PATH recalls
2 official recalls and safety alerts involving PATH since 2023-12, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.
This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with PATH or the relevant agency.
- 2025-12-24
US
PATH recalls MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100)
Following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance thresholds. These distortions can…
- 2023-12-06
US
PATH recalls SENTIERO SCREENING, mobile medical devices for hearing examinations, Model Number SOH100098, Part Number 100098-S
Affected lots of product may experience technical distortions to a greater extent than expected.