Zap Surgical Systems recalls

2 official recalls and safety alerts involving Zap Surgical Systems since 2022-09, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.

This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Zap Surgical Systems or the relevant agency.

  1. 2025-10-08
  2. US
    Drugs & medicalClass II

    Zap Surgical Systems recalls ZAP-X Radiosurgery System, Model: 300150

    If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software…

    FDA · DeviceZap Surgical SystemsWorldwide - US Nationwide distribution in the states of FL, CO, NJ, NY and the…Official notice →
  3. 2022-09-21
  4. US
    Drugs & medicalClass II

    Zap Surgical Systems recalls ZAP-X Radiosurgery System

    Software issue identified in cases of initial patient setup with large (greater-than-or-equal-to 1.5 degrees) rotational deviations, between digitally reconstructed radiographs, and pre-delivery X-ray images. In these…

    FDA · DeviceZap Surgical SystemsWorldwide distribution - US Nationwide distribution in the states of AZ, MD,…Official notice →