Drugs & medicalClass I

Abiomed, Inc. recalls Automated Impella Controller (AIC), for use with left heart support blood pump. Labeled as the following with corresponding Product Codes:…

Potential software error in the Automated Impella Controller ("AIC" ) when used in conjunction with left ventricular Impella devices.

Read the official notice →
Date
2026-06-24
Country
United States
Brand
Abiomed
Product
Automated Impella Controller (AIC), for use with left heart support blood pump. Labeled as the following with corresponding Product Codes:…
Classification
Class I
Units
Worldwide distribution - US Nationwide and the countries of Australia,…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Abiomed recalls

  1. 2026-07-22
  2. US
    Drugs & medicalClass I

    Abiomed, Inc. recalls ABIOMED Impella CP Set, Pump Set Code 0048-0003, containing BIOMED Impella Introducer, 14 Fr, 13 cm & 25 cm, Product Code 0052-3046. heart…

    A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr…

    FDA · DeviceAbiomed, Inc.Worldwide distribution. US Nationwide, Australia (AU), Austria (AT), Belgium…Official notice →
  3. 2026-07-01
  4. US
    Drugs & medicalClass I

    Abiomed, Inc. recalls Abiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer Kits. Product Codes: 1000413,…

    Potential for thrombus formation during prolonged use of the introducer.

    FDA · DeviceAbiomed, Inc.Worldwide - US Nationwide distribution in the states of FL, IL, MN, NJ, NY, VA…Official notice →
  5. 2026-06-24
  6. US
    Drugs & medicalClass I

    Abiomed, Inc. recalls Impella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP.

    Exposure to the low purge pressure occurrence may result in persistent low purge pressure alarms and, in some cases, interruption or loss of mechanical circulatory support.

    FDA · DeviceAbiomed, Inc.US distribution to GA, IA, NE, PA, TX, WA, WI. International distribution to…Official notice →
  7. 2026-06-03
  8. US
    Drugs & medicalClass I

    Abiomed, Inc. recalls Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged,…

    Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper…

    FDA · DeviceAbiomed, Inc.Worldwide distribution - US Nationwide and the countries of Canada, Denmark,…Official notice →
  9. US
    Drugs & medicalClass I

    Abiomed, Inc. recalls Automated Impella Controller (AIC); Product Code: 0042-0000-US;

    A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process;…

    FDA · DeviceAbiomed, Inc.Worldwide - US Nationwide distribution in the states of CA, CT, FL, LA, MA,…Official notice →
  10. 2026-04-08
  11. US
    Drugs & medicalClass I

    Abiomed, Inc. recalls Automated Impella Controller (AIC), labeled as the following with corresponding Product Codes: 1. Impella Controller, Packaged, CA;…

    Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.

    FDA · DeviceAbiomed, Inc.International distribution in the country of Australia, Canada, France,…Official notice →
All 30 Abiomed recalls →

Other drugs & medical recalls in the United States

  1. 2026-06-24
  2. US
    Drugs & medicalClass II

    Medline Industries, LP recalls Medline convenience kits: CMH CENTRAL LINE DRESSING CHANGE KIT EBSI1705A CVC/PICC/MIDLINE DRESSING CHANGE KIT EBSI1057A DIALYSIS DRESSING…

    Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.

    FDA · DeviceMedline Industries, LPUS Nationwide distribution. OUS distribution pending.Official notice →
  3. US
    Drugs & medicalClass II

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips MultiDiagnost Eleva with the following Model Numbers: Model # 708032 for MultiDiagnost Eleva; Model # 708034 for…

    Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection

    FDA · DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.US and ROW.Official notice →
  4. US
    Drugs & medicalClass I

    ARROW INTERNATIONAL, LLC recalls Description/REF: PI JACC KIT: 2L 5.5 FR X 15 CM CG+/ASK-41552-LTAC; PI JACC KIT: 3L 6 FR X 20 CM CG+/ASK-42063-LTAC; AGBA PI JACC KIT:…

    Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if…

    FDA · DeviceARROW INTERNATIONAL, LLCUS Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN,…Official notice →
  5. US
    Drugs & medicalClass I

    Datex-Ohmeda Inc recalls Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator) Product identifiers: 2082663-001,…

    Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting…

    FDA · DeviceDatex-Ohmeda IncPendingOfficial notice →
  6. US
    Drugs & medicalClass II

    Dabur India Limited recalls Med Pride, HYDROCORTISONE CREAM 1%, Net Wt. 1 oz (28.3g) tubes, Manufactured for: Shield Line LLC, 59 Hook Road, Bayonne, NJ 07002 USA,…

    CGMP Deviations; deficiencies observed during FDA inspection

    FDA · DrugDabur India LimitedUS Nationwide.Official notice →
  7. US
See all drugs & medical recalls →