- 2025-04-30
US
Globus Medical, Inc. recalls MAGEC 2 Rod, 5.0mm 90mm Standard; Version/Model: MC2-5090S;
Devices for the affected lot were not assembled according to product specifications. The distraction rod outer diameter was assembled with a 5.5mm rod versus the required 5.0mm.
- 2024-02-14
US
Globus Medical, Inc. recalls ExcelsiusGPS Cannula A, 9mm x 250mm, Cannula used with ball tip probes, Model No. 6143.2844
Devices may not be rendered sterile due to an internal gap that may be present on some devices causing increased difficulty cleaning prior to steam sterilization. There is an increased risk for infection, inflammation,…
FDA · DeviceGlobus Medical, Inc.Worldwide - US Nationwide and the countries of Australia, Austria, Belgium,…Official notice →US
Globus Medical, Inc. recalls Globus Medical Cranial Drill, 6.00mm - instruments for access, preparation, and placement of cranial devices with the ExcelsiusGPS. Ref:…
Sterile-packed Cranial Drills may not be sterile due to a breach in the packaging and/or may have discoloration near the drill tip.
FDA · DeviceGlobus Medical, Inc.Worldwide - US Nationwide and the countries of Czech Republic, Poland.Official notice →- 2024-01-24
US
Globus Medical, Inc. recalls ExcelsiusGPS Flat Panel Fluoroscopy Fixture Kit, Part number 6203.2100
A calibration error associated with the Flat Panel Fluoroscopy Fixture when used with ExcelsiusGPS for fluoroscopy workflow may lead to inaccurate navigation, which may result in device misplacement.
FDA · DeviceGlobus Medical, Inc.Domestic distribution to the following states: CA, CO, MT, SC, WA.…Official notice →