Drugs & medicalClass II

FUJIFILM Healthcare Americas Corporation recalls FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)

It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even…

Read the official notice →
Date
2026-02-11
Country
United States
Brand
FUJIFILM Healthcare Americas
Product
FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)
Classification
Class II
Units
US Nationwide distribution in the states of AK, CA, CT, FL, IL, NH, OH, WI.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More FUJIFILM Healthcare Americas recalls

  1. 2026-09-02
  2. US
    Drugs & medicalClass II

    FUJIFILM Healthcare Americas Corporation recalls FUJIFILM Synapse PACS software v7.4.310; radiological image processing system

    Software issue may cause the ruler measurement to be inaccurate.

    FDA · DeviceFUJIFILM Healthcare Americas CorporationWorldwide - US Nationwide distribution in the states of NY, VA, KS, OR, OH,…Official notice →
  3. 2025-10-15
  4. US
    Drugs & medicalClass II

    FUJIFILM Healthcare Americas Corporation recalls FDR Visionary Suite, CH-200 Model/Version Numbers: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33

    It was found that the shaft for mounting the X-ray tube unit on the support may break. In addition, it was found that the protective parts might not work effectively only when the shaft was broken while being operated…

    FDA · DeviceFUJIFILM Healthcare Americas CorporationDomestic: AK, CA, FL, IL, MA, NH, OH, WI.Official notice →
  5. 2025-08-06
  6. US
    Drugs & medicalClass II

    FUJIFILM Healthcare Americas Corporation recalls ASPRIRE Cristalle Mammography System (cleared under K212873) installed with the ASPIRE Cristalle Digital Breast…

    Devices had an unapproved slabbing software function enabled for use.

    FDA · DeviceFUJIFILM Healthcare Americas CorporationWorldwide distribution - US Nationwide and the countries of Canada, Chile,…Official notice →
  7. 2025-03-26
  8. US
    Drugs & medicalClass II

    FUJIFILM Healthcare Americas Corporation recalls Synapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200.

    The incorrect computed patient age is showing in VX for patients less than 3 months old.

    FDA · DeviceFUJIFILM Healthcare Americas CorporationDomestic: AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, KY, LA, MA, MN, MO, MS,…Official notice →
  9. 2024-10-02
  10. US
    Drugs & medicalClass II

    FUJIFILM Healthcare Americas Corporation recalls The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID)…

    When the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the…

    FDA · DeviceFUJIFILM Healthcare Americas CorporationUS Nationwide and WorldwideOfficial notice →
  11. 2024-06-05
  12. US
    Drugs & medicalClass II

    FUJIFILM Healthcare Americas Corporation recalls FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column,…

    Bolts on the CH-200 tube may rupture due to fatigue failure, causing the tube support to fall, could result in patient injury

    FDA · DeviceFUJIFILM Healthcare Americas CorporationUS Nationwide distribution in the states of AK, CA, FL, IL, NE, NH, OH, TN, WI.Official notice →
All 14 FUJIFILM Healthcare Americas recalls →

Other drugs & medical recalls in the United States

  1. 2026-02-11
  2. US
    Drugs & medicalClass II

    ETAC A/S recalls Brand Name: Molift Product Name: Molift 4-point sling bars (Accessory to mobile hoists) Model/Catalog Number: 1430116,1430117,1430115…

    A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as…

    FDA · DeviceETAC A/SWorldwide - US Nationwide distribution in the state of PA and the countries of…Official notice →
  3. US
    Drugs & medicalClass II

    Fujirebio Diagnostics, Inc. recalls Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Product Description:…

    Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging,…

    FDA · DeviceFujirebio Diagnostics, Inc.US Nationwide distribution in the states of Arizona, California, Florida,…Official notice →
  4. US
    Drugs & medicalClass II

    Maquet Cardiopulmonary Gmbh recalls Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;

    Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Venous Bubble Sensor. Excessive bending of the connecting cable can lead to full damage of the…

    FDA · DeviceMaquet Cardiopulmonary GmbhWorldwide Distribution: US (Nationwide) and OUS (International) to countries…Official notice →
  5. US
    Drugs & medicalClass II

    Wilson-Cook Medical Inc. recalls Product Name: Nasal Jejunal Feeding Tube REF: NJFT-10

    Nasal feeding tube packaged without the nasal transfer tube component.

    FDA · DeviceWilson-Cook Medical Inc.International distribution to the country of Australia.Official notice →
  6. US
    Drugs & medicalClass II

    Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These instruments…

    Devices which did not undergo thermoforming could deform and lose performance.

    FDA · DeviceOlympus Corporation of the AmericasWorldwide - US Nationwide distribution including in the states of AL, AR, AZ,…Official notice →
  7. US
    Drugs & medicalClass II

    Changchun Wancheng Bio-Electron Co., Ltd. recalls Male Fertility Sperm Test for Home Use (Cassette)

    Test strips and other tests including for near-patient testing, that require a 510(k) were distributed before they could be legally marketed to U.S. customers; devices may cause false or inaccurate diagnostic result by…

    FDA · DeviceChangchun Wancheng Bio-Electron Co., Ltd.US Nationwide distribution in the states of TX, GA, CA.Official notice →
See all drugs & medical recalls →