Drugs & medicalClass III

Becton Dickinson & Company recalls BD Vacutainer Luer Lok Access Device - intended as a sterile, single use device designed to provide a connection to a catheter port…

May shed contaminant particles into the urine specimen that may be misidentified as Red Blood Cells when used in conjunction with the Sysmex UF-1000i or UX-2000 Analyzers and lead to false elevations in RBC counts

Read the official notice →
Date
2022-03-02
Country
United States
Brand
Becton Dickinson &
Product
BD Vacutainer Luer Lok Access Device - intended as a sterile, single use device designed to provide a connection to a catheter port…
Classification
Class III
Units
US Nationwide distribution in the state of IL.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Becton Dickinson & recalls

  1. 2022-10-26
  2. US
    Drugs & medicalClass II

    Becton Dickinson & Company recalls BD Vacutainer Tube Sodium Fluoride: 100 mg, Potassium Oxalate: 20 mg, Catalog Number 367001. For collecting, transporting,…

    The affected product contains isobutylene which has recently demonstrated potential interference in determination of methanol by gas chromatography methods. The interference may lead to false positive results in…

    FDA · DeviceBecton Dickinson & CompanyDistribution US nationwide, Canada, New Zealand, and Hong Kong.Official notice →
  3. 2022-08-24
  4. US
    Drugs & medicalClass II

    Becton Dickinson & Company recalls BD PhaSealTM Y-Site Connector (C80)- airtight and leak-proof Closed System Drug Transfer Device (CSTD) that mechanically prohibits the…

    Incorrect expiration date listed on the product labels. The expiration date listed on the product label is beyond the expiration date of the product.

    FDA · DeviceBecton Dickinson & CompanyUS Nationwide Distribution CA FL GA IL MD MI OR PA PR TN TXOfficial notice →
  5. 2022-07-27
  6. US
    Drugs & medicalClass II

    Becton Dickinson & Company recalls BD Connecta Stopcock Without Extension Tube Catalog No. 394601 (OUS)

    May have the potential for leakage at the housing component of the stopcock, may result in delay or interruption in treatment, exposure to infusate and biohazardous material, under dosing/under infusion, contamination…

    FDA · DeviceBecton Dickinson & CompanyWorldwide distribution - US Nationwide distribution in the states of AK, AR,…Official notice →
  7. 2022-05-11
  8. US
    Drugs & medicalClass II

    Becton Dickinson & Co. recalls BD Synapsys Microbiology Informatics Solution, Catalog Number 444150, Software Version 4.10

    When updating BD Kiestra customers to BD Synapsys version 4.10, the custom antibiotic codes set by the customer are overwritten with the BD default antibiotic codes.

    FDA · DeviceBecton Dickinson & Co.Worldwide distribution US Nationwide distribution in the states of CA, IL, MI,…Official notice →
  9. 2022-02-16
  10. US
    Drugs & medicalClass II

    Becton Dickinson & Co. recalls BD Veritor Plus Analyzer - intended to provide rapid test results in near-patient settings. Catalog Number: 256066

    May have the potential to overheat and/or cause fire, the issue can occur when an unauthorized alternative AC power adapter, and not the one provided by BD is used to power the instrument.

    FDA · DeviceBecton Dickinson & Co.Worldwide distribution - US Nationwide and the countries of Armenia,…Official notice →
  11. 2021-11-03
  12. US
    Drugs & medicalClass II

    Becton Dickinson & Company recalls BD Syringe Tip Cap Bulk Sterile Convenience Pak-1000 units/shelf carton. Used to seal syringes and maintain container closure…

    Holes in product packaging impacts package integrity and may compromise sterility.

    FDA · DeviceBecton Dickinson & CompanyTXOfficial notice →
All 36 Becton Dickinson & recalls →

Other drugs & medical recalls in the United States

  1. 2022-03-02
  2. US
    Drugs & medicalClass II

    Biomet, Inc. recalls Comprehensive Shoulder System Primary Shoulder Stem, Micro Length, Porous Plasma, 11 MM, 55 MM Long Item Number: 113611

    The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be…

    FDA · DeviceBiomet, Inc.Worldwide distribution - US Nationwide and the countries of Canada, Ecuador,…Official notice →
  3. US
    Drugs & medicalClass II

    Collagen Matrix, Inc. recalls DuraMatrix-Onlay Collagen Dura Substitute Membrane 4 inches x 5 inches (10.0cm x 12.5cm) REF: CDSLM45

    Outer packaging (outer pouch) is not sealed compromising sterility and result in potential risk of patient infection, which could lead to revision surgery

    FDA · DeviceCollagen Matrix, Inc.DuraMatrix-Onlay distributed US Nationwide in the states of FL, IN, Puerto…Official notice →
  4. US
    Drugs & medicalClass II

    C.R. Bard Inc recalls BARD SURESTEP INTERMITTENT CATHETER TRAY, 14Fr, REF INTS14C

    There is a potential to leak from the catheter where it connects to the drain bag

    FDA · DeviceC.R. Bard IncUS Nationwide distribution in the states of GA, ME, NJ, WI, MN, SD, CO, ND,…Official notice →
  5. US
    Drugs & medicalClass II

    Stryker Sustainability Solutions recalls REF: 5921-018-135, Stryker Color Cuff, 18" (Red), Quick Connect Single Port, Single Bladder

    Mislabeling; dual-bladder tourniquet cuffs are labeled as single-bladder inflatable cuffs.

    FDA · DeviceStryker Sustainability SolutionsUS: KS OK TX OUS: NoneOfficial notice →
  6. US
    Drugs & medicalClass II

    Natus Neurology DBA Excel Tech., Ltd. (XLTEK) recalls IPELA FHD PTZ TILE KITS with Axis cameras used in combination with Natus Neuroworks/Sleepworks software

    Tile kits that allow one to install and situate a video camera within a patient environment are not fitting properly in some domestic institution ceiling tile grids that have a thinner style grid frame. The ceiling…

    FDA · DeviceNatus Neurology DBA Excel Tech., Ltd. (XLTEK)Worldwide distribution - US Nationwide distribution in the states of FL, IL,…Official notice →
  7. US
    Drugs & medicalClass II

    BALT USA, LLC recalls The Optima Coil System consists of an implantable embolization coil comprised of a platinum-tungsten alloy attached to a proximal…

    Due to Label/Pouch mix-up. Pouches incorrectly labeled. Incorrect shelf carton label or pouch label due to inadequate manufacturing line clearance. As a result of this labeling mix-up, there is the possibility of…

    FDA · DeviceBALT USA, LLCWorldwide distribution - U.S. Nationwide distribution in the states of FL and…Official notice →
See all drugs & medical recalls →