Pivotal Health Solutions recalls

2 official recalls and safety alerts involving Pivotal Health Solutions since 2024-07, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.

This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Pivotal Health Solutions or the relevant agency.

  1. 2025-12-03
  2. US
    Drugs & medicalClass II

    Pivotal Health Solutions, Inc. recalls Quantum Intersegmental Table, Model/Catalog Number: Q400; heated massage table

    The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl…

    FDA · DevicePivotal Health Solutions, Inc.Distribution to US nationwide, Canada and Japan.Official notice →
  3. 2024-07-17
  4. US
    Drugs & medicalClass II

    Pivotal Health Solutions, Inc. recalls Decompression Table Model E9011. This device applies cervical and/or lumbar decompression with practitioner definable target…

    This product is being recalled due to potentially insufficient rework on a part (E7130-B) on the lumbar section of the frame on 6 distributed tables. There were 4 additional tables with the faulty E7130-B part that…

    FDA · DevicePivotal Health Solutions, Inc.Distributed to CA, FL, TX, ID, GA, and TNOfficial notice →