MEDLINE INDUSTRIES, LP - Northfield recalls (1) Medline Sterile Hand Pack, Item DYNJ63380A, 3 kits per case, contains BD Chloraprep Hi-Lite Orange, 26 mL Applicator,…
The convenience trays had improper placement of some of the components that were in the nonsterile portion of the tray, thus, not receiving sterilization.
Read the official notice →- Date
- 2023-03-01
- Country
- United States
- Brand
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- (1) Medline Sterile Hand Pack, Item DYNJ63380A, 3 kits per case, contains BD Chloraprep Hi-Lite Orange, 26 mL Applicator, Vendor PN:…
- Classification
- Class II
- Units
- US Nationwide distribution in the states of AR, CO, and GA. There was no…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More MEDLINE INDUSTRIES, LP - Northfield recalls
- 2023-06-28
US
MEDLINE INDUSTRIES, LP - Northfield recalls MEDLINE THERAHONEY GEL, 100% LEPTOSPERMUM SCOPARIUM HONEY, a) REF MNK0005, b) REF MNK0015 wound gel
Medline issued a recall on TheraHoney Wound Gel due to the manufacturer notifying Medline of a sterilization dose audit failure on the wound gel and not augmenting the dose after failure per ISO 11137.
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide distribution - US Nationwide and the countries of SINGAPORE,…Official notice →- 2023-06-07
US
MEDLINE INDUSTRIES, LP - Northfield recalls Procedure packs: (1) Medline Sterile Arthroscopy Pack, REF: DYNJ64242B; (2) Medline Sterile Shoulder Pack, REF: DYNJ64250A;…
The kits were damaged by water.
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldDistribution was made to California. There was no foreign/government/military…Official notice →- 2023-04-26
US
MEDLINE INDUSTRIES, LP - Northfield recalls Centurion CirClamp with 1.3cm Bell & Insert Reprocessed, Reorder #CR230K, Sterile.
The CirClamp subassembly found in the kit was the incorrect size.
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS Nationwide distribution in the states of CA and TX. There was no…Official notice →- 2023-02-22
US
MEDLINE INDUSTRIES, LP - Northfield recalls Surgical convenience kits labeled as: MEDLINE ARM/MINOR DRAPE CSTM, Reorder Number DYNJCD0063
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
- 2023-02-15
US
MEDLINE INDUSTRIES, LP - Northfield recalls Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINOPLASTY CDS…
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide distribution - US Nationwide and the countries of Canada, Japan,…Official notice →- 2023-02-08
US
MEDLINE INDUSTRIES, LP - Northfield recalls MEDLINE UNITE FOOT & ANKLE DRILL BIT, 1.3MM Solid Core, REF MPN10013
The recall is due to observed intra-operative screw failure. When pre-drilled with ¿1.3mm drill bit for ¿2.0mm screws and ¿1.6mm drill bit for ¿2.4mm we have observed screw failure due to excessive torque.
Other drugs & medical recalls in the United States
- 2023-03-08
US
Alvogen, Inc recalls Levothyroxine Sodium Tablets, USP 112 mcg, 90 tablets per bottle, Rx Only, Manufactured by Lloyd Inc., Shenandoah, IA, 51601, Distributed…
Sub-Potent Drug: Out of specification for assay at the 24 month interval.
US
DuPont Nutrition USA, Inc recalls BD-102 NF, Microcrystalline Cellulose, NF, Ph. Eur., Jp., Net Content/Gross Weight 20.0 KG / 21.1 KG bulk container, Manufactured by:…
Failed Impurities/Degradation Specifications: Out of specification results obtained for conductivity.
FDA · DrugDuPont Nutrition USA, IncProduct was distributed nationwide in the United States and Puerto Rico, and…Official notice →- 2023-03-01
US
Siemens Medical Solutions USA, Inc recalls Luminos dRF Max with software VE10, VF10, VF11- A diagnostic imaging system for radiographic and fluoroscopic studie Model…
Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values
US
S & MOHR AMERICA, S.A. DE C.V. recalls abm Respiratory Care Standart Breating Circuit with Face Mask - Toddler REF BC21087 abm Respiratory Care Standart Breathing…
Incorrect expiration date on product label. The manufacturing date was listed as the expiration date. Correct expiration date is "22.07.2027"
FDA · DeviceS & MOHR AMERICA, S.A. DE C.V.U.S. Nationwide distribution in the state of MN.Official notice →US
Cardinal Health 200, LLC recalls Cardinal Health Presource Packs IV START TRAY I-Intended for IV access device CATALOG #: 01-5799A
Presource Kits contains the ICU MicroClave Clear Connector due to a manufacturing defect there is a gap between the connectors top and bottom housings which may lead to fluid leaks and loss of fluid path integrity
FDA · DeviceCardinal Health 200, LLCUS Nationwide distribution in the states of GA, PA.Official notice →US
Access Vascular, Inc recalls HydroMID 4Fr Single Lumen Maximal Barrier Kit-indicated for short term access (<30days) to the peripheral venous access system for…
Label with the incorrect component listed on the inner kit Tyvek header bag
FDA · DeviceAccess Vascular, IncUS Nationwide distribution in the states of CT, FL, NE, TN, TX, WA.Official notice →