Medtronic Perfusion Systems recalls Medtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94725; Cardiopulmonary bypass vascular catheter
Certain lots of product have the potential for a sterile barrier breach.
Read the official notice →- Date
- 2026-05-27
- Country
- United States
- Brand
- Medtronic Perfusion Systems
- Product
- Medtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94725; Cardiopulmonary bypass vascular catheter
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide and the countries of Australia,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Medtronic Perfusion Systems recalls
- 2026-08-05
US
Medtronic Perfusion Systems recalls EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula,…
Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ…
- 2026-07-22
US
Medtronic Perfusion Systems recalls Medtronic Bio-Medicus Adult Cannula Kit, Product Number/CFN 96530-115; Cardiopulmonary bypass vascular cannula
Medtronic received two complaints of loose foreign material confirmed as containing trace levels of blood.
FDA · DeviceMedtronic Perfusion SystemsWorldwide - US Nationwide distribution in the states of MD, CA, MI, WA, PA,…Official notice →- 2026-07-01
US
Medtronic Perfusion Systems recalls Octopus Evolution AS Tissue Stabilizer, Model TS2500
During the manufacturing process, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached…
- 2026-03-04
US
Medtronic Perfusion Systems recalls Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product Number TSMICS1
During the manufacturing, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the…
- 2025-11-26
US
Medtronic Perfusion Systems recalls Medtronic HR-ACT (High Range Activated Clotting Time Cartridges), REF 402-03
Medtronic personnel observed trace amounts of dry blood on the external carton label and Instructions for Use (IFU) in five boxes of one lot of ACT Cartridges.
FDA · DeviceMedtronic Perfusion SystemsUS Nationwide distribution in the states of Hawaii and New York.Official notice →US
Medtronic Perfusion Systems recalls MC3 VitalFlow Console, REF 58100; Blood pump of ecmo
As of August 6, 2025, Medtronic has received eleven reports of VitalFlow Consoles displaying an E70 error code during normal operation. When this occurs, the touch screen may become temporarily unresponsive and go…
Other drugs & medical recalls in the United States
- 2026-05-27
US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063), Azurion 3 M15 (722064), Azurion 7 B12…
The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance…
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Domestic: Nationwide; International: Canada, Algeria, Argentina, Australia,…Official notice →US
C.R. Bard Inc recalls Silastic Foley Catheters Silastic Brand Foley Catheter 30cc REF: 33416 33418 33420 33422 33424 Silastic Brand Foley Catheter 5cc REF:…
Due to stain present on the surface of affected foley catheters.
FDA · DeviceC.R. Bard IncU.S.: AK, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD,…Official notice →US
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion systems not configured with an optional auxiliary pan handle. Includes the below product descriptions and…
Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.
FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Nationwide distribution. International distribution to Albania, Algeria,…Official notice →US
Novapproach Spine, LLC recalls OneLIF Interbody Cage (REF/Description): 010-107-1007/Small 7 Degree 10mm X 7mm, 010-107-1007-2/Small 7 Degree 10mm X 7mm - Oblique,…
Intervertebral body fusion system straight inserter failed to properly attach to affected cages, which may result in an inability to securely engage the implant with the inserter, which could lead to procedural delay…
FDA · DeviceNovapproach Spine, LLCUS Nationwide distribution in the states of NY, CT, CA, TX, FL, MI.Official notice →US
ICU Medical, Inc. recalls Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; ChemoLock; Syringe Transfer…
Oncology sets/closed system transfer devices with extension sets may leak resulting in therapy delayed if noticed and corrected during setup, but if the leak occurs during therapy this may result in therapy…
FDA · DeviceICU Medical, Inc.US Nationwide distribution in the states of GA, MA, OH, NC, CA, ND, MI, TX,…Official notice →US
TANGENT ENDOSCOPY, LLC recalls Tangent Endoscopy, Tangent Single Use Digital Catheter. Model Numbers: (1) TNG4002-IND. (2) TNG-4007-6pk.
Affected devices exhibited fractures at the distal shaft tip without complete detachment.
FDA · DeviceTANGENT ENDOSCOPY, LLCUS Nationwide distribution in the states of North Carolina, Arizona, and Nevada.Official notice →