FUJIFILM Healthcare Americas Corporation recalls FDR Visionary Suite, CH-200 Model/Version Numbers: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33
It was found that the shaft for mounting the X-ray tube unit on the support may break. In addition, it was found that the protective parts might not work effectively only when the shaft was broken while being operated…
Read the official notice →- Date
- 2025-10-15
- Country
- United States
- Brand
- FUJIFILM Healthcare Americas
- Product
- FDR Visionary Suite, CH-200 Model/Version Numbers: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33
- Classification
- Class II
- Units
- Domestic: AK, CA, FL, IL, MA, NH, OH, WI.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More FUJIFILM Healthcare Americas recalls
- 2026-09-02
US
FUJIFILM Healthcare Americas Corporation recalls FUJIFILM Synapse PACS software v7.4.310; radiological image processing system
Software issue may cause the ruler measurement to be inaccurate.
FDA · DeviceFUJIFILM Healthcare Americas CorporationWorldwide - US Nationwide distribution in the states of NY, VA, KS, OR, OH,…Official notice →- 2026-02-11
US
FUJIFILM Healthcare Americas Corporation recalls FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)
It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even…
FDA · DeviceFUJIFILM Healthcare Americas CorporationUS Nationwide distribution in the states of AK, CA, CT, FL, IL, NH, OH, WI.Official notice →- 2025-08-06
US
FUJIFILM Healthcare Americas Corporation recalls ASPRIRE Cristalle Mammography System (cleared under K212873) installed with the ASPIRE Cristalle Digital Breast…
Devices had an unapproved slabbing software function enabled for use.
FDA · DeviceFUJIFILM Healthcare Americas CorporationWorldwide distribution - US Nationwide and the countries of Canada, Chile,…Official notice →- 2025-03-26
US
FUJIFILM Healthcare Americas Corporation recalls Synapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200.
The incorrect computed patient age is showing in VX for patients less than 3 months old.
FDA · DeviceFUJIFILM Healthcare Americas CorporationDomestic: AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, KY, LA, MA, MN, MO, MS,…Official notice →- 2024-10-02
US
FUJIFILM Healthcare Americas Corporation recalls The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID)…
When the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the…
- 2024-06-05
US
FUJIFILM Healthcare Americas Corporation recalls FDR Visionary Suite - Intended to generate digital or conventional radiographic images of the skull, spinal column,…
Bolts on the CH-200 tube may rupture due to fatigue failure, causing the tube support to fall, could result in patient injury
FDA · DeviceFUJIFILM Healthcare Americas CorporationUS Nationwide distribution in the states of AK, CA, FL, IL, NE, NH, OH, TN, WI.Official notice →
Other drugs & medical recalls in the United States
- 2025-10-15
US
Beckman Coulter Inc. recalls SYNCHRON Systems Phosphorus (PHOSm) Reagent REF: 467868 For In Vitro Diagnostic Use, Rx Only.
Specific SYNCHRON Systems Modular Chemistry Phosphorus (PHOSm REF 467868) reagent lots may fail calibration before the lot expiration date which causes a delay in obtaining patient results.
FDA · DeviceBeckman Coulter Inc.US: AL KY LA MA MD MS NC NE NJ NY OK PA TX WA WV OUS: Canada Mexico SingaporeOfficial notice →US
Abiomed, Inc. recalls Automated Impella Controller (AIC), used for left heart support blood pump, labeled as the following with corresponding Product Codes: 1.…
Potential cybersecurity vulnerabilities related to the operating system in the Automated Impella Controller (AIC).
FDA · DeviceAbiomed, Inc.Nationwide distribution. International distribution to United Arab Emirates,…Official notice →US
GE Medical Systems, SCS recalls AW Server 2.0, AW Server 3.0, AW Server 3.1, AW Server 3.2
Firm has identified a security vulnerability in AW Server products. If exploited, a malicious actor could compromise the confidentiality, integrity, and availability of patient data.
FDA · DeviceGE Medical Systems, SCSDomestic US distribution nationwide. International distribution worldwide.Official notice →US
Fresenius Medical Care Holdings, Inc. recalls Optiflux¿ High Flux E-beam Dialyzer Model Numbers: (1) 0500316E OPTIFLUX F160NRe 12/CS 1.5SA EBEAM, (2) 0500318E OPTIFLUX…
To provide further information to the user to ensure proper use of the medical device. the caps had been changed from a threaded version to a press-on version whereby the user must press the caps firmly onto the…
US
CORNEAT VISION, LTD. recalls CorNeat EverPatch, REF: 1036295, Permanent Tissue-Integrating Surgical Matrix, 5.0 mm x 605 mm x 100 ¿m
Reviewed complaints have identified that conjunctival wound dehiscence may result in early exposure of the surgical patch.
FDA · DeviceCORNEAT VISION, LTD.Worldwide - US Nationwide distribution in the states of AZ, CA, CO, CT, DC,…Official notice →US
ImaCor Inc. recalls ClariTEE Miniaturized TEE probe. Model Number: CLT-010. Ultrasound miniaturized TEE probe for use with Mobile Ultrasound, ImaCor model ZHH.
Production assembly error resulting in the potential for reversed articulation, where moving the lever forward caused the probe tip to anteflex rather than retroflex, and vice versa.
FDA · DeviceImaCor Inc.US Nationwide distribution in the states of South Carolina, New Jersey,…Official notice →