Think Surgical recalls
2 official recalls and safety alerts involving Think Surgical since 2024-03, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.
This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Think Surgical or the relevant agency.
- 2026-07-15
US
Think Surgical, Inc. recalls TMINI Femur Tracking Array REF 109003 UDI-DI code: M9461090030 TMINI Tibia Tracking Array REF 109004 UDI-DI code: M9461090040 TMINI…
Due to manufacturing/design issue where the knob on the tracking array may loosen causing potential misalignment of the array.
FDA · DeviceThink Surgical, Inc.U.S. Nationwide distribution in the states of AZ, CA, DE, FL, IA, KS, LA, MI,…Official notice →- 2024-03-27
US
Think Surgical, Inc. recalls Assembly, Array Guide TMINI, REF: 109101, part of the TMINI Knee Instrument Set, used with the TMINI Miniature Robotic System
There is a galling effect (cold-weld) caused by friction between the array fixation pins and the TMINI array guide holes due to insufficient clearance between the outer diameter of the fixation pins and the inner…
FDA · DeviceThink Surgical, Inc.US Nationwide distribution in the states of AL, OH, NV, TX, UT, CA, MO.Official notice →