Recall Atlas
Drugs & medicalClass II

Cadwell Industries Inc recalls Brand Name: IOMAX Product Name: IOMAX Cortical Module Model/Catalog Number: 190296-200 UDI-DI code: 00840067100653 Product Description:…

Stimulator device contains a defective printed circuit board assembly (PCBA) that may cause the device to unintentionally deliver electrical stimulation to the patient, or allow an unintended electrical current to…

Read the official notice →
Date
2026-10-21
Country
United States
Brand
Cadwell Industries
Product
Brand Name: IOMAX Product Name: IOMAX Cortical Module Model/Catalog Number: 190296-200 UDI-DI code: 00840067100653 Product Description:…
Classification
Class II
Units
Worldwide - US Nationwide distribution in the state of TN and the countries of…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Cadwell Industries recalls

  1. 2024-07-24
  2. Drugs & medicalClass II

    Cadwell Industries Inc recalls Cascade IOMAX Cortical Module, REF: 190296-200

    Electroneurodiagnostic medical device may contain a defective printed circuit board which may cause damage to the other internal components resulting in unintended electrical stimulation.

    FDA · DeviceCadwell Industries IncUS: IL CA OUS: Canada, Germany, India, Italy, Mexico, Poland, South Africa,…Official notice →
  3. 2024-05-29
  4. Drugs & medicalClass II

    Cadwell Industries Inc recalls Cascade IOMAX Cortical Module, REF: 190296-200

    There is the potential that an electroneurodiagnostic medical device insulator may be installed incorrectly which could result in arcing to patient connections resulting in potential unintended shock to the patient.

    FDA · DeviceCadwell Industries IncWorldwide - US Nationwide distribution in the state of MD and the countries of…Official notice →
All 3 Cadwell Industries recalls →

Other drugs & medical recalls in the United States

  1. 2026-10-21
  2. Drugs & medicalClass II

    Sarnova HC, Llc recalls Curaplex Capnography EtCO2/O2 Oral/Nasal Cannula Multipack (10 pk), Item Number 670272-KIT

    Affected kits were assembled with pediatric nasal cannula instead of the intended adult nasal cannula.

    FDA · DeviceSarnova HC, LlcUS Nationwide distribution in the states of IA, IL, IN, KY, MI, MN, MO, ND,…Official notice →
  3. Drugs & medicalClass II

    Boston Scientific Corporation recalls NaviFlex LW Pusher; Advx Panc. Stent 4/5F; Material Number (UPN): M00535030;

    Device was incorrectly manufactured with a guidewire exit port (referred to in the IFU as the RX port), which is not a design feature associated with this device configuration.

    FDA · DeviceBoston Scientific CorporationUS Nationwide distribution in the states of AL, CA, FL, GA, IL, IN, KY, MD,…Official notice →
  4. Drugs & medicalClass II

    Neuronetics, Inc. recalls NeuroStar Advanced Therapy System. Contains the NeuroStar HP Coil Assembly Part Number 81-01000-200. Affects the below NeuroStar Advanced…

    Cables are not in conformance with the product specifications required in the production of the HP Coil Assembly.

    FDA · DeviceNeuronetics, Inc.Worldwide distribution - US Nationwide and the countries of Japan and South…Official notice →
  5. Drugs & medicalClass II

    Becton Dickinson & Co. recalls BD Phoenix Panel NMIC-312. Catalog Number: 449082.

    Potential for quality control failures with Klebsiella pneumoniae ATCC BAA-1705 when using the CPO Ambler classification feature.

    FDA · DeviceBecton Dickinson & Co.US Nationwide distribution.Official notice →
  6. Drugs & medicalClass II

    Medtronic Sofamor Danek USA Inc recalls Grafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with an inert additive to yield a…

    Medtronic identified through donor record review that the donor for Medtronic batch A62881 did not meet all Medtronic donor eligibility requirements related to the presence of microbial organisms. While the batch…

    FDA · DeviceMedtronic Sofamor Danek USA IncUS Nationwide distribution in the states of OK, VA, CO, AR, FL, TX, MD, DE,…Official notice →
  7. Drugs & medicalClass II

    Jenavalve Technology Inc. recalls JenaValve Trilogy Delivery Catheter, Size 27MM REF: TRILOGY-DC-L

    Delivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device.

    FDA · DeviceJenavalve Technology Inc.US Nationwide distribution in the states of AZ, CA, FL, GA, IL, MA, MD, MI,…Official notice →