Drugs & medicalClass II

MAQUET GMBH recalls Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All Software…

Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from…

Read the official notice →
Date
2025-09-03
Country
United States
Brand
Maquet
Product
Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All Software…
Classification
Class II
Units
Worldwide - US Nationwide distribution in the states of KY, NY, TX and the…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Maquet recalls

  1. 2026-09-30
  2. US
    Drugs & medicalClass II

    Maquet SAS recalls PowerLED 300 Concealed Lighting System (PWD30+SF W K3 SPRING ARM A2 FC3 50-100nm) Model Numbers: 1. ARD568474114A, 2. ARD567501286;

    The potential for the device to detach at the junction between the spring arm and suspension arm and fall causing potential injury.

    FDA · DeviceMaquet SASUS Nationwide distribution.Official notice →
All 2 Maquet recalls →

Other drugs & medical recalls in the United States

  1. 2025-09-03
  2. US
    Drugs & medicalClass II

    VANTIVE US HEALTHCARE LLC recalls PrisMax V2-US, Hemodialysis Delivery System, Product Codes: 955626, 955558, 955627, 955701

    Low deaeration chamber level at post priming tests may lead to PrisMax System alarm T2309: Air Detected in Prime resulting in multiple additional priming cycles

    FDA · DeviceVANTIVE US HEALTHCARE LLCWorldwideOfficial notice →
  3. US
    Drugs & medicalClass II

    Onkos Surgical, Inc. recalls ELEOS SEGMENTAL LIMB SALVAGE SYSTEM, ELEOS STEM EXTENSION, STRAIGHT, SPLINED, SLOTTED. Canal filling stem implants.

    Unidentified substance of unknown impact was identified on devices during a retrospective evaluation.

    FDA · DeviceOnkos Surgical, Inc.US Nationwide distribution in the states of AK, AZ, CA, FL, GA, IL, IN, KS,…Official notice →
  4. US
    Drugs & medicalClass II

    Dentsply IH, Inc. recalls Brand Name: Atlantis IO FLO Product Name: Atlantis IO FLO-S Kit Model/Catalog Number: A04S / 6802034 Product Description: The Atlantis…

    The defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage approximately 2 rotations as opposed to…

    FDA · DeviceDentsply IH, Inc.Worldwide - US Nationwide distribution in the states of IL, LA, MA, MN, MO,…Official notice →
  5. US
    Drugs & medicalClass II

    THOR Photomedicine Ltd recalls Brand Name: THOR Product Name: NovoTHOR Gen 3.0 Reg PRX Model/Catalog Number: S2186 Software Version: N/A Product Description: The…

    The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.

    FDA · DeviceTHOR Photomedicine LtdWorldwide distribution: US (nationwide): AK, AZ, CA, CO, CT, FL, GA, HI, ID,…Official notice →
  6. US
    Drugs & medicalClass II

    Baxter Healthcare Corporation recalls Spectrum IQ Infusion Pump, Product Code 3570009

    Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having…

    FDA · DeviceBaxter Healthcare CorporationUS distribution to states of: AZ, FLOfficial notice →
  7. US
    Drugs & medicalClass III

    Philips Ultrasound, Inc recalls TRANSDUCER L12-5 Transducer Probe

    To provide clarification and labelling to define the useful life of ultrasound transducers in the field.

    FDA · DevicePhilips Ultrasound, IncUS Nationwide distribution.Official notice →
See all drugs & medical recalls →