Intuitive Surgical, Inc. recalls VESSEL SEALER EXTEND, Single-Use Instrument EndoWrist Instruments 8mm, da Vinci X' da Vinci Xi', REF 480422
Placing excessive tissue in the instrument jaws of the devices prior to sealing and transection can result in an insufficient seal.
Read the official notice →- Date
- 2022-06-08
- Country
- United States
- Brand
- Intuitive Surgical
- Product
- VESSEL SEALER EXTEND, Single-Use Instrument EndoWrist Instruments 8mm, da Vinci X' da Vinci Xi', REF 480422
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide and the countries/regions of Argentina,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Intuitive Surgical recalls
- 2022-11-02
US
Intuitive Surgical, Inc. recalls da Vinci X (IS4200) and Xi (IS4000) systems, which consists of finished devices Surgeon Console (SSC), Patient Cart (PSCART), Vision…
Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bipolar not connected to generator 3)…
FDA · DeviceIntuitive Surgical, Inc.US: MS, PA, WI, KY, TN, MI, FL, GA, NC, NV, LA, OH, IL, MO, IN, CA, AK, MD,…Official notice →- 2022-08-10
US
Intuitive Surgical, Inc. recalls The E-100 Generator (PN 374848-09) is a bipolar electrosurgical unit (ESU) designed to be used with the da Vinci Xi and X systems. The…
Due to preventative maintenance (PM) not being performed on E-100 generators which is a bipolar electrosugical unit designed to be used with the da Vinci Xi and X systems.
FDA · DeviceIntuitive Surgical, Inc.Worldwide distribution - U.S. Nationwide distribution in the states of AL, AK,…Official notice →- 2022-08-03
US
Intuitive Surgical, Inc. recalls Da Vinci S/Si Large Clip Applier, Model 420230 Da Vinci S/Si Medium-Large Clip Applier, Model 420327 Da Vinci S/Si Small Clip Applier,…
Due to the potential for unexpected motion caused by the partial disengagement of the instrument from the system.
FDA · DeviceIntuitive Surgical, Inc.Worldwide distribution - U.S. Nationwide distribution in the states of AL, AK,…Official notice →- 2022-05-04
US
Intuitive Surgical, Inc. recalls da Vinici X da Vinci Xi Sureform 45 REF 480445 da Vinci X da Vinci Xi Sureform 45 Curved-Tip REF 480545 da Vinici X da Vinci Xi Sureform…
As a result of complaint data review, it identified tissue pushout events occurring during firing of the SureForm stapler when tissue is pushed distally within the jaws, rather than being held in place and transected.…
FDA · DeviceIntuitive Surgical, Inc.Worldwide distribution - U.S. Nationwide distribution in the states of AL, AK,…Official notice →- 2022-03-09
US
Intuitive Surgical, Inc. recalls da Vinci Xi and X 12-8 mm Conductive Cannula Reducer, REF: 470381-11
Conductive Cannula Reducer metal tip may get dislodged from the plastic shaft. If unnoticed, additional surgery may be required to remove the metal tip. If the procedure proceeds without the metal tip, there is…
FDA · DeviceIntuitive Surgical, Inc.Worldwide distribution - US Nationwide distribution in the states of…Official notice →- 2022-01-05
US
Intuitive Surgical, Inc. recalls da Vinci SP Surgical systems with the following Model Name/ Model Number / UDI: ASSY,PSS,SP1098,P4/ 380601-44/ 00886874114605…
Issue was identified during internal engineering evaluation. Use of the system with the affected software version may experience either an inability to deliver energy or an inadvertent delivery of energy. Use may…
FDA · DeviceIntuitive Surgical, Inc.US Nationwide distribution in the states of Arizona, California, Connecticut,…Official notice →
Other drugs & medical recalls in the United States
- 2022-06-08
US
Becton, Dickinson and Company, BD Biosciences recalls BD Anti-Kappa PE (Cat. No. 346587) is an analyte specific reagent for the Anti-Kappa antibody, which specifically…
Anti-Kappa APC & Anti-Kappa PE analyte reagents may exhibit uncharacteristic flow profiles.
FDA · DeviceBecton, Dickinson and Company, BD BiosciencesWorldwide distribution - US Nationwide distribution in the states of AL, CA,…Official notice →US
Abbott recalls FIRMap" Catheter, 60mm Basket
Incorrect product labeling.
US
XVIVO PERFUSION AB recalls XVIVO Organ Chamber REF 19020
The weld seal of the primary pouch was incomplete, and the sterility of the product cannot be ensured/guaranteed.
FDA · DeviceXVIVO PERFUSION ABU.S.: CA, FL, MA, NC, NY, OH, PA, and TX O.U.S.: Austria, Canada, France,…Official notice →US
Bio-Rad Laboratories, Inc. recalls Resin UPDATE CD-ROM, REF: 250-3020 inside VARIANTnbs Sickle Cell Program Reorder Pack, REF: 250-3000
The problem is with the Bio-Rad VARIANT nbs Sickle Cell Program Resin Update CD-ROM software; Model Number: 250-3020, included in VARIANT nbs Sickle Cell Progam Reorder Pack #250-3000. This CD-ROM software included in…
FDA · DeviceBio-Rad Laboratories, Inc.US: AZ, AR, GA, CO, TN, LA, VA, FL, KY, IN, KS, MD, MN, MO, NC, NJ, NY, OH,…Official notice →US
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls The C304-HIS device features a guide wire to access the vein, a valve to reduce blood loss during the…
The firm's internal processes identified that certain lots may have an improperly sealed barrier that could potentially compromise sterility.
FDA · DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Worldwide distribution - US Nationwide distribution in the states of AL, AR,…Official notice →US
Ion Beam Applications S.A. recalls Proteus235- Proton Therapy System: to produce and deliver a proton beam for the treatment of patients with localized tumors and other…
First layer of Uniform Scanning treatment fields is sometimes irradiated with incorrect Scanning Magnet setpoints resulting in the field size may be smaller than expected and the irradiated area may receive more dose…
FDA · DeviceIon Beam Applications S.A.Worldwide distribution - US Nationwide distribution in the states of IL, VA,…Official notice →