GE Healthcare, LLC recalls SIGNA Pioneer nuclear magnetic resonance imaging system
Under certain conditions, some slices may be missing which can lead to a gap of anatomy in the 3D Volume images.
Read the official notice →- Date
- 2021-10-27
- Country
- United States
- Brand
- GE Healthcare
- Product
- SIGNA Pioneer nuclear magnetic resonance imaging system
- Classification
- Class II
- Units
- Distribution in United States to AZ, CA, FL, GA, IL, LA, MA, MD, MN, NC, NH,…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More GE Healthcare recalls
- 2022-02-02
US
GE Healthcare, LLC recalls GE Vscan Extend, being sold as "Vscan Extend Sector DICOM"
The firm identified that the Vscan Extend device will not power on because of depleted battery, due to the device not being turned off prior to battery removal or when being connected to a computer without AC power.
FDA · DeviceGE Healthcare, LLCDomestic: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY,…Official notice →- 2021-12-08
US
GE Healthcare, LLC recalls GE Centricity Universal Viewer Zero Footprint Client
There is a potential to display incomplete patient imaging study.
FDA · DeviceGE Healthcare, LLCWorldwide distribution - US Nationwide distribution in the states of AR, AZ,…Official notice →- 2021-12-01
US
GE Healthcare, LLC recalls Computed Tomography Systems Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition
The accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts. In this situation, the…
- 2021-10-13
US
GE Healthcare, LLC recalls Centricity Universal Viewer 7.0, Model 5829628-0XX. Radiological image processing system
A software anomaly exists in the Centricity Universal Viewer study management feature in which study changes are not propagated to either the Centricity Enterprise Archive (EA) or another Vendor Neutral Archive (VNA).
FDA · DeviceGE Healthcare, LLCAR, IL, ME, MS, NE, NJ, NY, and TN Belgium, Canada, Costa Rica, France,…Official notice →- 2021-10-06
US
GE Healthcare, LLC recalls Flow sensors that may be installed in the following GEHC anesthesia machines: (1) Aestiva MRI, GTIN 0080682102339; (2) Aestiva 7900; (3)…
The flow sensors built prior to June 2021 could have damaged tubes with small punctures or cuts that could cause leaks resulting in incorrect anesthesia machine tidal volumes potentially leading to over-delivery of…
FDA · DeviceGE Healthcare, LLCDistribution was nationwide, including PR. Foreign distribution was made to…Official notice →US
GE Healthcare, LLC recalls AMX Navigate Mobile X-Ray System
An unexpected column motion while attempting to park the column on the AMX Navigate mobile X-ray system.
Other drugs & medical recalls in the United States
- 2021-10-27
US
ADRIA SRL recalls Herniatome, 17G x 15cm, Lumbar Type, Model Number 5091745
The recalling firm was notified by the contract sterilizer that the process parameters of the sterilization cycles were not followed.
FDA · DeviceADRIA SRLThe products were distributed to the following US states: NJ and NY.Official notice →US
Smith & Nephew, Inc. recalls 6.5MMX165MM CANNULATED SCREW
The product label incorrectly indicated that the screws are fully threaded, but the screws contained in the package were partially threaded.
FDA · DeviceSmith & Nephew, Inc.Worldwide distribution - US Nationwide distribution in the states of OK, TN…Official notice →US
Canon Medical System, USA, INC. recalls Canon PET-CT SCANNER, Cartesion Prime, MODEL PCD-1000A
A software problem has been identified which could result in the diagnostic imaging system not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the…
FDA · DeviceCanon Medical System, USA, INC.US Nationwide distribution in the states of NV, AR MN.Official notice →US
Stradis Medical, LLC dba Stradis Healthcare recalls Foundation Kit
Firm unable to confirm that the tenaculum devices were produced in accordance with required design and manufacturing standards.
FDA · DeviceStradis Medical, LLC dba Stradis HealthcareDistributed nationwide to VA, SC, CA, PA, MA, FL, TN, NJ, GA, AZ, IL, OH, OK,…Official notice →US
Siemens Healthcare Diagnostics, Inc recalls Atellica IM 1600 Analyzer, SMN 11066000
Ancillary reagent packs used on Atellica IM 1300 and Atellica IM 1600 analyzers may overflow when pierced while on an analyzer located at an altitude of greater than 350m (1148ft) above sea level. As a result,…
FDA · DeviceSiemens Healthcare Diagnostics, IncDomestic distribution to AZ CA CO GA ID KS MO MT NC NE NM NV OH OK OR PA TN TX…Official notice →US
Philips Ultrasound Inc recalls EPIQ CVxi Diagnostic Ultrasound Systems, Product Code: 795232
Ultrasound system software issue can cause an EchoNavigator error notification, leading to an unresponsive system, necessitating system restart, which could lead to therapy/treatment delay and/or unnecessary…
FDA · DevicePhilips Ultrasound IncUS: MI, AZ, IA, TN, OH, CO, WI, NY, NJ, NC, LA, IL, CA, KY, IN, MD, TX, PA.…Official notice →