Drugs & medicalClass II

Merit Medical Systems, Inc. recalls Flex-Neck Catheter External Repair Kit REF CE-1400 LOT CE-1400/A

Product that was built for design verification testing was inadvertently distributed to customers.

Read the official notice →
Date
2023-04-19
Country
United States
Brand
Merit Medical Systems
Product
Flex-Neck Catheter External Repair Kit REF CE-1400 LOT CE-1400/A
Classification
Class II
Units
U.S. Nationwide distribution in the states of MD, MI, OH, and TX. O.U.S.: N/A
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Merit Medical Systems recalls

  1. 2024-05-29
  2. US
    Drugs & medicalClass II

    Merit Medical Systems, Inc. recalls Prelude Guide Sheath Roadster, REF PG5F45S018, 5F 0.076 (1.9 mm), 45 cm, STERILE EO, RX ONLY

    Their is a potential that sheaths labeled with .018 dilators may contain a .038 dilator.

    FDA · DeviceMerit Medical Systems, Inc.Worldwide - US Nationwide distribution in the state of CA and the countries of…Official notice →
  3. 2024-01-10
  4. US
    Drugs & medicalClass II

    Merit Medical Systems, Inc. recalls Merit Maestro Microcatheter, Reference Numbers: 28MC2111045/B, 28MC21110ST/B, 28MC2113045/B, 28MC21130SN/CNB, 28MC21130ST/B,…

    The sterility of microcatheter and infusion system devices cannot be guaranteed.

    FDA · DeviceMerit Medical Systems, Inc.Worldwide - US Nationwide distribution in the states of AL, AZ, CA, CO, CT,…Official notice →
  5. 2023-06-21
  6. US
    Drugs & medicalClass II

    Merit Medical Systems, Inc. recalls Aspira Pleural Drainage Catheter Insertion Tray 15.5F, REF: 4992509, Version A, Sterile EO, Rx Only

    There is a potential that valve assemblies will not open. preventing fluid from draining.

    FDA · DeviceMerit Medical Systems, Inc.Worldwide - US Nationwide distribution in the states of AK, AL, AR, AZ, CA,…Official notice →
  7. 2023-05-17
  8. US
    Drugs & medicalClass II

    Merit Medical Systems, Inc. recalls COBRA-OS, custom Sheath Introducer Kit, 4 French, REF: E01P2, Sterile EO, Single Use, RX Only, 5 Units

    Custom sheath Introducer contains incorrect needle size

    FDA · DeviceMerit Medical Systems, Inc.US Distribution to state of: MinnesotaOfficial notice →
  9. 2022-11-02
  10. US
    Drugs & medicalClass II

    Merit Medical Systems, Inc. recalls CorVocet Biopsy System/Catalog Code: CORA1610/B, CORA1610S/B, CORA1615/B, CORA1810/A, CORA1810/B, CORA1810S/B, CORA1815/A,…

    Biopsy Systems have components that are misaligned due to an issue with the manufacturing assembly equipment, which may lead to the needle ejecting when triggered outside the body, which may result in user or patient…

    FDA · DeviceMerit Medical Systems, Inc.Worldwide distribution - US National distribution including in the states of…Official notice →
  11. 2022-04-20
  12. US
    Drugs & medicalClass II

    Merit Medical Systems, Inc. recalls AccessPLUS Large Bore Hemostasis Valves, Catalog Code: 580758001, 580774001, IN8152/C, K04-01197, K05T-01863, K09-09102B, K09-10388A,…

    A design change made to a hemostasis valve to improve manufacturability caused an internal gap between the rotator and the Y-body, where the guidewire can get caught, resulting in difficulty advancing the guidewire…

    FDA · DeviceMerit Medical Systems, Inc.US: MN, PA, AZ, FL, IL, AK, NY, CA, TX, KY, WI, NC, NJ, VA, CO, IN, AR, MO,…Official notice →
All 22 Merit Medical Systems recalls →

Other drugs & medical recalls in the United States

  1. 2023-04-19
  2. US
    Drugs & medicalClass II

    Philips North America Llc recalls CombiDiagnost R90 1.0 and 1.1 multi-functional general R/F system, Model Numbers 709030 and 709031

    If the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell…

    FDA · DevicePhilips North America LlcUS Nationwide. Global Distribution.Official notice →
  3. US
    Drugs & medicalClass II

    Solta Medical Inc recalls thermage THERMAGE CPT SYSTEM REF TG-2B *** Solta Medical, Inc. thermage RF GENERATOR REF TG-2B *** Solta Medical, Inc. Dermatologic and…

    A service unit was potentially improperly tested and calibrated during manufacturing on a failed test station.

    FDA · DeviceSolta Medical IncU.S. Distribution to state of: MDOfficial notice →
  4. US
    Drugs & medicalClass II

    Shanghai United Imaging Healthcare Co., Ltd. recalls Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography (PET) and Computed Tomography (CT), Model…

    Due to a software issue where the process of patient scanning, the scatter correction may occasionally fail with will potentially cause a failure of the PET image reconstruction and generation of PET images.

    FDA · DeviceShanghai United Imaging Healthcare Co., Ltd.U.S. Nationwide distribution in the state of TX. O.U.S.: Not provided.Official notice →
  5. US
    Drugs & medicalClass II

    Outset Medical, Inc. recalls Outset Tablo Model Numbers (REF): PN-0006000 PN-0007001 Tablo is a next generation, self-contained hemodialysis system (Hemodialysis…

    An observed trend of high conductivity dialysate alarms in a hemodialysis system as a result of introducing a software tool into the manufacturing process.

    FDA · DeviceOutset Medical, Inc.U.S. Nationwide distribution including in the states of AK, AL, AR, CA, CO,…Official notice →
  6. US
    Drugs & medicalClass II

    Church & Dwight Inc recalls Easy Read App associated with First Response Pregnancy Test Sticks - Intended use of converting analog pregnancy test results into text…

    Church & Dwight First Response Pregnancy Kits was marketed with the Easy Read App without a new 510k. There is also a potential for Easy Read App to be unable to reads the test results (i.e., no results reported on the…

    FDA · DeviceChurch & Dwight IncUS Nationwide distribution.Official notice →
  7. US
    Drugs & medicalClass I

    Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS), Large Volume Pump, Model No. LVP-0004

    Fluid ingress that can cause a loss of electrical funtion and failure of the Set ID Sensory, resulting in the prevention of infusion or halt of infusion.

    FDA · DeviceFresenius Kabi USA, LLCDomestic: CA, CO, NJ, WI, & UT. No foreign distribution.Official notice →
See all drugs & medical recalls →