- 2026-08-19
US
Remel Europe Ltd. recalls Thermo Fisher Diagnostics Neisseria meningitidis Gp C 2ml/VL, Product Number R30166901; invitro diagnostics
Certain lots of Neisseria meningitidis Gp C 2ml/VL, R30166901, are not performing to IFU criteria not positive to ATCC13102 within 60 Seconds.
- 2026-04-29
US
Remel Europe Ltd. recalls Oxoid Agglutinating Sera, Salmonella 9-0 R30957301
Affected lots are not performing to IFU criteria showing cross reactions with 2-0 within 60 seconds. Cross-reactivity may result in the misidentification of a Salmonella 9-O serotype as a 2-O serotype.
FDA · DeviceRemel Europe Ltd.Worldwide - US Nationwide distribution to the states of AL, CA, FL, IA, IL,…Official notice →