Drugs & medicalClass I

Abiomed, Inc. recalls Impella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP.

Exposure to the low purge pressure occurrence may result in persistent low purge pressure alarms and, in some cases, interruption or loss of mechanical circulatory support.

Read the official notice →
Date
2026-06-24
Country
United States
Brand
Abiomed
Product
Impella CP Set with SmartAssist. Product Codes: 1000080, 1000851, 0048-0014, 0048-0024-JP.
Classification
Class I
Units
US distribution to GA, IA, NE, PA, TX, WA, WI. International distribution to…
Category
Drugs & medical
Source
FDA · Device

Copied automatically from the US Food and Drug Administration. Details may be shortened or out of date; the official notice is the authoritative source. Disclaimer

What to do

  1. Check the product name, strength and lot number against the official notice.
  2. Do not stop a prescribed medicine without talking to your doctor or pharmacist first.
  3. Ask the pharmacy or supplier about a replacement or refund.

General guidance from Recall Atlas. The official notice from the US Food and Drug Administration has the exact models, batch codes and remedy, and it takes priority.

More Abiomed recalls

  1. 2026-09-23
  2. US
    Drugs & medicalClass I

    Abiomed, Inc. recalls Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged,…

    The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.

    FDA · DeviceAbiomed, Inc.Domestic: Nationwide Distribution. International: United Arab Emirates,…Official notice →
  3. 2026-07-22
  4. US
    Drugs & medicalClass I

    Abiomed, Inc. recalls ABIOMED Impella CP Set, Pump Set Code 0048-0003, containing BIOMED Impella Introducer, 14 Fr, 13 cm & 25 cm, Product Code 0052-3046. heart…

    A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr…

    FDA · DeviceAbiomed, Inc.Worldwide distribution. US Nationwide, Australia (AU), Austria (AT), Belgium…Official notice →
  5. 2026-07-01
  6. US
    Drugs & medicalClass I

    Abiomed, Inc. recalls Abiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer Kits. Product Codes: 1000413,…

    Potential for thrombus formation during prolonged use of the introducer.

    FDA · DeviceAbiomed, Inc.Worldwide - US Nationwide distribution in the states of FL, IL, MN, NJ, NY, VA…Official notice →
  7. 2026-06-24
  8. US
    Drugs & medicalClass I

    Abiomed, Inc. recalls Automated Impella Controller (AIC), for use with left heart support blood pump. Labeled as the following with corresponding Product Codes:…

    Potential software error in the Automated Impella Controller ("AIC" ) when used in conjunction with left ventricular Impella devices.

    FDA · DeviceAbiomed, Inc.Worldwide distribution - US Nationwide and the countries of Australia,…Official notice →
  9. 2026-06-03
  10. US
    Drugs & medicalClass I

    Abiomed, Inc. recalls Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged,…

    Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper…

    FDA · DeviceAbiomed, Inc.Worldwide distribution - US Nationwide and the countries of Canada, Denmark,…Official notice →
  11. US
    Drugs & medicalClass I

    Abiomed, Inc. recalls Automated Impella Controller (AIC); Product Code: 0042-0000-US;

    A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process;…

    FDA · DeviceAbiomed, Inc.Worldwide - US Nationwide distribution in the states of CA, CT, FL, LA, MA,…Official notice →
All 30 Abiomed recalls →

Other drugs & medical recalls in the United States

  1. 2026-06-24
  2. US
    Drugs & medicalClass II

    Howmedica Osteonics Corp. recalls TRI TS BASEPLATE SIZE 3. Part Number: 5521-B-300.

    Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot…

    FDA · DeviceHowmedica Osteonics Corp.US Nationwide distribution in the states of AZ, CA, GA, MI, NC, NJ, NY, WV.Official notice →
  3. US
    Drugs & medicalClass II

    Accriva Diagnostics, Inc. recalls directCHECK ACT-LR whole blood quality control, level 2 Model/Catalog Number: DCGLR-2

    Assayed Whole blood control contains labeling with incorrect performance range.

    FDA · DeviceAccriva Diagnostics, Inc.Worldwide - US Nationwide distribution including in the states of AL, AR, AZ,…Official notice →
  4. US
    Drugs & medicalClass II

    AMD Medicom Inc. recalls Ritmed Neurological Sponges, Description/REF: 0.5x0.5in, Non-Sterile/55992C; 0.5x1.0in, Non-Sterile/55993C; 0.5x3.0in, Non-Sterile/55994C;…

    Neurological sponges used to moisten or absorb biological fluid during neurological surgery and may also be used to protect the brain during incision, may not consistently meet endotoxin specification for devices that…

    FDA · DeviceAMD Medicom Inc.US: MT, NC, VA, CA, TX, OH, TN, IL, RIOfficial notice →
  5. US
    Drugs & medicalClass II

    Smith & Nephew, Inc. recalls TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816

    Humeral nail due to a manufacturing error is missing distal holes and if implanted may lead to revision surgery but, if the missing holes are identified before implantation, an equivalent backup can be used, or a bent…

    FDA · DeviceSmith & Nephew, Inc.International distribution to the countries of BE, DE, IT.Official notice →
  6. US
    Drugs & medicalClass II

    Medtronic Navigation, Inc.-Boxborough recalls Medtronic O-arm O2 Imaging System. Model Number: BI70002000.

    Potential for image artifacts caused by an anomaly in the O-arm O2 Imaging System s detector panel firmware.

    FDA · DeviceMedtronic Navigation, Inc.-BoxboroughWorldwide distribution - US Nationwide and the countries of Albania, Algeria,…Official notice →
  7. US
    Drugs & medicalClass II

    Steris Corporation recalls HarmonyAIR A-Series Surgical Lighting System

    The the plus and minus buttons (+/-) used to control light intensity on the light heads may detach.

    FDA · DeviceSteris CorporationWorldwide - US Nationwide distribution.Official notice →
See all drugs & medical recalls →