Drugs & medicalClass II

B. Braun Medical, Inc. recalls Dispensing Pins for Air Inlet Filter, Product Catalog No. 412005, DP2500 Dispensing Pin W/Valve. A valve for aspiration or injection of…

This lot may have an elevated risk of leakage from the Air-Inlet Filter of the device. Leakage of the Air-Inlet Filter may increase the risk of decreasing the dose of medication delivered to the patient, cause delays…

Read the official notice →
Date
2022-03-16
Country
United States
Brand
B. Braun Medical
Product
Dispensing Pins for Air Inlet Filter, Product Catalog No. 412005, DP2500 Dispensing Pin W/Valve. A valve for aspiration or injection of…
Classification
Class II
Units
US Nationwide distribution in the states of AZ, CA, FL, LA, MI, MO, MS, NC,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More B. Braun Medical recalls

  1. 2022-10-26
  2. US
    Drugs & medicalClass II

    B. Braun Medical, Inc. recalls ULTRASITE US04TL Smallbore T-Port Extension Set, Product Code 474515, 100 units per carton. to provide needle-free access on I.V. and…

    The extension set has a high probability of failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of bloodstream…

    FDA · DeviceB. Braun Medical, Inc.Distribution US nationwide, Canada and Saudi Arabia.Official notice →
  3. 2022-10-12
  4. US
    Drugs & medicalClass II

    B Braun Medical Inc recalls Introcan Safety FEP 14G, 2.2x32MM-US, Product Code 4251890-02. A passive anti-needle device to provide venous or arterial access.

    Potential for leakage at the catheter hub.

    FDA · DeviceB Braun Medical IncDomestic distribution to AZ CA FL GA MD MI MO MS NC NY OH PA TN TX WA. Foreign…Official notice →
  5. 2022-04-20
  6. US
    Drugs & medicalClass II

    B Braun Medical Inc recalls Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (END) - VERSION 2, Article Code Number…

    Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration…

    FDA · DeviceB Braun Medical IncUS Nationwide distribution.Official notice →
  7. 2022-03-30
  8. US
    Drugs & medicalClass II

    B Braun Medical Inc recalls PERFUSOR SPACE - US VERSION. Intended Use: syringe infusion pump that can be used to configure customized, tailor-made solutions Catalog…

    Units not properly inspected prior to release to the market, may not function as indicated in the product labeling, resulting in syringe detection and infusion accuracy that can result in delays in therapy,…

    FDA · DeviceB Braun Medical IncUS Distribution to states of: MA, VTOfficial notice →
  9. 2022-03-23
  10. US
  11. 2021-10-06
  12. US
    Drugs & medicalClass II

    B. Braun Medical, Inc. recalls PUMP SET, 15 DR, 3 CARESITE, 127 in.single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of…

    Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing and bloodstream infections.

    FDA · DeviceB. Braun Medical, Inc.Worldwide distribution - US Nationwide distribution and the country of Canada.Official notice →
All 53 B. Braun Medical recalls →

Other drugs & medical recalls in the United States

  1. 2022-03-16
  2. US
    Drugs & medicalClass II

    CELLTRION USA INC recalls Celltrion DiaTrust COVID-19 Ag Rapid Test, Reference No. CT-P60 D-2 02

    Kits were labeled with incorrect expiration dates.

    FDA · DeviceCELLTRION USA INCThe products were distributed to the following US states: CA, CT, FL, HI, MA,…Official notice →
  3. US
    Drugs & medicalClass II

    Maquet Cardiovascular, LLC recalls HEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR VASCULAR GRAFT- D: 7 mm; L: 60 cm Indicated for use in the replacement or repair of…

    Outer labelling (HEMASHIELD GOLD Knitted Microvel Double Velour Bifurcated Vascular Graft, D: 16 x 8 mm; L: 40 cm) does not correspond to the (inner) sterile packaging labelling (HEMASHIELD GOLD Knitted Microvel Double…

    FDA · DeviceMaquet Cardiovascular, LLCUS Nationwide distribution in the state of MO.Official notice →
  4. US
    Drugs & medicalClass II

    Stryker Corporation recalls 1688 Camera Control Unit (CCU), Catalog number:1688010000. Used in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and…

    A software defect in the camera control unit (CCU) will cause the image on the monitor to flip upside-down into an incorrect orientation. The potential of harms are conversion to open procedure, additional medical…

    FDA · DeviceStryker CorporationWorldwide distribution. US nationwide including Puerto Rico. OUS: BR, AE, NL,…Official notice →
  5. US
    Drugs & medicalClass II

    Heartware, Inc. recalls HeartWare Ventricular Assist Device (HVAD) System

    Updated cleaning instructions are being provided for the Controller AC Adapter, DC Adapter, and Battery to provide clarity to avoid cleaning the power source connector pins, as this could remove the lubricant that is…

    FDA · DeviceHeartware, Inc.Distribution throughout the United States. OUS distribution in Argentina,…Official notice →
  6. US
    Drugs & medicalClass II

    Philips North America Llc recalls Philips Allura Xper, UNIQ(Allura Xper system-China) Interventional Fluoroscopic X-ray System products with a Laird chiller WITHOUT an…

    Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room,…

    FDA · DevicePhilips North America LlcUS Nationwide Foreign: Argentina Austria Bangladesh Brazil Chile China Denmark…Official notice →
  7. US
    Drugs & medicalClass II

    Carl Zeiss Meditec, Inc. recalls CIRRUS HD-OCT, Models 5000 AngioPlex and 6000 AngioPlex

    Optic nerve head angiography scan to be turned off due to its distribution without pre-market clearance.

    FDA · DeviceCarl Zeiss Meditec, Inc.U.S Nationwide Distribution.: OH, CA, IN, AR, IA, LA, WI, OK, IL, OR, AL, KY,…Official notice →
See all drugs & medical recalls →