Drugs & medicalClass II

CareFusion 303, Inc. recalls BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES Tower, REF: 352, BD Pyxis…

Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to…

Read the official notice →
Date
2025-12-31
Country
United States
Brand
CareFusion
Product
BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES Tower, REF: 352, BD Pyxis…
Classification
Class II
Units
Worldwide - US Nationwide distribution including in the states of CO, IN, OH,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More CareFusion recalls

  1. 2026-02-04
  2. US
    Drugs & medicalClass II

    CareFusion 303, Inc. recalls BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a modular infusion pump…

    Due to a number of modules flashed with a date and time associated with the daylight savings time (DST) adjustment that may cause connectivity issue with hospital networks. This issue only affects the Interoperability…

    FDA · DeviceCareFusion 303, Inc.U.S. Nationwide distribution in the states of AZ, IN, OH, and VA.Official notice →
  3. 2026-01-14
  4. US
    Drugs & medicalClass II

    CareFusion 303, Inc. recalls BD Pyxis ES Enterprise Server Catalog Numbers 1115-00, 137672-02,137672-03 137671-02, 137671-03, 137670-02, 137670-03, 139676-01,…

    Due a software issue that may result in equipment not receiving timely updates of patient, medication, or related information from enterprise systems.

    FDA · DeviceCareFusion 303, Inc.Worldwide - U.S. Nationwide distribution including in the states of AK,AL,AR,…Official notice →
  5. 2026-01-07
  6. US
  7. 2025-11-26
  8. US
    Drugs & medicalClass I

    CareFusion 303, Inc. recalls BD Alaris infusion Pump Module 8100, REF: 8100; KIT, BEZEL ASSY 8100BD, REF: 49000437; KIT ASSY ONE PIECE BEZEL 8100, REF: 49000270. Used…

    If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred…

    FDA · DeviceCareFusion 303, Inc.Worldwide distribution. US nationwide including Puerto Rico, Guam, American…Official notice →
  9. 2025-11-05
  10. US
    Drugs & medicalClass II

    CareFusion 303, Inc. recalls BD Pyxis Pro MedStation Main, REF: 1155-00

    Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.

    FDA · DeviceCareFusion 303, Inc.Worldwide distribution - US Nationwide and the countries of Canada, United…Official notice →
  11. 2025-10-15
  12. US
    Drugs & medicalClass I

    CareFusion 303, Inc. recalls BD Alaris Pump Infusion Sets for the following reference numbers: C24101E, Amber Tubing, 15 Micron Filter, 0.2 Micron Filter, SmartSite…

    Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion…

    FDA · DeviceCareFusion 303, Inc.US: Nationwide OUS: United Arab Emirates Australia Belgium Bahrain Botswana…Official notice →
All 64 CareFusion recalls →

Other drugs & medical recalls in the United States

  1. 2025-12-31
  2. US
    Drugs & medicalClass II

    ICU Medical, Inc. recalls LifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02

    For certain customers, the Change Summary review option ceases to operate as expected for any migrated library after the migration has been completed. Users must use alternate methods to review changes. Failure to do…

    FDA · DeviceICU Medical, Inc.US Nationwide distribution in the states of CA, IA, LA, NY, PA, TN, VA.Official notice →
  3. US
    Drugs & medicalClass II

    Intuitive Surgical, Inc. recalls da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40

    Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be…

    FDA · DeviceIntuitive Surgical, Inc.US Nationwide distribution in the states of MN, FL, CO, CA, NC, KS, TX, SC,…Official notice →
  4. US
    Drugs & medicalClass II

    Howmedica Osteonics Corp. recalls The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrared (IR)…

    When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.

    FDA · DeviceHowmedica Osteonics Corp.Worldwide distribution - US Nationwide and the countries of Canada, Germany,…Official notice →
  5. US
    Drugs & medicalClass I

    Alcon Research LLC recalls Brand Name: Alcon Laboratories, Inc. Product Name: Custom Pak Model/Catalog Numbers: 9903-20, 17188-04, 19762-03, 19203-04, 19203-04,…

    Ophthalmic procedure packs may have incomplete seals affecting sterility.

    FDA · DeviceAlcon Research LLCUS Nationwide distribution in the states of CA, ND, UT.Official notice →
  6. US
    Drugs & medicalClass I

    Draeger, Inc. recalls Draeger Vapor 2000. Unheated, calibrated anesthetic vaporizer.

    A certain component of affected devices was not delivered within specification and contained impurities.

    FDA · DeviceDraeger, Inc.Worldwide - US Nationwide distribution in the states of AK, CA, FL, IA, KS,…Official notice →
  7. US
    Drugs & medicalClass II

    AGFA Healthcare Corp. recalls Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800

    It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.

    FDA · DeviceAGFA Healthcare Corp.U.S. Nationwide distribution.Official notice →