Remel Europe recalls

2 official recalls and safety alerts involving Remel Europe since 2026-04, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.

This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Remel Europe or the relevant agency.

  1. 2026-08-19
  2. US
    Drugs & medicalClass II

    Remel Europe Ltd. recalls Thermo Fisher Diagnostics Neisseria meningitidis Gp C 2ml/VL, Product Number R30166901; invitro diagnostics

    Certain lots of Neisseria meningitidis Gp C 2ml/VL, R30166901, are not performing to IFU criteria not positive to ATCC13102 within 60 Seconds.

    FDA · DeviceRemel Europe Ltd.US Nationwide distribution.Official notice →
  3. 2026-04-29
  4. US
    Drugs & medicalClass II

    Remel Europe Ltd. recalls Oxoid Agglutinating Sera, Salmonella 9-0 R30957301

    Affected lots are not performing to IFU criteria showing cross reactions with 2-0 within 60 seconds. Cross-reactivity may result in the misidentification of a Salmonella 9-O serotype as a 2-O serotype.

    FDA · DeviceRemel Europe Ltd.Worldwide - US Nationwide distribution to the states of AL, CA, FL, IA, IL,…Official notice →