American Contract Systems, Inc. recalls Pacemaker, COPM11B; Medical convenience kit

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance…

Read the official notice →
Date
2024-10-02
Country
United States
Brand
American Contract Systems
Product
Pacemaker, COPM11B; Medical convenience kit
Classification
Class II
Units
US Nationwide distribution in the states of MO, MN, MA, OH, NE.
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More American Contract Systems recalls

  1. 2025-04-02
  2. US
    Drugs & medicalClass II

    American Contract Systems, Inc. recalls Brand Name: Banner University Med Ctr. Product Name: Cardiac Cath Lab Pack-170278 Model/Catalog Number: GSCC40G Software…

    Sterility assurance with procedure trays

    FDA · DeviceAmerican Contract Systems, Inc.US: AZ OUS: NoneOfficial notice →
  3. 2025-03-26
  4. US
    Drugs & medicalClass II

    American Contract Systems Inc recalls CRANI PACK, Model/Catalog Number WECR41D, medical procedure kit.

    During an internal investigation, ACS was unable to confirm that the CODMAN component achieved the required Sterility Assurance Level and therefore is not suitable for sterilization according to ACS procedures.

    FDA · DeviceAmerican Contract Systems IncUS: MIOfficial notice →
  5. US
    Drugs & medicalClass II

    American Contract Systems Inc recalls MAJOR VASCULAR PACK , Model No UTMV78X-01 UTMV78Y UTMV78Y-01 UTMV78X-01 UTMV78Y-02

    ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

    FDA · DeviceAmerican Contract Systems IncUS Nationwide distribution in the state of OH.Official notice →
  6. 2024-08-14
  7. US
    Drugs & medicalClass II

    American Contract Systems, Inc. recalls American Contract Systems Spine Pack convenience kit

    Product released with insufficient aeration time impacting the kits sterilized on the same cycle or in the same aeration chamber

    FDA · DeviceAmerican Contract Systems, Inc.US Nationwide distribution in the states of CA, KS, ME, MN, MO, NE, and SD.Official notice →
  8. 2024-07-17
  9. US
    Drugs & medicalClass II

    American Contract Systems Inc recalls TOTAL KNEE - 302497- Procedure tray Catalog Number: ANTK12AF

    Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

    FDA · DeviceAmerican Contract Systems IncUS Nationwide distribution.Official notice →
  10. 2024-02-07
  11. US
    Drugs & medicalClass II

    American Contract Systems, Inc. recalls Procedural convenience kits and trays, cardiac, labeled as: a) EP LAB BI V IMPLANT PACK, kit number AGBI64J; b) LAB DEVICE…

    During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked"…

    FDA · DeviceAmerican Contract Systems, Inc.US Nationwide distribution.Official notice →
All 22 American Contract Systems recalls →

Other drugs & medical recalls in the United States

  1. 2024-10-02
  2. US
    Drugs & medicalClass II

    Hologic Inc recalls Selenia Dimensions/ 3Dimensions-Digital breast tomosynthesis Software Version Dimensions 1.12.0 and 3Dimensions 2.3.0 system software as…

    Identified an issue with Selenia Dimensions 1.12.0 and 3Dimensions 2.3.0 system software when used in combination with C-View software in which the quality of the final C-View synthesized 2D image is impacted and may…

    FDA · DeviceHologic IncWorldwide distribution - US Nationwide and the countries of Belgium, Canada,…Official notice →
  3. US
    Drugs & medicalClass II

    Becton Dickinson & Co. recalls BD Phoenix M50 instrument -Intended for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically…

    Escherichia coli (E. coli) misidentification for clinical samples that were tested on BD Phoenix" M50 instrument and BD Phoenix" 100, may lead to misdiagnosis and inappropriate treatment of infections caused by E. coli,

    FDA · DeviceBecton Dickinson & Co.Worldwide distribution - US Nationwide and the countries of Algeria,…Official notice →
  4. US
    Drugs & medicalClass II

    Waldemar Link GmbH & Co. KG (Corp. Hq.) recalls LINK Universal Handle, with quick coupling, Stainless Steel, straight, Item Number 130-394/01

    Some lots of the LINK Universal Handle, with quick coupling, Stainless Steel, straight may suffer an immediate material failure intraoperatively when the instrument is used for the first time. The instrument can break…

    FDA · DeviceWaldemar Link GmbH & Co. KG (Corp. Hq.)US Nationwide distribution in the states of FL, KS, MI, NJ, VA.Official notice →
  5. US
    Drugs & medicalClass II

    Karl Storz Endoscopy recalls Karl Storz - Endoskope , REF: 27092AMA, Hopkins Telescope 6, Rx only, CE 0123

    Various medical device product IFUs contain reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA.

    FDA · DeviceKarl Storz EndoscopyUS Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT,…Official notice →
  6. US
    Drugs & medicalClass II

    Siemens Medical Solutions USA, Inc. recalls ACUSON Juniper 1.0, 1.5, 2.0, 2.5 Diagnostic Ultrasound Systems, REF: 11335791, with software. ACUSON Juniper Select 2.0,…

    If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion…

    FDA · DeviceSiemens Medical Solutions USA, Inc.US: MS, WA, TX, NJ, GA, PN, NY, AL, MN, NE, IN, CT, CA, IL, TN, WI, ND, NM,…Official notice →
  7. US
    Drugs & medicalClass II

    GE Healthcare recalls GE Healthcare Centricty PACS, V2, V3, V4, V6, V7; Radiological imaging processing system.

    GE HealthCare has become aware of an issue where image storage is not possible when Centricity PACS reaches the maximum number of unique image identifiers (Ckeys). If this occurs, it can result in acquisition of a…

    FDA · DeviceGE HealthcareWorldwide distribution.Official notice →
See all drugs & medical recalls →