American Contract Systems Inc recalls CRANI PACK, Model/Catalog Number WECR41D, medical procedure kit.
During an internal investigation, ACS was unable to confirm that the CODMAN component achieved the required Sterility Assurance Level and therefore is not suitable for sterilization according to ACS procedures.
Read the official notice →- Date
- 2025-03-26
- Country
- United States
- Brand
- American Contract Systems
- Product
- CRANI PACK, Model/Catalog Number WECR41D, medical procedure kit.
- Classification
- Class II
- Units
- US: MI
- Category
- Drugs & medical
- Source
- FDA · Device
Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer
What to do
- Check the product name, strength and lot number against the official notice.
- Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
- Ask the pharmacy or supplier about a replacement or refund.
General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.
More American Contract Systems recalls
- 2025-09-03
US
American Contract Systems Inc recalls AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) ANTERIOR HIP PACK, MODEL/ITEM NUMBER AMAH40N; 2)TOTAL…
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS…
- 2025-06-18
US
American Contract Systems, Inc. recalls LOWER EXTREMITY, Model: NCEX20I; LOWER EXTREMITY, Model: NCEX20J; TOTAL KNEE PACK, Model: NCTK31K; TOTAL KNEE PACK, Model:…
Test results for cast padding component included in procedure trays shows Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals exceeded residual limits, which may lead to a possible loss or lack of functionality…
FDA · DeviceAmerican Contract Systems, Inc.US Nationwide distribution in the state of CO.Official notice →- 2025-04-02
US
- 2025-03-26
US
American Contract Systems Inc recalls MAJOR VASCULAR PACK , Model No UTMV78X-01 UTMV78Y UTMV78Y-01 UTMV78X-01 UTMV78Y-02
ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.
FDA · DeviceAmerican Contract Systems IncUS Nationwide distribution in the state of OH.Official notice →- 2024-10-02
US
American Contract Systems, Inc. recalls Pacemaker, COPM11B; Medical convenience kit
ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance…
FDA · DeviceAmerican Contract Systems, Inc.US Nationwide distribution in the states of MO, MN, MA, OH, NE.Official notice →- 2024-08-14
US
American Contract Systems, Inc. recalls American Contract Systems Spine Pack convenience kit
Product released with insufficient aeration time impacting the kits sterilized on the same cycle or in the same aeration chamber
FDA · DeviceAmerican Contract Systems, Inc.US Nationwide distribution in the states of CA, KS, ME, MN, MO, NE, and SD.Official notice →
Other drugs & medical recalls in the United States
- 2025-04-02
US
Strides Pharma, Inc. recalls Testosterone Gel 1%, 5 grams, 30 Unit-dose Packets in a carton, Rx only, Distributed by: Strides Pharma Inc., East Brunswick, NJ, NDC…
Presence of foreign substance: Presence of Benzene.
US
- 2025-03-26
US
FUJIFILM Healthcare Americas Corporation recalls Synapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200.
The incorrect computed patient age is showing in VX for patients less than 3 months old.
FDA · DeviceFUJIFILM Healthcare Americas CorporationDomestic: AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, KY, LA, MA, MN, MO, MS,…Official notice →US
Smiths Medical ASD, Inc. recalls CAPNOCHECK BCI Airway Adapter; Model Number BSR; Product Code WW1100
Airway adapter does not meet specification and may break off when attaching a valve or tube. Device failure may result in inaccurate readings, a break in airway circuit, and loss of ventilation. A break in the airway…
FDA · DeviceSmiths Medical ASD, Inc.US Nationwide distribution. International distribution pending.Official notice →US
ZIMMER ORTHOPEDICS MANFACTURING LIMITED-GALWAY recalls Comprehensive Shoulder Stem, Comprehensive Micro Stem, Comprehensive Mini Stem; uncemented shoulder prosthesis,…
Five complaints received where surgeon was unable to remove and/or connect the inserter with the stem during surgery due to excess material remaining within the taper. Issue may lead to clinically significant extension…
FDA · DeviceZIMMER ORTHOPEDICS MANFACTURING LIMITED-GALWAYUS distribution nationwide. International distribution to Dominican Republic,…Official notice →US
Merge Healthcare, Inc. recalls VERICIS, Merge Cardio, Model/Catalog Number: Version 9.x, Software Version: 9.0.6 and 9.0.8; VERICIS is an integrated cardiovascular…
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements…