Drugs & medicalClass II

Cardinal Health 200, LLC recalls Cardinal Health Presource Packs IV START TRAY I-Intended for IV access device CATALOG #: 01-5799A

Presource Kits contains the ICU MicroClave Clear Connector due to a manufacturing defect there is a gap between the connectors top and bottom housings which may lead to fluid leaks and loss of fluid path integrity

Read the official notice →
Date
2023-03-01
Country
United States
Brand
Cardinal Health
Product
Cardinal Health Presource Packs IV START TRAY I-Intended for IV access device CATALOG #: 01-5799A
Classification
Class II
Units
US Nationwide distribution in the states of GA, PA.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Cardinal Health recalls

  1. 2023-10-18
  2. US
    Drugs & medicalClass II

    Cardinal Health 200, LLC recalls Procedure pack: Presource PBDS Cat. PLMHCSBHF, Hermann Hospital MHHS, Kit, C Section B MHHS, Sterile.

    Packs contain recalled components (equipment drapes, table cover, and surgical room turnover (SRT) kits).

    FDA · DeviceCardinal Health 200, LLCWorldwide distribution - US Nationwide distribution in the states of AZ, CA,…Official notice →
  3. 2023-07-26
  4. US
    Drugs & medicalClass II

    Cardinal Health 200, LLC recalls NovaPlus Infant Heel Warmer, Instant Squeeze Activation, Catalog Number V11460-010

    Cardinal Health has received an increase in burst/leaking complaints from mid-April to May regarding the above listed lot of the Novalplus Infant Heel Warmer with tape.

    FDA · DeviceCardinal Health 200, LLCUS Nationwide distribution.Official notice →
  5. 2023-06-14
  6. US
    Drugs & medicalClass II

    Cardinal Health 200, LLC recalls Presource Kit, SMA21LPUHJ GYN Laparoscopy Pack. Convenience kit used for Laparoscopic surgery

    Inadvertently released for distibution without sterility assurance.

    FDA · DeviceCardinal Health 200, LLCDistribution in the US to GeorgiaOfficial notice →
  7. 2023-01-25
  8. US
    Drugs & medicalClass II

    Cardinal Health 200, LLC recalls Cardinal Health Jackson-Pratt Channel Drain, 32 Fr, REF JP-2292

    Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including lack of radiopacity in wound drains, dull trocars impacting product…

    FDA · DeviceCardinal Health 200, LLCWorldwide distribution - US Nationwide and the countries of Canada, Saudi…Official notice →
  9. 2022-12-14
  10. US
    Drugs & medicalClass II

    Cardinal Health recalls Dover Closed Urethral Tray with Vinyl Catheter 14 Fr/Ch (4.7 ml) 5ml Catheter Product Code: 2480 Intended for urinary drainage from the…

    Occlusion (blockage) of the bubble inlet port of the urinary drainage bag which may prevent urine draining into the bag. If the bladder is not emptied within a clinically appropriate time, this may increase the risk of…

    FDA · DeviceCardinal HealthNationwideOfficial notice →
  11. 2022-11-16
  12. US
    Drugs & medicalClass II

    Cardinal Health 200, LLC recalls Cardinal Health GEL PACK REUSABLE, SMALL, Cat. 70204

    Users are being informed that the product should not be used on infants/neonates. A caution statement will be added to the labeling stating: Not for use on Infants/Neonates .

    FDA · DeviceCardinal Health 200, LLCnationwide and CanadaOfficial notice →
All 41 Cardinal Health recalls →

Other drugs & medical recalls in the United States

  1. 2023-03-01
  2. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc recalls Luminos dRF Max with software VE10, VF10, VF11- A diagnostic imaging system for radiographic and fluoroscopic studie Model…

    Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values

    FDA · DeviceSiemens Medical Solutions USA, IncUS Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls (1) Medline Sterile Hand Pack, Item DYNJ63380A, 3 kits per case, contains BD Chloraprep Hi-Lite Orange, 26 mL Applicator,…

    The convenience trays had improper placement of some of the components that were in the nonsterile portion of the tray, thus, not receiving sterilization.

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS Nationwide distribution in the states of AR, CO, and GA. There was no…Official notice →
  4. US
    Drugs & medicalClass III

    S & MOHR AMERICA, S.A. DE C.V. recalls abm Respiratory Care Standart Breating Circuit with Face Mask - Toddler REF BC21087 abm Respiratory Care Standart Breathing…

    Incorrect expiration date on product label. The manufacturing date was listed as the expiration date. Correct expiration date is "22.07.2027"

    FDA · DeviceS & MOHR AMERICA, S.A. DE C.V.U.S. Nationwide distribution in the state of MN.Official notice →
  5. US
    Drugs & medicalClass II

    Access Vascular, Inc recalls HydroMID 4Fr Single Lumen Maximal Barrier Kit-indicated for short term access (<30days) to the peripheral venous access system for…

    Label with the incorrect component listed on the inner kit Tyvek header bag

    FDA · DeviceAccess Vascular, IncUS Nationwide distribution in the states of CT, FL, NE, TN, TX, WA.Official notice →
  6. US
    Drugs & medicalClass II

    Access Vascular, Inc recalls HydroMID 4Fr Single Lumen Maximal Barrier Kit- Indicated for short term access (<30days) to the peripheral venous access system for…

    Product Mislabeled on the outer bag and inner kit Tyvek header bag

    FDA · DeviceAccess Vascular, IncUS Nationwide distribution in the states of FL, TX, WA.Official notice →
  7. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc. recalls Atellica IM Folate 140 test kit- For in vitro diagnostic use in the quantitative determination of folate in human serum and…

    Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.

    FDA · DeviceSiemens Healthcare Diagnostics, Inc.Nationwide Foreign: Albania¿¿¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿ Bahrain¿¿¿¿¿¿¿¿…Official notice →