Drugs & medicalClass II

CardioQuip, LLC recalls CardioQuip Modular Cooler-Heater, Model:MCH-1000(i), Power: 115V-60hz,16A

There is a potential for the heating-chamber to malfunction resulting in melting/burning of the heater chamber components.

Read the official notice →
Date
2023-12-13
Country
United States
Brand
CardioQuip
Product
CardioQuip Modular Cooler-Heater, Model:MCH-1000(i), Power: 115V-60hz,16A
Classification
Class II
Units
US Distribution: TX, IN, IL, WI, FL, NJ, PA, KY, VA, WV, CT, CA, WA, AL, IA
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More CardioQuip recalls

  1. 2023-12-27
  2. US
    Drugs & medicalClass II

    CardioQuip, LLC recalls CardioQuip Modular Cooler-Heater, Model: MCH-1000(m), Power: 115V-60Hz, 16A

    There is the potential that cooler-heater devices may contain bacterial contamination which could result in patient infection.

    FDA · DeviceCardioQuip, LLCUS: California, Texas, Pennsylvania, Washington, Ohio, Nebraska OUS: NoneOfficial notice →
  3. 2021-10-13
  4. US
    Drugs & medicalClass II

    CardioQuip, LLC recalls CardioQuip, Modular Cooler-Heater 1000(i), Refrigeration Module, Model: MCH-10RMS, UDI: (01)00860000846127 used with heat exchange devices.

    There is the potential bacterial including Nontuberculous mycobacterial (NTM) to grow in the water system of heater-cooler devices which could result in contaminated water potentially aerosolized into the operating…

    FDA · DeviceCardioQuip, LLCWorldwide distribution - US Nationwide distribution including in the states of…Official notice →
  5. US
    Drugs & medicalClass II

    CardioQuip, LLC recalls CardioQuip, Modular Cooler-Heater, Model: MCH-1000(i), CE 2797, UDI: (01)00860000846103

    Labeling does not include guidance for proper water-quality maintenance and device inspection for heater cooler devices.

    FDA · DeviceCardioQuip, LLCWorldwide distribution - US Nationwide distribution in the states of AK, AL,…Official notice →
All 4 CardioQuip recalls →

Other drugs & medical recalls in the United States

  1. 2023-12-13
  2. US
    Drugs & medicalClass II

    AURIS HEALTH INC recalls Monarch Platform REF MON-000006 100-240V- 50/60Hz 1440w

    Their is the potential that software issues may result in flipped image of the virtual Bronchoscope view.

    FDA · DeviceAURIS HEALTH INCUS Distribution: AK, AL, AZ, CA, CO, CT, DC, DE, FL, GA, IL, IN, KY, MA, MI,…Official notice →
  3. US
    Drugs & medicalClass II

    Elekta, Inc. recalls ELEKTA LIMITED Elekta Infinity, REF XRT 0511, XRT 0521, XRT 0531, XRT 0541; Medical Charged-Particle Radiation Therapy System Accelerator,…

    Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there…

    FDA · DeviceElekta, Inc.worldwideOfficial notice →
  4. US
    Drugs & medicalClass II

    Abiomed, Inc. recalls Automated Impella Controller (AIC), product number 0042-040-US with software versions V8.4 and V8.4.1.

    Pump not detected as connected to controller due to software issue.

    FDA · DeviceAbiomed, Inc.US: AZ, CA, FL, NY & VA.Official notice →
  5. US
    Drugs & medicalClass II

    Carestream Health, Inc. recalls DRX Revolution Mobile X-Ray System

    Unexpected failure of electrical components within the CPI generator.

    FDA · DeviceCarestream Health, Inc.Worldwide Distribution: US (Nationwide Distribution); and OUS (Foreign):…Official notice →
  6. US
    Drugs & medicalClass II

    Stryker Medical Division of Stryker Corporation recalls ProCuity bed series, model number 3009, item number: 3009PX-LE-400.

    Patient beds are missing test values for electrical safety tests presenting risk of tissue burn or electrical shock.

    FDA · DeviceStryker Medical Division of Stryker CorporationDistribution US Nationwide including Puerto Rico, Australia, Canada, Chile,…Official notice →
  7. US
    Drugs & medicalClass I

    Maquet Medical Systems USA recalls CARDIOHELP Emergency Drive, Part Numbers 701048002 and 701076205. The CARDIOHELP System is a blood oxygenation and carbon dioxide…

    Product removal due to possible blocking or impairment of the CARDIOHELP Emergency Drive. This issue can lead to the patient being exposed to inappropriate low blood flow. Potential associated harms are ischemia and…

    FDA · DeviceMaquet Medical Systems USAUS Nationwide - Worldwide Distribution: Domestic distribution to: AL, CA, DC,…Official notice →
See all drugs & medical recalls →