Drugs & medicalClass II

Canon Medical System, USA, INC. recalls Canon PET-CT SCANNER, Cartesion Prime, MODEL PCD-1000A

A software problem has been identified which could result in the diagnostic imaging system not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the…

Read the official notice →
Date
2021-10-27
Country
United States
Brand
Canon Medical System
Product
Canon PET-CT SCANNER, Cartesion Prime, MODEL PCD-1000A
Classification
Class II
Units
US Nationwide distribution in the states of NV, AR MN.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Canon Medical System recalls

  1. 2024-02-28
  2. US
    Drugs & medicalClass II

    Canon Medical System, USA, INC. recalls The INFX-8000F is designed to take advantage of the latest technological innovations to reduce dose and to save time. A…

    VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may…

    FDA · DeviceCanon Medical System, USA, INC.US: CA, LAOfficial notice →
  3. 2023-08-16
  4. US
    Drugs & medicalClass II

    Canon Medical System, USA, INC. recalls INFX-8000C XGCP-930AA TABLE SIDE CONSOLE, INFX-8000C XGCP-930BA TABLE SIDE CONSOLE

    The C-arm, used to set the position and angle of the x-ray exposure field, may not function due to the tableside console PWB, which needs replacement. There may be a delay in the procedure, but procedures can continue…

    FDA · DeviceCanon Medical System, USA, INC.US Nationwide Distribution: NY, CO, IL, CA, MD, MN, WAOfficial notice →
  5. 2022-11-16
  6. US
    Drugs & medicalClass II

    Canon Medical System, USA, INC. recalls Canon Aplio ultrasound system, Model: Aplio i900 TUS-AI900, Aplio i800 TUS-AI800, Aplio i700 TUS-AI700, Aplio a550 CUS-AA550,…

    After completion, ultrasound system's Stress Echo may not terminate if user does not manually terminate the protocol. During next patient exam, an abnormal confirmation message will be displayed "Same view was saved in…

    FDA · DeviceCanon Medical System, USA, INC.US: MT, MO, NC, SC, MN, WA, KY, TN, AR, IL, IA, MI, WV, KS, CA, HA, CO, LA,…Official notice →
  7. 2022-09-07
  8. US
    Drugs & medicalClass II

    Canon Medical System, USA, INC. recalls INFX-8000V. For radiographic and fluoroscopic studies and intervention.

    The fluoroscopic dose rate might exceed the conforming value during biplane fluoroscopy.

    FDA · DeviceCanon Medical System, USA, INC.US Nationwide DistributionOfficial notice →
  9. 2022-07-06
  10. US
    Drugs & medicalClass II

    Canon Medical System, USA, INC. recalls System: Alphenix 4D CT in combination with CAS-930A (C arm Support) Model Number: INFX-8000C/BF, INFX-8000C/BH, INFX-8000C/SW

    CT operation may be restricted by an interlock which is a result of an system error and the CT system may stop operating properly after an attempt to cancel the error message has been initiated.

    FDA · DeviceCanon Medical System, USA, INC.U.S.: CA, CO, FL, NY, and WY O.U.S.: NoneOfficial notice →
  11. 2022-01-19
  12. US
    Drugs & medicalClass II

    Canon Medical System, USA, INC. recalls Flat Panel Detector, Model Number: TFP-800A, used with the Alphenix Systems, Item Numbers: INFX-8000V, INFX-8000C, INFX-8000F,…

    Abnormal images such as horizontal stripes may be displayed and recovery may not not be possible by restarting the system. The worst case scenario is an operator who is working without a workable image may remove a…

    FDA · DeviceCanon Medical System, USA, INC.US Nationwide distribution in the states of DE, FL, LA, MT, NJ, OH, and WV.Official notice →
All 16 Canon Medical System recalls →

Other drugs & medical recalls in the United States

  1. 2021-10-27
  2. US
    Drugs & medicalClass II

    Medical Action Industries, Inc. 306 recalls Kit: Central Line Dres Change 20/Cs

    The ChloraPrep 3ml kit component has a shorter expiry date than the labeled kit expiry date.

    FDA · DeviceMedical Action Industries, Inc. 306Distribution to a single US distributor in MichiganOfficial notice →
  3. US
    Drugs & medicalClass II

    ADRIA SRL recalls Herniatome, 17G x 15cm, Lumbar Type, Model Number 5091745

    The recalling firm was notified by the contract sterilizer that the process parameters of the sterilization cycles were not followed.

    FDA · DeviceADRIA SRLThe products were distributed to the following US states: NJ and NY.Official notice →
  4. US
    Drugs & medicalClass II

    Smith & Nephew, Inc. recalls 6.5MMX165MM CANNULATED SCREW

    The product label incorrectly indicated that the screws are fully threaded, but the screws contained in the package were partially threaded.

    FDA · DeviceSmith & Nephew, Inc.Worldwide distribution - US Nationwide distribution in the states of OK, TN…Official notice →
  5. US
    Drugs & medicalClass II

    GE Healthcare, LLC recalls SIGNA Pioneer nuclear magnetic resonance imaging system

    Under certain conditions, some slices may be missing which can lead to a gap of anatomy in the 3D Volume images.

    FDA · DeviceGE Healthcare, LLCDistribution in United States to AZ, CA, FL, GA, IL, LA, MA, MD, MN, NC, NH,…Official notice →
  6. US
    Drugs & medicalClass II

    Stradis Medical, LLC dba Stradis Healthcare recalls Foundation Kit

    Firm unable to confirm that the tenaculum devices were produced in accordance with required design and manufacturing standards.

    FDA · DeviceStradis Medical, LLC dba Stradis HealthcareDistributed nationwide to VA, SC, CA, PA, MA, FL, TN, NJ, GA, AZ, IL, OH, OK,…Official notice →
  7. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics, Inc recalls Atellica IM 1600 Analyzer, SMN 11066000

    Ancillary reagent packs used on Atellica IM 1300 and Atellica IM 1600 analyzers may overflow when pierced while on an analyzer located at an altitude of greater than 350m (1148ft) above sea level. As a result,…

    FDA · DeviceSiemens Healthcare Diagnostics, IncDomestic distribution to AZ CA CO GA ID KS MO MT NC NE NM NV OH OK OR PA TN TX…Official notice →