- 2024-12-04
US
Shimadzu Corp. Analytical & Measuring Instruments Div. recalls Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL
There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and…
FDA · DeviceShimadzu Corp. Analytical & Measuring Instruments Div.Worldwide - US Nationwide distribution in the states of DE, PA and the country…Official notice →US
Shimadzu Corp. Analytical & Measuring Instruments Div. recalls Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL
There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and…