…Official notice →- 2022-03-30
US
Outset Medical, Inc. recalls Outset Tablo Console Model Number: PN-0003000 Model Number: PN-0006000
Due to a component in the hemodialysis console there is the possibility of heat-related damage that may occur with the device.
FDA · DeviceOutset Medical, Inc.U.S. Nationwide distribution in the states of AL, AK, AZ, CA, CO, CT, DC, DE,…Official notice →
Other drugs & medical recalls in the United States
- 2024-10-09
US
Siemens Healthcare Diagnostics, Inc. recalls Atellica IM aTPO Assay, Material Number 10995466 (100 Test) and 10995467 (500 Test)
Positive bias in patient results at and below the cut-off of 60 U/mL with the lots listed in the table below when compared to alternate lots on the Atellica IM analyzer.
FDA · DeviceSiemens Healthcare Diagnostics, Inc.Worldwide distribution - US Nationwide.Official notice →US
Healgen Scientific Llc recalls Healgen Series Reagent strips for Urinalysis are composed of several color pads aligned on a test strip. Each pad is employed for testing…
The reason for this recall is Healgen Scientific Ltd. has determined that their Urinary Reagent Strip products have shown reduced sensitivity, or potential false negative results for the detection of urinary protein by…
FDA · DeviceHealgen Scientific LlcU.S. Nationwide distribution in the states of CA, GA and TX.Official notice →US
Percussionaire Corporation recalls Percussionaire, Inc., REF: P5-TEE-20, Qty: 20, Rx Only. IPV (Intrapulmonary Percussive Ventilation) therapy In-Line Valve
Updated labeling to ensure that blue cap on in-line valve is removed, prior to delivery of (Intrapulmonary percussive ventilation) IPV therapy.
FDA · DevicePercussionaire CorporationUS: Nationwide OUS: Israel Japan Russia Switzerland Canada TurkeyOfficial notice →US
Becton, Dickinson and Company, BD Bio Sciences recalls BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSDuet Sample Preparation System…
Sample preparation system with software may cause: 1) Multidispense feature enabled with 2-tube assay: Specimen volume dispensed into tube1 at least 3% higher than volume in tube2, or 2) Absolute counts calculated…
FDA · DeviceBecton, Dickinson and Company, BD Bio SciencesUS: NJ, UT, WI, TX, ID, AK, MN, MA, MD, TN, MI, CA, VA OUS: Ireland, Austria,…Official notice →US
Young Dental Manufacturing I, LLC recalls BIOTROL PURIT Clean-It General Purpose Ultrasonic Cleaner, REF PC016, net contents 16 fl. oz., makes 16 gallons of solution.
The bulk product used in the recalling firm's repackaged finished product was recalled due to bacterial contamination.
FDA · DeviceYoung Dental Manufacturing I, LLCDistribution was made to AZ, ID, MD, NV, NY, PA, SC, TN, TX, WI. There was no…Official notice →US
Philips North America recalls IntelliVue G7m Anesthesia Gas Module, Product Number 866173
The 866173 lntelliVue G7m Anesthesia Gas Modules could have a component issue that may trigger an interruption of the gas measurement. This issue concerns the G7m's sample gas pump which could stop working, producing a…