B Braun Medical Inc recalls Introcan Safety FEP 14G, 2.2x32MM-US, Product Code 4251890-02. A passive anti-needle device to provide venous or arterial access.
Potential for leakage at the catheter hub.
Read the official notice →- Date
- 2022-10-12
- Country
- United States
- Brand
- B. Braun Medical
- Product
- Introcan Safety FEP 14G, 2.2x32MM-US, Product Code 4251890-02. A passive anti-needle device to provide venous or arterial access.
- Classification
- Class II
- Units
- Domestic distribution to AZ CA FL GA MD MI MO MS NC NY OH PA TN TX WA. Foreign…
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More B. Braun Medical recalls
- 2022-12-14
US
B. Braun Medical, Inc. recalls 0.9% Sodium Chloride Injection USP 1000 mL Single-dose container Rx only NDC 0264-5802-00 B. Braun Medical Inc. Bethlehem, PA 18018-3524…
Lack of sterility assurance: Bags have the potential to leak.
- 2022-10-26
US
B. Braun Medical, Inc. recalls CARESITE SMALLBORE EXT SET T-PORT 5 IN., Product Code 470114, 100 units per carton. The A6 Access Device is a valve intended for the…
The extension set has a high probability of potential failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of…
FDA · DeviceB. Braun Medical, Inc.Distribution US nationwide, Canada and Saudi Arabia.Official notice →US
B. Braun Medical, Inc. recalls ULTRASITE US04TL Smallbore T-Port Extension Set, Product Code 474515, 100 units per carton. to provide needle-free access on I.V. and…
The extension set has a high probability of failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of bloodstream…
FDA · DeviceB. Braun Medical, Inc.Distribution US nationwide, Canada and Saudi Arabia.Official notice →- 2022-04-20
US
B Braun Medical Inc recalls Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (END) - VERSION 2, Article Code Number…
Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration…
- 2022-03-30
US
B Braun Medical Inc recalls PERFUSOR SPACE - US VERSION. Intended Use: syringe infusion pump that can be used to configure customized, tailor-made solutions Catalog…
Units not properly inspected prior to release to the market, may not function as indicated in the product labeling, resulting in syringe detection and infusion accuracy that can result in delays in therapy,…
- 2022-03-23
US
B. Braun Medical, Inc. recalls 0.9% Sodium Chloride Injection USP, 250 mL Excel Container, Rx only, B. Braun Medical Inc., Bethlehem, PA, NDC 0264-7800-20
Lack of sterility assurance: leaking bags
Other drugs & medical recalls in the United States
- 2022-10-19
US
Justified Laboratories recalls Skincell Mole & Skin Corrector Serum, 1 fl. oz./30 mL jars, Distributed By: Justified Laboratoires, Arlington, TX 76011 USA UPC 8 10077…
Marketed Without An Approved NDA/ANDA; Unapproved new drug
- 2022-10-12
US
Hologic, Inc recalls The BioZorb LP Marker is an implantable radiopaque marker used to facilitate visualization of a soft tissue site. The BioZorb LP Marker is…
Hologic BioZorb LP Marker, (Product number F0221), lot number 22A0RL, contained removable labels with an expiration date that indicated the product had already expired. The expiration date on the removable label is…
FDA · DeviceHologic, IncUS Nationwide distribution including in the states of FL, IL, MD, NJ, AZ, TX,…Official notice →US
Philips Respironics, Inc. recalls Wisp and Wisp Youth Nasal Mask
No contraindication/warning regarding magnetic components in CPAP masks for patients with implantable devices or metallic splinters in the eyes.
FDA · DevicePhilips Respironics, Inc.Worldwide distribution - US Nationwide distribution including in the states of…Official notice →US
Synthes (USA) Products LLC recalls Hammertoe Continuous Compression Implant (CCI) Kit Part Number: 46.239.001
Drill Pin (drill bit) is not compatible with the Drill Template included in the kit due to a packaging error, result in a surgical delay and may cause soft tissue damage if a larger incision is required to accommodate…
US
Zimmer, Inc. recalls G7 Acetabular System, Acetabular Shell, Item numbers Item Numbers: 010000663 (52 MM E), 010000667 (60 MM G). used in hip arthroplasty
The outer sterile package cavity has a corner wall thickness that is below the specification. The thin corner wall could potentially crack during transit. Potential risks include non-clinically or clinically…
US
Wright Medical Technology, Inc. recalls stryker EVOLVE, Proline Stem, CoCr, Size: +4, OD: 9.5mm, REF 496S275, Wright s EVOLVE Proline Plus Radial Head and Repair System…
The outer label does not match the device inside the package. The affected lot of EVOLVE Proline Stem 7.5mm +2 contains a size 9.5mm +4 stem, while the affected lot of EVOLVE Proline Stem 9.5mm +4 contains a size 7.5mm…