Icotec Ag recalls VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702
During internal testing (at manufacturer site) of two torque wrench devices, it was detected that the torque limiting function was out of specification. The specified torque limiting value is 12 Nm. The measuring…
Read the official notice →- Date
- 2022-12-28
- Country
- United States
- Brand
- Icotec
- Product
- VADER Pedicle System Torque Wrench, Hudson Coupling, 12 Nm, REF 42-702
- Classification
- Class II
- Units
- US Nationwide distribution in the state of GA.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Icotec recalls
- 2024-01-17
US
Icotec Ag recalls VADER pedicle system, torque wrench, Catalog Number 42-702
icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the…
Other drugs & medical recalls in the United States
- 2022-12-28
US
Natus Medical Incorporated recalls The Natus Cranial Access Kit is designed to be a complete, single use and disposable accessory and instrumentation set that is used to…
Due to Cranial Access Kits containing recalled sterile surgical drapes. The sterile surgical drapes are being recalled due to the liner on the adhesive component is difficult to remove without damaging the product and…
FDA · DeviceNatus Medical IncorporatedU.S. Nationwide distribution in the states of AK, CA, CT, DC, FL, GA, IA, ID,…Official notice →US
Fresenius Medical Care Holdings, Inc. recalls 2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy in a healthcare…
(1)Software version 2.74 upgrade and (2)hardware related cause that affects only CDX Machines, has been identified as the UI-MICs Board is sensitive to Electromagnetic Interference (EMI) which will cause the false…
US
GE Medical Systems, LLC recalls B105P Patient Monitor (with E-module slot option), REF 6160000-001-XXXXXX,
GE Healthcare has become aware that if certain B1x5P / B1x5M patient monitors use "mmHg" or "kPa" as the unit for CO2 measurement, the displayed value for CO2 measurement can be inaccurate when used in a location that…
US
Corin Ltd recalls Corin BIOLOX DELTA MOD HEAD, 036mm EX LONG +8mm, REF 104.3615
The BIOLOX Delta Ceramic Head size 36XL is labelled as the size 32XL and vice-versa.
FDA · DeviceCorin LtdUS Nationwide distribution in the states of FL, GA, MA, MI, Ok, PA, TX.Official notice →US
Young Dental Manufacturing I, LLC recalls Denticator PICK-A-DENT, periodontal aid - intradental pick, REF 621714
mislabeled product
FDA · DeviceYoung Dental Manufacturing I, LLCWorldwide distribution - US Nationwide distribution in the states of IA, NY,…Official notice →US
DeVilbiss Healthcare LLC recalls DeVilbiss, 525 5-liter Oxygen Concentrator Models 525DS, 525DS-Q
The firm has received complaints of the device becoming hot. The firm has updated the user manual to include warnings that applied parts in contact with the patient may in some situations exceed 41 degrees Celsius…