Dentsply IH, Inc. recalls Brand Name: Atlantis IO FLO Product Name: Atlantis IO FLO-S Kit Model/Catalog Number: A04S / 6802034 Product Description: The Atlantis…
The defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage approximately 2 rotations as opposed to…
Read the official notice →- Date
- 2025-09-03
- Country
- United States
- Brand
- Dentsply IH, Inc.
- Product
- Brand Name: Atlantis IO FLO Product Name: Atlantis IO FLO-S Kit Model/Catalog Number: A04S / 6802034 Product Description: The Atlantis…
- Classification
- Class II
- Units
- Worldwide - US Nationwide distribution in the states of IL, LA, MA, MN, MO,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
Other drugs & medical recalls in the United States
- 2025-09-03
US
VANTIVE US HEALTHCARE LLC recalls PrisMax V3 control Unit-US, Hemodialysis Delivery System, Product Code 955724
Unexpected PrisMax System alarm T0830: Blood Leak Detected, when no blood leak is present, or difficulty normalizing the Blood Leak Detector (BLD) leading to PrisMax System alarms T1313: BLD Normalize Failed, T0853:…
US
VANTIVE US HEALTHCARE LLC recalls PrisMax V2-US, Hemodialysis Delivery System, Product Codes: 955626, 955558, 955627, 955701
Low deaeration chamber level at post priming tests may lead to PrisMax System alarm T2309: Air Detected in Prime resulting in multiple additional priming cycles
US
Onkos Surgical, Inc. recalls ELEOS SEGMENTAL LIMB SALVAGE SYSTEM, ELEOS STEM EXTENSION, STRAIGHT, SPLINED, SLOTTED. Canal filling stem implants.
Unidentified substance of unknown impact was identified on devices during a retrospective evaluation.
FDA · DeviceOnkos Surgical, Inc.US Nationwide distribution in the states of AK, AZ, CA, FL, GA, IL, IN, KS,…Official notice →US
MAQUET GMBH recalls Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All Software…
Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from…
FDA · DeviceMAQUET GMBHWorldwide - US Nationwide distribution in the states of KY, NY, TX and the…Official notice →US
THOR Photomedicine Ltd recalls Brand Name: THOR Product Name: NovoTHOR Gen 3.0 Reg PRX Model/Catalog Number: S2186 Software Version: N/A Product Description: The…
The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.
FDA · DeviceTHOR Photomedicine LtdWorldwide distribution: US (nationwide): AK, AZ, CA, CO, CT, FL, GA, HI, ID,…Official notice →US
Baxter Healthcare Corporation recalls Spectrum IQ Infusion Pump, Product Code 3570009
Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having…