Intuitive Surgical, Inc. recalls da Vinci Xi and X 12-8 mm Conductive Cannula Reducer, REF: 470381-11
Conductive Cannula Reducer metal tip may get dislodged from the plastic shaft. If unnoticed, additional surgery may be required to remove the metal tip. If the procedure proceeds without the metal tip, there is…
Read the official notice →- Date
- 2022-03-09
- Country
- United States
- Brand
- Intuitive Surgical
- Product
- da Vinci Xi and X 12-8 mm Conductive Cannula Reducer, REF: 470381-11
- Classification
- Class II
- Units
- Worldwide distribution - US Nationwide distribution in the states of…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Intuitive Surgical recalls
- 2022-08-03
US
Intuitive Surgical, Inc. recalls Da Vinci S/Si Large Clip Applier, Model 420230 Da Vinci S/Si Medium-Large Clip Applier, Model 420327 Da Vinci S/Si Small Clip Applier,…
Due to the potential for unexpected motion caused by the partial disengagement of the instrument from the system.
FDA · DeviceIntuitive Surgical, Inc.Worldwide distribution - U.S. Nationwide distribution in the states of AL, AK,…Official notice →- 2022-06-08
US
Intuitive Surgical, Inc. recalls VESSEL SEALER EXTEND, Single-Use Instrument EndoWrist Instruments 8mm, da Vinci X' da Vinci Xi', REF 480422
Placing excessive tissue in the instrument jaws of the devices prior to sealing and transection can result in an insufficient seal.
FDA · DeviceIntuitive Surgical, Inc.Worldwide distribution - US Nationwide and the countries/regions of Argentina,…Official notice →- 2022-05-04
US
Intuitive Surgical, Inc. recalls da Vinici X da Vinci Xi Sureform 45 REF 480445 da Vinci X da Vinci Xi Sureform 45 Curved-Tip REF 480545 da Vinici X da Vinci Xi Sureform…
As a result of complaint data review, it identified tissue pushout events occurring during firing of the SureForm stapler when tissue is pushed distally within the jaws, rather than being held in place and transected.…
FDA · DeviceIntuitive Surgical, Inc.Worldwide distribution - U.S. Nationwide distribution in the states of AL, AK,…Official notice →- 2022-01-05
US
Intuitive Surgical, Inc. recalls da Vinci SP Surgical systems with the following Model Name/ Model Number / UDI: ASSY,PSS,SP1098,P4/ 380601-44/ 00886874114605…
Issue was identified during internal engineering evaluation. Use of the system with the affected software version may experience either an inability to deliver energy or an inadvertent delivery of energy. Use may…
FDA · DeviceIntuitive Surgical, Inc.US Nationwide distribution in the states of Arizona, California, Connecticut,…Official notice →- 2021-12-01
US
Intuitive Surgical, Inc. recalls The da Vinci Xi and X surgical systems (IS4000 and IS4200) each consist of three main components: the Surgeon Console (SC), the Vision…
Incorrect expiration date on its package labeling (shipper box and inner carton label).
FDA · DeviceIntuitive Surgical, Inc.Worldwide distribution - U.S. Nationwide distribution in the states of AL, CA,…Official notice →- 2021-10-20
US
Intuitive Surgical, Inc. recalls Ion Flexison Biopsy Needle, 19G, Extension Range : 0-3 cm, Ref: 490104-09 and 49004-10, 5-pack, Sterile EO, Rx only, UDI:00886874115978
Potential for the sheath tip on the biopsy needle to separate from the sheath shaft.
FDA · DeviceIntuitive Surgical, Inc.US: AZ, AR, CA, CO, DC, FL, GA, IL, IN, IO, MD, MA, MI, MN, MO, NV, NJ, NY,…Official notice →
Other drugs & medical recalls in the United States
- 2022-03-09
US
Brasseler USA, Medical L.L.C. recalls Brasseler USA, K-WIRE, Double Trocar Smooth, REF KM172-26-35S
One lot of product was distributed in unsealed packaging
FDA · DeviceBrasseler USA, Medical L.L.C.US Nationwide distribution in the states of OH, MI, AZ, FL, CA, TN, TX, SC, PA…Official notice →US
K2M, Inc recalls Everest MI XT Inner Dilator, Catalog Number 5101-90167
Stryker received two (2) complaints for units from specific lots of Inner and Outer Dilators not fitting together properly during a surgical procedure. Some dilators from these lots and from other lots were…
US
Microbiologics Inc recalls QC Sets and Panels: KWIK-STIK Cepheid Smart GBS QC Set (Live Culture). Sets include: 1. 0864E4 Streptococcus agalactiae (Group B, lb)…
The Lactobacillus acidophilus negative control (item 0243 single KWIK-STIK) contained in the 8164 Cepheid Xpert GBS QC Set and 8165 Cepheid Smart GBS QC Set was found to test positive for Group B Streptococcus (GBS) in…
FDA · DeviceMicrobiologics IncUS Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, IA, IL,…Official notice →US
Microbiologics Inc recalls Microbiologics QC Sets and Panels: Helix Elite. Extended Enteric Bacterial Verification Panel (Inactivated Pellet) includes 21 pellets…
Real-time shelf life testing failed at 24 months
FDA · DeviceMicrobiologics IncWorldwide distribution - US Nationwide distribution in the states of AK, CA,…Official notice →US
MEDLINE INDUSTRIES, LP - Northfield recalls MEDLINE Micro-Kill Bleach Germicidal Bleach Wipes
Micro-Kill Bleach Wipes contain out of specification (low) levels of Sodium Hypochlorite.
FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldWorldwide distribution - US Nationwide and the countries of Panama and Bermuda.Official notice →US
ROi CPS LLC recalls Regard, TOTAL HIP - SPRINGFIELD, Item No. 8800297010, Case Quantity: 2, Sterile
Products are mislabeled with the incorrect product identification on individual kit labeling.