- 2022-04-20
US
Nextremity Solutions recalls Nextra Hammertoe Correction System 3.5 Middle and 3.2 Proximal Saw Cut SKU: NX-3532K-SC Orthopedic surgical instrument
Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery
FDA · DeviceNextremity SolutionsDistribution US Nationwide, Japan, and the NetherlandsOfficial notice →- 2022-03-16
US
Nextremity Solutions recalls Nextra CH Cannulated Hammertoe Mini Instrument Kit-indicated for small bone reconstruction limited to inter-digital repair and fusion of…
During insertion of the middle phalanx implant into bone, the driver may deform the implant locking mechanism and result in inadequate mating force between the middle and proximal phalanx implants, resulting in…