ICU Medical recalls
29 official recalls and safety alerts involving ICU Medical since 2022-09, from FDA · Device, FDA · Drug. Newest first. Check the official notice for exact models and batch codes.
This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with ICU Medical or the relevant agency.
- 2026-09-23
US
ICU Medical, Inc. recalls CADD Yellow Medication Cassette Reservoir with NRFit connector US Product Codes: 21-7600-24, 21-7609-24 OUS Product Codes: 21-7609-24…
Reservoirs may exhibit medication leakage at the joint between luer connector and tubing. Issue could result in delay or interruption of therapy, under delivery of medication, exposure to infectious or toxic agents, or…
FDA · DeviceICU Medical, Inc.Worldwide - US Nationwide distribution in the states of AR, CA, FL, IL, IN,…Official notice →US
ICU Medical, Inc. recalls Oncology sets with drip chambers
Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not…
FDA · DeviceICU Medical, Inc.Worldwide - US Nationwide distribution including in the states and territories…Official notice →- 2026-07-29
US
ICU Medical Inc. recalls Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the patient's…
ICU Medical became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate solvent application at the tubing to return connector bond.
- 2026-07-22
US
ICU Medical, Inc. recalls ICU Medical Plum Duo Infusion Pump, v1.2.3: 1) Item Number 400020401 and 2) Item Number 400020403; infusion pump
In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate.
FDA · DeviceICU Medical, Inc.Worldwide distribution - US Nationwide and the country of Canada.Official notice →- 2026-06-17
US
ICU Medical, Inc. recalls Plum Duo Precision IV Pump, 40002-0403
Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or…
FDA · DeviceICU Medical, Inc.Worldwide - US Nationwide distribution in the states of ID, IL, IN, KS, MA,…Official notice →- 2026-05-27
US
ICU Medical, Inc. recalls Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; ChemoLock; Syringe Transfer…
Oncology sets/closed system transfer devices with extension sets may leak resulting in therapy delayed if noticed and corrected during setup, but if the leak occurs during therapy this may result in therapy…
FDA · DeviceICU Medical, Inc.US Nationwide distribution in the states of GA, MA, OH, NC, CA, ND, MI, TX,…Official notice →- 2026-01-28
US
ICU Medical, Inc. recalls ICU Medical Plum Duo Infusion Pump, List Numbers: 400020401 (US), 400021001 (Philippines)
Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at…
FDA · DeviceICU Medical, Inc.Worldwide distribution - US Nationwide and the country of Philippines.Official notice →US
ICU Medical, Inc. recalls Tego Connector: REF: 011-D1000, 01C-D1000, 055-D1000, D1000; Conector Tego: REF: LAT-D1000
Needle free access device intended as accessory to vascular access used in hemodialysis or accessory to intravascular administration sets has a silicone seal which may bulge at the top, separate from the body, tear;…
FDA · DeviceICU Medical, Inc.Worldwide - US Nationwide distribution including in the states of FL, AZ, RI,…Official notice →- 2026-01-21
US
ICU Medical Inc. recalls Pulsator Arterial Blood Sampling Kit, Item Numbers 4044EG-1 4080P-2 4084P-1 4080PE 4080PG
Affected lots of syringes may have a crack on the syringe collar, which may lead to delayed or repeated blood draw/fluid delivery. This issue may also lead to blood leakage during sampling of arterial blood or leakage…
FDA · DeviceICU Medical Inc.US Nationwide distribution in the states of CA, FL, IL, IN, MA, MI, MN, MO,…Official notice →US
ICU Medical, Inc. recalls Plum Duo Infusion System, List Number 40002-04-01
ICU Medical received one lot of defective primary speakers from our supplier that may fail to produce an audible sound. If the primary speaker in your Plum Duo Infusion Pump fails, the pump may not produce audible…
- 2025-12-31
US
ICU Medical, Inc. recalls LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02
Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they…
FDA · DeviceICU Medical, Inc.US Nationwide distribution in the states of CA, IA, KY, LA, MD, MS, NY, PA,…Official notice →US
ICU Medical, Inc. recalls LifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02
For certain customers, the Change Summary review option ceases to operate as expected for any migrated library after the migration has been completed. Users must use alternate methods to review changes. Failure to do…
FDA · DeviceICU Medical, Inc.US Nationwide distribution in the states of CA, IA, LA, NY, PA, TN, VA.Official notice →- 2025-11-05
US
ICU Medical, Inc. recalls ICU Medical ChemoLock Closed Vial Spike w/Skirt, 5 Units REF: CL-80S-5
Port weld of drug transfer device may separate or break during use and potentially result in a leak.
FDA · DeviceICU Medical, Inc.Worldwide distribution: US (Nationwide) including states of: AK AL AR AZ CA CO…Official notice →- 2025-10-29
US
ICU Medical, Inc. recalls IV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23" (58 cm) 150 ml Burette Set…
IV Gravity burette administration set burette component is missing an internal shutoff valve intended to stop fluid flow, which if used may result in delay in therapy during setup, over delivery/unrestricted flow, or…
FDA · DeviceICU Medical, Inc.Worldwide distribution: US (nationwide) to states of: TX, SC and OUS (foreign)…Official notice →- 2025-09-10
US
ICU Medical Inc. recalls Smiths Medical CADD Medication Cassette Reservoir with clamp and female Luer. Nonvented stopper included. 50mL, Product Code 21-7001-24
Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.
- 2025-08-13
US
ICU Medical, Inc. recalls Plum Duo Infusion System, List Number: 400020401
ICU Medical identified two sequences of programming events and alarm interactions that may cause the user interface to become unresponsive.
- 2025-07-23
US
ICU Medical, Inc. recalls SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve, REF: SCXT3-5-2000; SwabCap, 200-ct. Box, REF: SCXT3-2000
Due to a manufacturing issue, disinfecting cap for needle-free connectors may have an incomplete seal between the foil lid and plastic container, which may result in isopropyl alcohol evaporation from the sponge, which…
FDA · DeviceICU Medical, Inc.Worldwide - US Nationwide distribution including in the states of TX, SD, ID,…Official notice →- 2025-06-25
US
ICU Medical, Inc. recalls ICU Medical Plum Solos, Item number 400011001; infusion pump
Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager…
FDA · DeviceICU Medical, Inc.Worldwide - US Nationwide distribution in the states of FL, KY, VA, MS, MA,…Official notice →- 2025-05-21
US
ICU Medical, Inc. recalls LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software SUite v2.1 and v2.2
Firm has identified Drug Library Management defects in the software: 1) DLM software does not allow user to create percentage dose rate or rate change values with certain limits, potentially resulting in over-delivery…
FDA · DeviceICU Medical, Inc.Worldwide - US Nationwide distribution in the states of CA, FL, IA, KY, LA,…Official notice →- 2025-03-19
US
ICU Medical, Inc. recalls POTASSIUM CHLORIDE Inj., 10 mEq total in 100 mL, 100 mEq/L flexible container, Rx Only, ICU Medical, Inc., Lake Forest, Illinois, 60045,…
Labeling: Label Error on Declared Strength. Overwrap labeled as Potassium Chloride Inj 10 mEq may contain flexible containers of Potassium Chloride Inj 20 mEq
US
ICU Medical, Inc. recalls POTASSIUM CHLORIDE Inj., 20 mEq total in 100 mL flexible container 24 x case, 200 mEq/L, Rx Only, ICU Medical, Inc., Lake Forest,…
Labeling: Label Error on Declared Strength. Cases labeled POTASSIUM CHLORIDE 20 mEq, may contain flexible containers with overwrap mislabeled as 10 mEq. The correct dosage strength of 20 mEq is printed on the labeling…
- 2024-12-18
US
ICU Medical, Inc. recalls ChemoLock Close Vial Spike REF CL-80 The Chemolock Closed System Transfer Device prevents the transfer of environmental contaminants. The…
Due to finished goods incorrectly packaged and labeled.
FDA · DeviceICU Medical, Inc.Worldwide - U.S. Nationwide distribution in the state of IL and the country of…Official notice →- 2024-11-27
US
ICU Medical, Inc. recalls Allegedly counterfeit batteries manufactured by CSB batteries being used with Plum Infusion Systems: Plum 360" Infusion System, Pump List…
ICU Medical has received reports of allegedly counterfeit CSB batteries being used with Plum Infusion Systems. While these non-OEM batteries are visually similar to the Plum batteries supplied by ICU Medical, they are…
FDA · DeviceICU Medical, Inc.Worldwide distribution: US (nationwide) including Puerto Rico and OUS…Official notice →- 2024-08-21
US
ICU Medical Inc recalls Replacement Battery List Number SUB0000594 and SUB0000864 found in Plum A+ & Plum A+3 Infusion Systems, Pump List Numbers: 11005, 11971,…
Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the…
FDA · DeviceICU Medical IncWorldwide Distribution: US (nationwide) and OUS (foreign) countries of:…Official notice →- 2023-08-30
US
ICU Medical Inc recalls Plum 360 Infusion System, List Numbers: 300100405, 300100409, 300100413, 300100414, 300100415, 300100418, 300100451, 300100452, 300100453,…
Due to a manufacturing defect of a supplier provided component, there is a potential that the audible signal for an alarm may not sound under certain conditions.
- 2023-05-24
US
ICU Medical Inc recalls Replacement Battery List Number SUB0000864, found in Plum 360 Infusion System, List Number 30010
Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the…
- 2022-11-23
US
ICU Medical Inc recalls 0.9% Sodium Chloride Injection, USP, 100 mL flexible container bag, Rx Only, ICU Medical, Inc., Lake Forest, Illinois, 60045, USA, NDC…
Lack of assurance of sterility: Bags have the potential to leak in the flexible containers which may compromise sterility.
- 2022-09-21
US
ICU Medical, Inc. recalls 6.5 IN (17 cm) APPX 0.78 ml, EXT SET w/Clave CLEAR, Clave 4-WAY Stopcock, Spiros, RED CAP. Accessory for intravascular administration set.
Due to a manufacturing defect which results in a visible gap between the connector's top and bottom housing.
FDA · DeviceICU Medical, Inc.U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN,…Official notice →- 2022-09-14
US
ICU Medical, Inc. recalls Product Description/Item: ChemoLock Bag Spike, 10 units/CL-10-10; ChemoLock BAG SPIKE w/ ADDITIVE PORT, DRY SPIKE/CL-12; APPX…
Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and exposure to caustic…
FDA · DeviceICU Medical, Inc.Worldwide distribution - US Nationwide distribution including in the states of…Official notice →