Hobbs Medical recalls

4 official recalls and safety alerts involving Hobbs Medical since 2022-03, from FDA · Device. Newest first. Check the official notice for exact models and batch codes.

This list only includes recalls Recall Atlas has collected since 2021-10. A product not listed here may still be affected; check with Hobbs Medical or the relevant agency.

  1. 2024-11-27
  2. US
    Drugs & medicalClass II

    Hobbs Medical, Inc. recalls Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4723

    Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the…

    FDA · DeviceHobbs Medical, Inc.Worldwide - US Nationwide distribution in the states of CT, GA, IN, MI, MN,…Official notice →
  3. 2022-10-19
  4. US
    Drugs & medicalClass II

    Hobbs Medical, Inc. recalls Hobbs Medical Polypectomy Snare Catalog Number: 4575-OLY

    Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and…

    FDA · DeviceHobbs Medical, Inc.Worldwide distribution - US Nationwide and the countries of Canada, Austria,…Official notice →
  5. 2022-07-27
  6. US
    Drugs & medicalClass II

    Hobbs Medical, Inc. recalls Hobbs Medical Polypectomy Snare, Catalog No. 7202

    During use, the sheath of the device may accordion at the proximal end where the sheath meets the shrink band. Kinking of the sheath may prevent the snare from fully closing, and injury may result.

    FDA · DeviceHobbs Medical, Inc.Domestic distribution to NC, TX, and WA. Foreign distribution to Canada.Official notice →
  7. 2022-03-09
  8. US
    Drugs & medicalClass II

    Hobbs Medical, Inc. recalls Bronchial Microbiology Brush, Catalog Number 4310 and 4320. Microbial sampling is performed by compressing a spring as the base of brush…

    The labeling for the Microbiology Brush identifies the device as a Bronchial Microbiology Brush. The device is only cleared for gastrointestinal use and is not cleared for bronchial use.

    FDA · DeviceHobbs Medical, Inc.US Nationwide distribution.Official notice →