Drugs & medicalClass II

ICU Medical, Inc. recalls LifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02

For certain customers, the Change Summary review option ceases to operate as expected for any migrated library after the migration has been completed. Users must use alternate methods to review changes. Failure to do…

Read the official notice →
Date
2025-12-31
Country
United States
Brand
ICU Medical
Product
LifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02
Classification
Class II
Units
US Nationwide distribution in the states of CA, IA, LA, NY, PA, TN, VA.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More ICU Medical recalls

  1. 2026-01-21
  2. US
    Drugs & medicalClass II

    ICU Medical Inc. recalls Pulsator Arterial Blood Sampling Kit, Item Numbers 4044EG-1 4080P-2 4084P-1 4080PE 4080PG

    Affected lots of syringes may have a crack on the syringe collar, which may lead to delayed or repeated blood draw/fluid delivery. This issue may also lead to blood leakage during sampling of arterial blood or leakage…

    FDA · DeviceICU Medical Inc.US Nationwide distribution in the states of CA, FL, IL, IN, MA, MI, MN, MO,…Official notice →
  3. US
    Drugs & medicalClass II

    ICU Medical, Inc. recalls Plum Duo Infusion System, List Number 40002-04-01

    ICU Medical received one lot of defective primary speakers from our supplier that may fail to produce an audible sound. If the primary speaker in your Plum Duo Infusion Pump fails, the pump may not produce audible…

    FDA · DeviceICU Medical, Inc.Worldwide - US Nationwide distribution.Official notice →
  4. 2025-12-31
  5. US
    Drugs & medicalClass II

    ICU Medical, Inc. recalls LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02

    Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they…

    FDA · DeviceICU Medical, Inc.US Nationwide distribution in the states of CA, IA, KY, LA, MD, MS, NY, PA,…Official notice →
  6. 2025-11-05
  7. US
    Drugs & medicalClass II

    ICU Medical, Inc. recalls ICU Medical ChemoLock Closed Vial Spike w/Skirt, 5 Units REF: CL-80S-5

    Port weld of drug transfer device may separate or break during use and potentially result in a leak.

    FDA · DeviceICU Medical, Inc.Worldwide distribution: US (Nationwide) including states of: AK AL AR AZ CA CO…Official notice →
  8. 2025-10-29
  9. US
    Drugs & medicalClass I

    ICU Medical, Inc. recalls IV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23" (58 cm) 150 ml Burette Set…

    IV Gravity burette administration set burette component is missing an internal shutoff valve intended to stop fluid flow, which if used may result in delay in therapy during setup, over delivery/unrestricted flow, or…

    FDA · DeviceICU Medical, Inc.Worldwide distribution: US (nationwide) to states of: TX, SC and OUS (foreign)…Official notice →
  10. 2025-09-10
  11. US
    Drugs & medicalClass II

    ICU Medical Inc. recalls Smiths Medical CADD Medication Cassette Reservoir with clamp and female Luer. Nonvented stopper included. 50mL, Product Code 21-7001-24

    Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.

    FDA · DeviceICU Medical Inc.Worldwide distribution.Official notice →
All 29 ICU Medical recalls →

Other drugs & medical recalls in the United States

  1. 2025-12-31
  2. US
    Drugs & medicalClass II

    Microbiologics Inc recalls KWIK-STIK 2-Pack, Campylobacter jejuni subsp. jejuni derived from ATCC 29428, Catalog Number 0325P

    Affected units may fail to recover the target microorganism Campylobacter jejuni subsp. jejuni derived from ATCC 29428.

    FDA · DeviceMicrobiologics IncWorldwide - US Nationwide.Official notice →
  3. US
    Drugs & medicalClass II

    Zimmer, Inc. recalls Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509420…

    Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue…

    FDA · DeviceZimmer, Inc.US Nationwide distribution in the states of AL, AZ, CA, FL, IN, KY, MA, MO,…Official notice →
  4. US
    Drugs & medicalClass II

    ICU Medical, Inc. recalls LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02

    Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they…

    FDA · DeviceICU Medical, Inc.US Nationwide distribution in the states of CA, IA, KY, LA, MD, MS, NY, PA,…Official notice →
  5. US
    Drugs & medicalClass II

    Intuitive Surgical, Inc. recalls da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40

    Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be…

    FDA · DeviceIntuitive Surgical, Inc.US Nationwide distribution in the states of MN, FL, CO, CA, NC, KS, TX, SC,…Official notice →
  6. US
    Drugs & medicalClass II

    Howmedica Osteonics Corp. recalls The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrared (IR)…

    When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.

    FDA · DeviceHowmedica Osteonics Corp.Worldwide distribution - US Nationwide and the countries of Canada, Germany,…Official notice →
  7. US
    Drugs & medicalClass II

    CareFusion 303, Inc. recalls BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES Tower, REF: 352, BD Pyxis…

    Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to…

    FDA · DeviceCareFusion 303, Inc.Worldwide - US Nationwide distribution including in the states of CO, IN, OH,…Official notice →