Drugs & medicalClass II

Alcon Research, LLC recalls Alcon Custom Pak Surgical Procedure Pack: 16998-11 2519351H SALAHUDDIN 2.4 PACK FRESNO SURGICAL HOSP 18392-03 14HDFF CATARACT SWAGEL…

Due to latex gloves being include in Custom Pak. Kits are supposed to include non-latex gloves.

Read the official notice →
Date
2022-07-20
Country
United States
Brand
Alcon Research
Product
Alcon Custom Pak Surgical Procedure Pack: 16998-11 2519351H SALAHUDDIN 2.4 PACK FRESNO SURGICAL HOSP 18392-03 14HDFF CATARACT SWAGEL…
Classification
Class II
Units
U.S.: AZ, CA, MO, and NY O.U.S.: None
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Alcon Research recalls

  1. 2024-04-03
  2. US
    Drugs & medicalClass II

    Alcon Research, LLC recalls SYRINGE,BULB,2OZ/60CC,GREEN, Part: 300055070, packaged inside Alcon Custom-Pak Surgical Procedure Packs, Products: 18018-06, 3865-44,…

    Medline Bulb irrigation syringes recalled due to potential for packaging sterile barrier to be breached, and these syringes were packed inside surgical procedure packs that were subsequently terminally sterilized, so…

    FDA · DeviceAlcon Research, LLCUS Nationwide distribution in the states of OH, WV.Official notice →
  3. 2023-11-15
  4. US
    Drugs & medicalClass II

    Alcon Research, LLC recalls Alcon Custom-Pak, Surgical Procedure Pack REF 14057-09, 45K W/DUOVISC CATARACT, N VIRGINIA EYE SURG CTR

    Custom surgical procedure packs contain latex gloves but are labeled as "latex free".

    FDA · DeviceAlcon Research, LLCUS: VA OUS: NoneOfficial notice →
  5. 2023-10-18
  6. US
    Drugs & medicalClass II

    Alcon Research, LLC recalls Constellation Vision System, REF: 8065751150

    Ophthalmic microsurgical vision systems have a new Printed Circuit Board (PCB) design that could lead to higher voltage to a PCB component, that could cause component failure and loss of pressure control for the use of…

    FDA · DeviceAlcon Research, LLCUS: GA, IL, PROfficial notice →
  7. 2023-08-23
  8. US
    Drugs & medicalClass II

    Alcon Research, LTD. recalls Centurion FMS Pak

    Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.

    FDA · DeviceAlcon Research, LTD.Worldwide distribution - US Nationwide and the countries of Aland Islands,…Official notice →
  9. 2023-04-05
  10. US
    Drugs & medicalClass II

    Alcon Research, LLC recalls Alcon Custom Pak Surgical Procedure Pack: Pak Number & Pak Description: 17533-04 DR GROSS LPS LLC 18704-03 CATARACT SURGI CARE 18382-03 DR…

    Sterility; Steril surgical procedure packs contain incomplete seals.

    FDA · DeviceAlcon Research, LLCUS: AL, AR, CA, IA, ID, IN, KY, MN, NC, ND, NJ, OH, PA, TX, OUS: NoneOfficial notice →
  11. 2022-12-07
  12. US
    Drugs & medicalClass II

    Alcon Research, LLC recalls Alcon Custom Pak Surgical Procedure Pack

    Due to difficulty removing the liner on the adhesive components of the affected drapes and as a result may render the product unusable. In addition, the manufacturer of the drapes has observed an increase in reported…

    FDA · DeviceAlcon Research, LLCWorldwide Distribution: U.S.(nationwide) to states of: AK, AL, AR, AZ, CA, CO,…Official notice →
All 22 Alcon Research recalls →

Other drugs & medical recalls in the United States

  1. 2022-07-20
  2. US
    Drugs & medicalClass II

    Waldemar Link GmbH & Co. KG (Mfg Site) recalls Tibial Component Modular. orthopedic prosthesis. Model Nos: Small, W 16-2817/32 Medium, W 16-2817/35 Large, W 16-2817/37

    There is a risk that blind screws of the modular tibial component cannot be loosened intraoperatively. This may lead to prolongation of surgery due to an intraoperative change in procedure.

    FDA · DeviceWaldemar Link GmbH & Co. KG (Mfg Site)Domestic distribution to AL, CA, FL, GA, IL, IN, KS, MI, MN, NV, NY, OH, TN,…Official notice →
  3. US
    Drugs & medicalClass II

    DePuy Orthopaedics, Inc. recalls MODULAR CATHCART FRACTURE HEAD HIP BALL, sizes 41mm to 60mm inclusive; metallic femoral head prosthesis, Product No. 136341000,…

    The Cathcart Ball has an inherent +5mm offset which is not currently specified in its Instructions for Use or Surgical Technique Guide. The STG has been updated to reflect the clarification that the +5mm offset should…

    FDA · DeviceDePuy Orthopaedics, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  4. US
    Drugs & medicalClass II

    Ventana Medical Systems Inc recalls Ventana HE 600 System, automated slide preparer, for use in laboratories.

    There is a potential for fluid leak inside the Ventana HE 600 instrument which could result in electrical short and fire

    FDA · DeviceVentana Medical Systems IncUS: AL AR AZ CA CO CT DC FL GA HI IA IL IN KS KY MA MD MI MN MO MS MT NC NE NJ…Official notice →
  5. US
    Drugs & medicalClass II

    Siemens Healthcare Diagnostics Inc recalls epoc BGEM Test Card 25 Test Cards/box. In vitro diagnostic device for the quantitative testing of samples of heparinized or…

    Discrepant (low bias) pO2 results, could be an incorrect diagnosis of hypoxemia and initiation of unnecessary or inappropriate respiratory support.

    FDA · DeviceSiemens Healthcare Diagnostics IncAL CA CO FL IA ID KS NM OK TXOfficial notice →
  6. US
    Drugs & medicalClass II

    Fresenius Medical Care Holdings, Inc. recalls NATURALYTE Liquid Acid 3.0K 3.0Ca- indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for…

    Alternate Bottle Caps labeled with "Sealed for your Protection" on the foil not qualified for use

    FDA · DeviceFresenius Medical Care Holdings, Inc.NatiowideOfficial notice →
  7. US
    Drugs & medicalClass III

    Invacare Corporation recalls Invacare Platinum 5NXG Oxygen Concentrator, Model Numbers P5NXG1, P5NXGC1 (Canada specific model)

    Between March 2, 2021 and December 31, 2021, the product label used in production did not contain the "Rx Only" statement as required.

    FDA · DeviceInvacare CorporationDomestic distribution to the following states: ALASKA ARIZONA CALIFORNIA…Official notice →
See all drugs & medical recalls →