Drugs & medicalClass II

Elekta, Inc. recalls ELEKTA LIMITED Elekta Infinity, REF XRT 0511, XRT 0521, XRT 0531, XRT 0541; Medical Charged-Particle Radiation Therapy System Accelerator,…

Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there…

Read the official notice →
Date
2023-12-13
Country
United States
Brand
Elekta
Product
ELEKTA LIMITED Elekta Infinity, REF XRT 0511, XRT 0521, XRT 0531, XRT 0541; Medical Charged-Particle Radiation Therapy System Accelerator,…
Classification
Class II
Units
worldwide
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Elekta recalls

  1. 2026-01-21
  2. US
    Drugs & medicalClass II

    Elekta, Inc. recalls MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)

    Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial…

    FDA · DeviceElekta, Inc.US Nationwide distribution in the state of TX.Official notice →
  3. 2025-11-12
  4. US
    Drugs & medicalClass II

    Elekta, Inc. recalls Brand Name: Leksell Stereotactic System Product Name: Leksell Vantage Stereotactic System Software Version: n/a Product Description: The…

    Due to increased risk that paint flakes from the engravings may separate from the device and there is the possibility that the flake may inadvertently enter the wound

    FDA · DeviceElekta, Inc.Worldwide - U.S. Nationwide distribution in the states of AR, CA, CO, FL, ID,…Official notice →
  5. 2025-02-19
  6. US
    Drugs & medicalClass II

    Elekta, Inc. recalls MOSAIQ Oncology Information System

    Image offsets calculated by oncology Information System software may be incorrect for Linear Accelerator (LINACs) not characterized at IEC61217 scale.

    FDA · DeviceElekta, Inc.US: PA, AL, FL, IA UT OUS: Austria Belgium Bolivia Bosnia and Herzegowina…Official notice →
  7. 2023-06-14
  8. US
    Drugs & medicalClass II

    Elekta, Inc. recalls Elekta Unity Image-Guided Radiation Therapy System containing an MR gradient coil.

    There is a low possibility that an electrical connector in the MR gradient coil will overheat on Elekta Unity systems.

    FDA · DeviceElekta, Inc.US Nationwide distribution in the states of FL, IA, NJ, NY, OK, PA, TN, WI,…Official notice →
  9. 2023-06-07
  10. US
    Drugs & medicalClass II

    Elekta, Inc. recalls Elekta Medical Linear Accelerator (ELMA) included in the below platforms: a. Elekta Infinity; b. Versa HD; c. Elekta Synergy

    A component within a limited number of Elekta Linacs has been realized to be non-compliant to IEC 60601

    FDA · DeviceElekta, Inc.Worldwide distribution. US distribution to AR, IN, LA, OK, TN, VA, WA.…Official notice →
  11. 2023-03-15
  12. US
    Drugs & medicalClass II

    Elekta Inc recalls Monaco RTP System. Used to make treatment plans for patients with prescriptions for external beam radiation therapy.

    Re-optimization, after adding contours without forced density outside the external structure, may result in inaccurate dose presentation.

    FDA · DeviceElekta IncNationwide distribution in the US. International distribution to Albania,…Official notice →
All 18 Elekta recalls →

Other drugs & medical recalls in the United States

  1. 2023-12-13
  2. US
    Drugs & medicalClass II

    IHB OPERATIONS B.V. recalls Freedom Boom models series: F110, F120, F130, F320, F330, F350, F410, and F420 with built in 42A Lightweight Spring Arm part numbers…

    Weakened connection point on the adjustable suspension arm, potentially causing damage or a break.

    FDA · DeviceIHB OPERATIONS B.V.US: AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, LA, MA, MB, MD, ME, MI,…Official notice →
  3. US
    Drugs & medicalClass II

    BioFire Diagnostics, LLC recalls BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, used with Maine Molecular Quality Controls, Inc (MMQCI)…

    Routine quality control testing of affected blood culture identification panel with affected control panel may need to be amended, due to control panel/instrument characteristic changes, that lead C. tropicalis to be…

    FDA · DeviceBioFire Diagnostics, LLCWorldwide Distribution: US (nationwide): DE, OH, NY, IL, KY, FL, GA, KS, CA,…Official notice →
  4. US
    Drugs & medicalClass II

    AURIS HEALTH INC recalls Monarch Platform REF MON-000006 100-240V- 50/60Hz 1440w

    Their is the potential that software issues may result in flipped image of the virtual Bronchoscope view.

    FDA · DeviceAURIS HEALTH INCUS Distribution: AK, AL, AZ, CA, CO, CT, DC, DE, FL, GA, IL, IN, KY, MA, MI,…Official notice →
  5. US
    Drugs & medicalClass II

    Abiomed, Inc. recalls Automated Impella Controller (AIC), product number 0042-040-US with software versions V8.4 and V8.4.1.

    Pump not detected as connected to controller due to software issue.

    FDA · DeviceAbiomed, Inc.US: AZ, CA, FL, NY & VA.Official notice →
  6. US
    Drugs & medicalClass II

    CardioQuip, LLC recalls CardioQuip Modular Cooler-Heater, Model:MCH-1000(i), Power: 115V-60hz,16A

    There is a potential for the heating-chamber to malfunction resulting in melting/burning of the heater chamber components.

    FDA · DeviceCardioQuip, LLCUS Distribution: TX, IN, IL, WI, FL, NJ, PA, KY, VA, WV, CT, CA, WA, AL, IAOfficial notice →
  7. US
    Drugs & medicalClass II

    Carestream Health, Inc. recalls DRX Revolution Mobile X-Ray System

    Unexpected failure of electrical components within the CPI generator.

    FDA · DeviceCarestream Health, Inc.Worldwide Distribution: US (Nationwide Distribution); and OUS (Foreign):…Official notice →
See all drugs & medical recalls →