Drugs & medicalClass II

Becton Dickinson & Co. recalls BD Veritor Plus Analyzer - intended to provide rapid test results in near-patient settings. Catalog Number: 256066

May have the potential to overheat and/or cause fire, the issue can occur when an unauthorized alternative AC power adapter, and not the one provided by BD is used to power the instrument.

Read the official notice →
Date
2022-02-16
Country
United States
Brand
Becton Dickinson &
Product
BD Veritor Plus Analyzer - intended to provide rapid test results in near-patient settings. Catalog Number: 256066
Classification
Class II
Units
Worldwide distribution - US Nationwide and the countries of Armenia,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Becton Dickinson & recalls

  1. 2022-10-26
  2. US
    Drugs & medicalClass II

    Becton Dickinson & Company recalls BD Vacutainer Tube Sodium Fluoride: 100 mg, Potassium Oxalate: 20 mg, Catalog Number 367001. For collecting, transporting,…

    The affected product contains isobutylene which has recently demonstrated potential interference in determination of methanol by gas chromatography methods. The interference may lead to false positive results in…

    FDA · DeviceBecton Dickinson & CompanyDistribution US nationwide, Canada, New Zealand, and Hong Kong.Official notice →
  3. 2022-08-24
  4. US
    Drugs & medicalClass II

    Becton Dickinson & Company recalls BD PhaSealTM Y-Site Connector (C80)- airtight and leak-proof Closed System Drug Transfer Device (CSTD) that mechanically prohibits the…

    Incorrect expiration date listed on the product labels. The expiration date listed on the product label is beyond the expiration date of the product.

    FDA · DeviceBecton Dickinson & CompanyUS Nationwide Distribution CA FL GA IL MD MI OR PA PR TN TXOfficial notice →
  5. 2022-07-27
  6. US
    Drugs & medicalClass II

    Becton Dickinson & Company recalls BD Connecta Stopcock Without Extension Tube Catalog No. 394601 (OUS)

    May have the potential for leakage at the housing component of the stopcock, may result in delay or interruption in treatment, exposure to infusate and biohazardous material, under dosing/under infusion, contamination…

    FDA · DeviceBecton Dickinson & CompanyWorldwide distribution - US Nationwide distribution in the states of AK, AR,…Official notice →
  7. 2022-05-11
  8. US
    Drugs & medicalClass II

    Becton Dickinson & Co. recalls BD Synapsys Microbiology Informatics Solution, Catalog Number 444150, Software Version 4.10

    When updating BD Kiestra customers to BD Synapsys version 4.10, the custom antibiotic codes set by the customer are overwritten with the BD default antibiotic codes.

    FDA · DeviceBecton Dickinson & Co.Worldwide distribution US Nationwide distribution in the states of CA, IL, MI,…Official notice →
  9. 2022-03-02
  10. US
    Drugs & medicalClass III

    Becton Dickinson & Company recalls BD Vacutainer Luer Lok Access Device - intended as a sterile, single use device designed to provide a connection to a catheter port…

    May shed contaminant particles into the urine specimen that may be misidentified as Red Blood Cells when used in conjunction with the Sysmex UF-1000i or UX-2000 Analyzers and lead to false elevations in RBC counts

    FDA · DeviceBecton Dickinson & CompanyUS Nationwide distribution in the state of IL.Official notice →
  11. 2021-11-03
  12. US
    Drugs & medicalClass II

    Becton Dickinson & Company recalls BD Syringe Tip Cap Bulk Sterile Convenience Pak-1000 units/shelf carton. Used to seal syringes and maintain container closure…

    Holes in product packaging impacts package integrity and may compromise sterility.

    FDA · DeviceBecton Dickinson & CompanyTXOfficial notice →
All 36 Becton Dickinson & recalls →

Other drugs & medical recalls in the United States

  1. 2022-02-16
  2. US
    Drugs & medicalClass II

    Synthes (USA) Products LLC recalls Universal Chuck-The Universal Chuck is an Instrument Handle provided as an alternative instrument available for use in the…

    Top cap may loosen and detach if the Universal Chuck becomes jammed and the user attempts to manually free the device, internal ball bearings may fall out onto the surgical field and cause surgical delay or, and if not…

    FDA · DeviceSynthes (USA) Products LLCUS Nationwide distribution.Official notice →
  3. US
    Drugs & medicalClass II

    MEDLINE INDUSTRIES, LP - Northfield recalls CURAD REUSABLE GEL HOT/COLD PACK

    The product contains natural rubber latex but the label states that it does not.

    FDA · DeviceMEDLINE INDUSTRIES, LP - NorthfieldUS and Bahamas, Bermuda, Panama, Lebanon, Barbados, Honduras, AnguillaOfficial notice →
  4. US
    Drugs & medicalClass II

    DePuy Orthopaedics, Inc. recalls ATTUNE REVISION POSTERIOR FEMORAL AUGMENT SZ 9 12MM-intended as a Component in total knee arthroplasty (TKA) revision surgeries Product…

    May have adhesive residue on the posterior surface of the femoral augment, if undetected prior to implantation, soft tissue irritation and/or infection may occur.

    FDA · DeviceDePuy Orthopaedics, Inc.US Nationwide distribution in the states of TN, TX.Official notice →
  5. US
    Drugs & medicalClass II

    GE Healthcare, LLC recalls GE Healthcare MR superconducting magnets, a component of GE Healthcare SIGNA Artist system, nuclear magnetic resonance imaging system.

    The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.

    FDA · DeviceGE Healthcare, LLCDistribution was nationwide, including Puerto Rico and the Virgin Islands.…Official notice →
  6. US
    Drugs & medicalClass II

    Microbiologics Inc recalls QC Sets and Panels, KWIK-STIK AST-GP (4 Strains) QC Set. KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control…

    Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used…

    FDA · DeviceMicrobiologics IncWorldwide Distribution. US nationwide including Puerto Rico and Guam,…Official notice →
  7. US
    Drugs & medicalClass II

    Microbiologics Inc recalls QC Sets and Panels: KWIK-STIK" 2 Pack. Common Name: QC Sets and Panels GP Comprehensive QC Set. Each KWIK-STIK unit contains a lyophilized…

    Potential contamination with Escherichia coli Staphylococcus epidermidis and S. warneri.

    FDA · DeviceMicrobiologics IncWorldwide distribution - US Nationwide distribution in the states of AR, AZ,…Official notice →