Drugs & medicalClass II

Shanghai United Imaging Healthcare Co., Ltd. recalls Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography (PET) and Computed Tomography (CT), Model…

Due to a software issue where the process of patient scanning, the scatter correction may occasionally fail with will potentially cause a failure of the PET image reconstruction and generation of PET images.

Read the official notice →
Date
2023-04-19
Country
United States
Brand
Shanghai United Imaging Healthcare
Product
Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography (PET) and Computed Tomography (CT), Model uMI 550 Part Number:…
Classification
Class II
Units
U.S. Nationwide distribution in the state of TX. O.U.S.: Not provided.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Shanghai United Imaging Healthcare recalls

  1. 2026-10-07
  2. US
    Drugs & medicalClass II

    Shanghai United Imaging Healthcare Co., Ltd. recalls Computed Tomography X-ray System, Model: uCT 530 with software

    Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images…

    FDA · DeviceShanghai United Imaging Healthcare Co., Ltd.US Nationwide distribution in the states of CA, PR, UT, WV, ND, NE, NC, TX,…Official notice →
  3. 2022-10-19
  4. US
    Drugs & medicalClass II

    Shanghai United Imaging Healthcare Co., Ltd. recalls Positron Emission Tomography and Computed Tomography System, Model: uMI 550

    The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and respiratory signal loss…

    FDA · DeviceShanghai United Imaging Healthcare Co., Ltd.Distribution to US states of TX, WI, CA, and MIOfficial notice →
  5. 2022-05-18
  6. US
    Drugs & medicalClass II

    Shanghai United Imaging Healthcare Co., Ltd. recalls Digital Medical X-ray Imaging System, Model: uDR 596i, uDR 592h

    X-ray imaging system positioning image and protocol label is reversed for Flexion and Extension on C-Spine and L-Spine, this will cause the image to be incorrectly labeled and may cause the patient to have a repeat…

    FDA · DeviceShanghai United Imaging Healthcare Co., Ltd.US Nationwide distribution in the state of TX.Official notice →
  7. 2021-10-06
  8. US
    Drugs & medicalClass II

    Shanghai United Imaging Healthcare Co., Ltd. recalls uEXPLORER PET/CT System - Product Usage: intended to be operated by qualified healthcare professionals to assist in…

    The EXAM acquisition workflow could occasionally freeze after canceling the PET scan.

    FDA · DeviceShanghai United Imaging Healthcare Co., Ltd.U.S. Nationwide distribution in the states of CA, IL, IN, TX, and WI.Official notice →
All 5 Shanghai United Imaging Healthcare recalls →

Other drugs & medical recalls in the United States

  1. 2023-04-19
  2. US
    Drugs & medicalClass III

    AGILENT TECHNOLOGIES INC./US recalls Agilent Resolution ctDx FIRST Sample Collection Kit, 500032

    Distributed sample collection kit with an unapproved instruction for use.

    FDA · DeviceAGILENT TECHNOLOGIES INC./USU.S. Nationwide distribution in the states of CA, GA, IN, NJ, and TX. O.U.S.:…Official notice →
  3. US
    Drugs & medicalClass II

    Philips North America Llc recalls CombiDiagnost R90 1.0 and 1.1 multi-functional general R/F system, Model Numbers 709030 and 709031

    If the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell…

    FDA · DevicePhilips North America LlcUS Nationwide. Global Distribution.Official notice →
  4. US
    Drugs & medicalClass II

    Solta Medical Inc recalls thermage THERMAGE CPT SYSTEM REF TG-2B *** Solta Medical, Inc. thermage RF GENERATOR REF TG-2B *** Solta Medical, Inc. Dermatologic and…

    A service unit was potentially improperly tested and calibrated during manufacturing on a failed test station.

    FDA · DeviceSolta Medical IncU.S. Distribution to state of: MDOfficial notice →
  5. US
    Drugs & medicalClass II

    Merit Medical Systems, Inc. recalls Flex-Neck Catheter External Repair Kit REF CE-1400 LOT CE-1400/A

    Product that was built for design verification testing was inadvertently distributed to customers.

    FDA · DeviceMerit Medical Systems, Inc.U.S. Nationwide distribution in the states of MD, MI, OH, and TX. O.U.S.: N/AOfficial notice →
  6. US
    Drugs & medicalClass II

    Outset Medical, Inc. recalls Outset Tablo Model Numbers (REF): PN-0006000 PN-0007001 Tablo is a next generation, self-contained hemodialysis system (Hemodialysis…

    An observed trend of high conductivity dialysate alarms in a hemodialysis system as a result of introducing a software tool into the manufacturing process.

    FDA · DeviceOutset Medical, Inc.U.S. Nationwide distribution including in the states of AK, AL, AR, CA, CO,…Official notice →
  7. US
    Drugs & medicalClass II

    Church & Dwight Inc recalls Easy Read App associated with First Response Pregnancy Test Sticks - Intended use of converting analog pregnancy test results into text…

    Church & Dwight First Response Pregnancy Kits was marketed with the Easy Read App without a new 510k. There is also a potential for Easy Read App to be unable to reads the test results (i.e., no results reported on the…

    FDA · DeviceChurch & Dwight IncUS Nationwide distribution.Official notice →
See all drugs & medical recalls →