DeVilbiss Healthcare LLC recalls DeVilbiss, 525 5-liter Oxygen Concentrator Models 525DS, 525DS-Q
The firm has received complaints of the device becoming hot. The firm has updated the user manual to include warnings that applied parts in contact with the patient may in some situations exceed 41 degrees Celsius…
Read the official notice →- Date
- 2022-12-28
- Country
- United States
- Brand
- DeVilbiss Healthcare
- Product
- DeVilbiss, 525 5-liter Oxygen Concentrator Models 525DS, 525DS-Q
- Classification
- Class II
- Units
- US Nationwide. Foreign distribution pending.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More DeVilbiss Healthcare recalls
- 2023-05-10
US
DeVilbiss Healthcare LLC recalls DEVILBISS HEALTHCARE 1025 SERIES OXYGEN CONCENTRATOR, Model Numbers 1025DS, 1025UK, 1025KS, and 1025KS-AR
The instruction manual is being replaced with an updated version that is in alignment with labeling requirements specified in the IEC-60101-1 standard.
- 2022-11-23
US
DeVilbiss Healthcare LLC recalls 125 series iGO2 Portable Oxygen Concentrator-indicated for the administration of supplemental oxygen. Models: 125D, 125D-XB
An unapproved adhesive used to manufacture the accumulator tank sub assembly (a part of the patient gas pathway)
FDA · DeviceDeVilbiss Healthcare LLCUS Distribution in states of: CA, FL, OH, MI, MN, NY, PA, and TX; and…Official notice →
Other drugs & medical recalls in the United States
- 2022-12-28
US
Icotec Ag recalls VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702
During internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring…
US
Corin Ltd recalls Corin BIOLOX DELTA MOD HEAD, 036mm EX LONG +8mm, REF 104.3615
The BIOLOX Delta Ceramic Head size 36XL is labelled as the size 32XL and vice-versa.
FDA · DeviceCorin LtdUS Nationwide distribution in the states of FL, GA, MA, MI, Ok, PA, TX.Official notice →US
Young Dental Manufacturing I, LLC recalls Denticator PICK-A-DENT, periodontal aid - intradental pick, REF 621714
mislabeled product
FDA · DeviceYoung Dental Manufacturing I, LLCWorldwide distribution - US Nationwide distribution in the states of IA, NY,…Official notice →US
ConvaTec, Inc recalls ConvaTec AQUACEL Foam Ag Adhesive, wound dressing advesive
ConvaTec are voluntarily recalling a batch of AQUACEL¿ Foam Ag Adhesive 10cm x 10cm (x 10 pack) due to the identification of visual contamination in the form of brown spots present on several dressings.
US
GE Medical Systems, LLC recalls Centricity Universal Viewer 6.0. Used to display medical images (Including mammograms) and data from various imaging sources.
There is the potential that after installation of the IAS tool, some studies can remain hidden without user notification. Access to those hidden studies requires GE Healthcare assistance, which can result in a delay in…
FDA · DeviceGE Medical Systems, LLCNationwide distribution to CA, CO, DE, FL, IA, IL, MA, MI, MN, MO, MS, MT, ND,…Official notice →US