Isopure Corp recalls Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO)
Isopure received a report of a hazardous incident involving a device which had experienced an electrical short that resulted in a thermal event, damaging some of the device components. The firm is alerting customers to…
Read the official notice →- Date
- 2022-07-20
- Country
- United States
- Brand
- Isopure
- Product
- Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO)
- Classification
- Class II
- Units
- Domestic distribution to AL AZ CO DE FL GA IN KS KY LA MA MD ME MN MO NC NY OK…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Isopure recalls
- 2022-09-21
US
Isopure Corp recalls Acidified Concentrate Distribution and Storage-The Isopure ACDS (Acidified Concentrate Distribution and Storage) with remote fill is…
Acidified Concentrate Distribution System potential electrical safety hazard may result in electric shock, arching
Other drugs & medical recalls in the United States
- 2022-07-20
US
Cordis US Corp recalls PALMAZ GENESIS Transhepatic Biliary Stent on OPTA PRO .035" Delivery System, Catalog numbers PG2990BPS, PG2990BPX, PG3990BPS, PG3990BPX.
Potential for stent dislodgement and associated failures related to two specific sizes of the device.
FDA · DeviceCordis US CorpUS Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE,…Official notice →US
Cook Incorporated recalls Pressure Monitoring Set, Single-lumen uncoated central venous catheter. RPN C-PMS-250, GPN G02838
The products may contain excess coating material on the outside and/or interior of the needle component. Potential adverse events that may occur if an affected product is used include a delay in the procedure, a…
US
US
Medtronic MiniMed recalls MiniMed 640G Insulin Pump (O.U.S. Version)
Due to battery cap deterioration, this may potentially result in an incomplete battery circuit (similar to when battery cap is not attached) and a loss of power/functionality of the insulin infusion pumps.
FDA · DeviceMedtronic MiniMedU.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY,…Official notice →US
Spectranetics Corporation recalls Philips Laser System, REF LAS-100, Intermittent Operation, Duty cycle dependent upon device in use 100 - 240 VAC, 50/60HZ, 16 AM
The LAS-100 Laser system may detect an inoperable hardware component during power up, which results in an error code and the system not being operable until code is cleared.
FDA · DeviceSpectranetics CorporationUS: AR AZ CA CO CT FL GA HI IA IL IN KS LA MA MD ME MI MN MO NC NJ NY OK PA TN…Official notice →US
Zeltiq Aesthetics, Inc recalls CoolMax ZELTIQ Vacuum Applicator REF BRZ-AP1-080-000
Executing a voluntary discontinuation and removal of parallel plate applicators due to an increase complaint rate for adverse event Paradoxical Hyperplasia (HP) during the 2019 to 2021 timeframe.
FDA · DeviceZeltiq Aesthetics, IncU.S. AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY,…Official notice →