ImaCor recalls

2 official recalls and safety alerts involving ImaCor since 2024-11, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.

This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with ImaCor or the relevant agency.

  1. 2025-10-15
  2. US
    Drugs & medicalClass II

    ImaCor Inc. recalls ClariTEE Miniaturized TEE probe. Model Number: CLT-010. Ultrasound miniaturized TEE probe for use with Mobile Ultrasound, ImaCor model ZHH.

    Production assembly error resulting in the potential for reversed articulation, where moving the lever forward caused the probe tip to anteflex rather than retroflex, and vice versa.

    FDA · DeviceImaCor Inc.US Nationwide distribution in the states of South Carolina, New Jersey,…Official notice →
  3. 2024-11-06
  4. US
    Drugs & medicalClass II

    ImaCor Inc. recalls ClariTEE Probe- Miniature transesophageal echo ultrasound probe for use with ImaCor ultrasound diagnostic platforms. Model/Catalog Number:…

    Epoxy used to seal the edge of the transducer into its paddle exceeded specification. This variability could attenuate the ultrasound signal with a potential impact on the quality of images obtained from the probes.

    FDA · DeviceImaCor Inc.FL ,GA, NJ, TNOfficial notice →