PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion systems not configured with an optional auxiliary pan handle. Includes the below product descriptions and…
Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.
Read the official notice →- Date
- 2026-05-27
- Country
- United States
- Brand
- Philips Medical Systems Nederland
- Product
- Philips Azurion systems not configured with an optional auxiliary pan handle. Includes the below product descriptions and corresponding…
- Classification
- Class II
- Units
- Nationwide distribution. International distribution to Albania, Algeria,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Philips Medical Systems Nederland recalls
- 2026-06-10
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion 3M15, Model Numbers: 722064, 722222 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with…
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without…
- 2026-06-03
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063, 722221), Azurion 3 M15 (722064, 722222),…
It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific…
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide and the countries of Canada,…Official notice →- 2026-05-27
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063), Azurion 3 M15 (722064), Azurion 7 B12…
The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance…
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Domestic: Nationwide; International: Canada, Algeria, Argentina, Australia,…Official notice →- 2026-05-20
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221, 722229), Azurion 3 M15 (722222, 722230),…
Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or…
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide and the country of Canada.Official notice →- 2026-04-22
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips AneurysmFlow. Model Number: 001015. AneurysmFlow is a software medical device (Interventional Tool) intended to…
Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information…
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide and the countries of Argentina,…Official notice →- 2026-04-08
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Vue Motion V12. Product Number: 1017979.
Potential that mis-ordered frames in Vue Motion during dynamic cine runs may cause images frames to display out of sequence.
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Worldwide distribution - US Nationwide and the countries of Algeria, Andorra,…Official notice →
Other drugs & medical recalls in the United States
- 2026-06-03
US
AVEVA Drug Delivery Systems, Inc. recalls Nicotine Transdermal System Patch, 14 mg, 14 patches per box, Aveva, Distributed by: RUGBY LABORATORIES, Indianapolis, IN…
Failed Impurities/Degradation Specifications: Elevated levels of oxidative-related impurities, exceeding shelf-life specifications during stability testing of individual units.
- 2026-05-27
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063), Azurion 3 M15 (722064), Azurion 7 B12…
The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance…
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Domestic: Nationwide; International: Canada, Algeria, Argentina, Australia,…Official notice →US
C.R. Bard Inc recalls Silastic Foley Catheters Silastic Brand Foley Catheter 30cc REF: 33416 33418 33420 33422 33424 Silastic Brand Foley Catheter 5cc REF:…
Due to stain present on the surface of affected foley catheters.
FDA · DeviceC.R. Bard IncU.S.: AK, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD,…Official notice →US
Medtronic Perfusion Systems recalls Medtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94725; Cardiopulmonary bypass vascular catheter
Certain lots of product have the potential for a sterile barrier breach.
FDA · DeviceMedtronic Perfusion SystemsWorldwide distribution - US Nationwide and the countries of Australia,…Official notice →US
Novapproach Spine, LLC recalls OneLIF Interbody Cage (REF/Description): 010-107-1007/Small 7 Degree 10mm X 7mm, 010-107-1007-2/Small 7 Degree 10mm X 7mm - Oblique,…
Intervertebral body fusion system straight inserter failed to properly attach to affected cages, which may result in an inability to securely engage the implant with the inserter, which could lead to procedural delay…
FDA · DeviceNovapproach Spine, LLCUS Nationwide distribution in the states of NY, CT, CA, TX, FL, MI.Official notice →US
ICU Medical, Inc. recalls Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; ChemoLock; Syringe Transfer…
Oncology sets/closed system transfer devices with extension sets may leak resulting in therapy delayed if noticed and corrected during setup, but if the leak occurs during therapy this may result in therapy…
FDA · DeviceICU Medical, Inc.US Nationwide distribution in the states of GA, MA, OH, NC, CA, ND, MI, TX,…Official notice →