Becton Dickinson & Co. recalls BD Phoenix Panel NMIC-312. Catalog Number: 449082.
Potential for quality control failures with Klebsiella pneumoniae ATCC BAA-1705 when using the CPO Ambler classification feature.
Read the official notice →- Date
- 2026-10-21
- Country
- United States
- Brand
- Becton Dickinson &
- Product
- BD Phoenix Panel NMIC-312. Catalog Number: 449082.
- Classification
- Class II
- Units
- US Nationwide distribution.
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More Becton Dickinson & recalls
- 2026-06-10
Becton Dickinson & Company recalls BD¿Spinal Tray with BD¿Whitacre Needle 22 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) with dextrose (8.25%), 2 mL…
This is a sub-recall of a product recall from Huons Co, Ltd., the manufacturer of Bupivacaine Hydrochloride in Dextrose Injection, USP. The recall is due to quality issues identified during an FDA manufacturing…
FDA · DeviceBecton Dickinson & CompanyNationwide distribution to AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS,…Official notice →- 2026-05-06
Becton Dickinson & Co. recalls BD GasPak EZ CO2 Pouch System. Catalog Number: 260684. for microbiologic use
BD identified an issue affecting certain lots of BD GasPak EZ CO2 and BD GasPak EZ Campy Gas Generating Pouch Systems, where gas generating sachets may produce carbon dioxide levels below specification.
FDA · DeviceBecton Dickinson & Co.Worldwide distribution - US Nationwide and the country of Hong Kong.Official notice →- 2026-03-18
Becton Dickinson & Company recalls 10mL Non-Sterile BD Luer-Lok Tip Syringe Only. Catalog Number: 301029. 850 bulk non-sterile units per case.
During the bulk packaging process, some Luer Slip syringes were incorrectly included in a batch designated for Luer Lok syringes.
FDA · DeviceBecton Dickinson & CompanyUS Nationwide distribution in the states of California and Texas.Official notice →- 2025-11-26
Becton Dickinson & Company recalls BD Luer Tip Caps; Catalog Number(s) or Model Number(s): 308341;
BD has confirmed through internal testing that certain device trays failed routine biocompatibility testing.
FDA · DeviceBecton Dickinson & CompanyWorldwide - US Nationwide distribution including in the states of AL, AZ, CA,…Official notice →- 2025-11-19
Becton Dickinson & Co. recalls BD Kiestra" Isolate Suspension Cuvette Array; Catalog No.: 246100;
The 2D barcodes on these cuvettes contain formatting errors that prevent the analyzing system from correctly reading the expiration date. As a result, the instrument enters an error state and the cuvettes cannot be used.
FDA · DeviceBecton Dickinson & Co.Worldwide - US Nationwide distribution in the states of CO, CT, KY, NJ, TN,…Official notice →- 2025-10-29
Becton Dickinson & Co. recalls BD EpiCenter Microbiology Data Management System, labeled as the following: 1. Computer System Epicenter F/G, Catalog Number: 441002. 2.…
Product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of…
FDA · DeviceBecton Dickinson & Co.Worldwide distribution. US Nationwide including Guam and Puerto Rico, Algeria,…Official notice →
Other drugs & medical recalls in the United States
- 2026-10-21
Boston Scientific Corporation recalls NaviFlex LW Pusher; Advx Panc. Stent 4/5F; Material Number (UPN): M00535030;
Device was incorrectly manufactured with a guidewire exit port (referred to in the IFU as the RX port), which is not a design feature associated with this device configuration.
FDA · DeviceBoston Scientific CorporationUS Nationwide distribution in the states of AL, CA, FL, GA, IL, IN, KY, MD,…Official notice →Neuronetics, Inc. recalls NeuroStar Advanced Therapy System. Contains the NeuroStar HP Coil Assembly Part Number 81-01000-200. Affects the below NeuroStar Advanced…
Cables are not in conformance with the product specifications required in the production of the HP Coil Assembly.
FDA · DeviceNeuronetics, Inc.Worldwide distribution - US Nationwide and the countries of Japan and South…Official notice →Cadwell Industries Inc recalls Brand Name: IOMAX Product Name: IOMAX Cortical Module Model/Catalog Number: 190296-200 UDI-DI code: 00840067100653 Product Description:…
Stimulator device contains a defective printed circuit board assembly (PCBA) that may cause the device to unintentionally deliver electrical stimulation to the patient, or allow an unintended electrical current to…
FDA · DeviceCadwell Industries IncWorldwide - US Nationwide distribution in the state of TN and the countries of…Official notice →Medtronic Sofamor Danek USA Inc recalls Grafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with an inert additive to yield a…
Medtronic identified through donor record review that the donor for Medtronic batch A62881 did not meet all Medtronic donor eligibility requirements related to the presence of microbial organisms. While the batch…
FDA · DeviceMedtronic Sofamor Danek USA IncUS Nationwide distribution in the states of OK, VA, CO, AR, FL, TX, MD, DE,…Official notice →Jenavalve Technology Inc. recalls JenaValve Trilogy Delivery Catheter, Size 27MM REF: TRILOGY-DC-L
Delivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device.
FDA · DeviceJenavalve Technology Inc.US Nationwide distribution in the states of AZ, CA, FL, GA, IL, MA, MD, MI,…Official notice →Instrumentation Laboratory recalls HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009802028. in vitro diagnostic (OUS only)
Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.
FDA · DeviceInstrumentation LaboratoryWorldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, FL,…Official notice →