Drugs & medicalClass II

Alcon Research, LLC recalls Alcon Custom-Pak, Surgical Procedure Pack REF 14057-09, 45K W/DUOVISC CATARACT, N VIRGINIA EYE SURG CTR

Custom surgical procedure packs contain latex gloves but are labeled as "latex free".

Read the official notice →
Date
2023-11-15
Country
United States
Brand
Alcon Research
Product
Alcon Custom-Pak, Surgical Procedure Pack REF 14057-09, 45K W/DUOVISC CATARACT, N VIRGINIA EYE SURG CTR
Classification
Class II
Units
US: VA OUS: None
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Alcon Research recalls

  1. 2024-06-05
  2. US
    Drugs & medicalClass II

    Alcon Research LLC recalls ClearCut S Safety Sideport Knife 1.2mm Dual Bevel Model/Catalog Number: 8065771541 Software Version: N/A Product Description: Ophthalmic…

    Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Corneal/Scleral V-Lance Ophthalmic knives.

    FDA · DeviceAlcon Research LLCWorldwide distribution - US Nationwide and the countries of Argentina,…Official notice →
  3. 2024-05-01
  4. US
    Drugs & medicalClass II

    Alcon Research, LLC recalls Various Alcon Custom Paks Surgical Procedure Pack containing Oasis 27Ga Chang Hydrodissection Cannulas. These Custom Paks are surgical…

    Due to complaints of particulates, residue, occlusions, and bent tips for sterile cannulas.

    FDA · DeviceAlcon Research, LLCWorldwide distribution - U.S. Nationwide including in the states of AK, AR,…Official notice →
  5. 2024-04-03
  6. US
    Drugs & medicalClass II

    Alcon Research, LLC recalls SYRINGE,BULB,2OZ/60CC,GREEN, Part: 300055070, packaged inside Alcon Custom-Pak Surgical Procedure Packs, Products: 18018-06, 3865-44,…

    Medline Bulb irrigation syringes recalled due to potential for packaging sterile barrier to be breached, and these syringes were packed inside surgical procedure packs that were subsequently terminally sterilized, so…

    FDA · DeviceAlcon Research, LLCUS Nationwide distribution in the states of OH, WV.Official notice →
  7. 2023-10-18
  8. US
    Drugs & medicalClass II

    Alcon Research, LLC recalls Constellation Vision System, REF: 8065751150

    Ophthalmic microsurgical vision systems have a new Printed Circuit Board (PCB) design that could lead to higher voltage to a PCB component, that could cause component failure and loss of pressure control for the use of…

    FDA · DeviceAlcon Research, LLCUS: GA, IL, PROfficial notice →
  9. 2023-08-23
  10. US
    Drugs & medicalClass II

    Alcon Research, LTD. recalls Centurion FMS Pak

    Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.

    FDA · DeviceAlcon Research, LTD.Worldwide distribution - US Nationwide and the countries of Aland Islands,…Official notice →
  11. 2023-04-05
  12. US
    Drugs & medicalClass II

    Alcon Research, LLC recalls Alcon Custom Pak Surgical Procedure Pack: Pak Number & Pak Description: 17533-04 DR GROSS LPS LLC 18704-03 CATARACT SURGI CARE 18382-03 DR…

    Sterility; Steril surgical procedure packs contain incomplete seals.

    FDA · DeviceAlcon Research, LLCUS: AL, AR, CA, IA, ID, IN, KY, MN, NC, ND, NJ, OH, PA, TX, OUS: NoneOfficial notice →
All 22 Alcon Research recalls →

Other drugs & medical recalls in the United States

  1. 2023-11-15
  2. US
    Drugs & medicalClass II

    Encore Medical, LP recalls RSP Humeral Socket Insert Size 36mm Semi-Constrained +4 mm, REF: 509-01-436, orthopedic shoulder implant component.

    Two different humeral socket Insert devices were swapped during packaging, which could result in surgeons not having the correct device for surgery, which could cause a delay in surgery.

    FDA · DeviceEncore Medical, LPUS: MN, TX, IN, OH, SC, RI, ME, NJ, NY, KS, LA, FL, CAOfficial notice →
  3. US
    Drugs & medicalClass II

    Philips North America Llc recalls DigitalDiagnost 4 Chest/Emergency, used to acquire, process, store, display and export digital radiographic images.

    Potential for units suspended on the ceiling to fail and fall.

    FDA · DevicePhilips North America LlcNationwide distribution.Official notice →
  4. US
    Drugs & medicalClass II

    Family Dollar Stores, Llc. recalls MEDIC's CHOICE CONTACT LENS SOLUTION 12FLOZ, SKU 998935

    Products were stored outside of labeled temperature requirements.

    FDA · DeviceFamily Dollar Stores, Llc.Distributed to Family Dollar stores in the following states: GA, AL, MS, FL,…Official notice →
  5. US
    Drugs & medicalClass II

    Percussionaire Corporation recalls Monitron II Waveform Analyzer, REF: S00015-1, used with ventilators.

    Note this recall occurred in 2020 and 2021. Reports of screen freeze on visual displays of waveform analyzers that are paired with ventilators. Subsequently, waveform analyzers were serviced to install PAL…

    FDA · DevicePercussionaire CorporationUS: CA, ND, MS, MDOfficial notice →
  6. US
    Drugs & medicalClass II

    Encore Medical, LP recalls DJO EMPOWR Knee Punch Handle, REF: 801-05-040

    Knee punch handle tip is breaking during surgery due to the lack of heat treatment, which may lead to a disruption in surgery.

    FDA · DeviceEncore Medical, LPUS: VA, TN, MN, TX, IL, IN, KY, OH, SC, MS, RI, ME, NJ, NY, MD, CA, KS, LA,…Official notice →
  7. US
    Drugs & medicalClass II

    Stryker Corporation recalls Disposable Spetzler-Malis Bipolar forceps, 23CM SM DISP BAYONET 1.5MM TIP, Catalog Number 6760-230-015

    An error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.

    FDA · DeviceStryker CorporationUS Nationwide. Australia, Canada, Japan, Korea, Singapore.Official notice →
See all drugs & medical recalls →